Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test

INTRODUCTION: Aortic valve bioprostheses ring fracture in valve-in-valve procedures has shown low complication rates and presents as an option in the treatment of patients at high risk for conventional surgery, avoiding high transvalvular gradients, which are associated with increased mortality. Som...

Descripción completa

Detalles Bibliográficos
Autores principales: Mota, Andre Lupp, Gaia, Diego Felipe, da Fonseca, José Honório Palma
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Cirurgia Cardiovascular 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9162417/
https://www.ncbi.nlm.nih.gov/pubmed/35605215
http://dx.doi.org/10.21470/1678-9741-2021-0272
_version_ 1784719700076265472
author Mota, Andre Lupp
Gaia, Diego Felipe
da Fonseca, José Honório Palma
author_facet Mota, Andre Lupp
Gaia, Diego Felipe
da Fonseca, José Honório Palma
author_sort Mota, Andre Lupp
collection PubMed
description INTRODUCTION: Aortic valve bioprostheses ring fracture in valve-in-valve procedures has shown low complication rates and presents as an option in the treatment of patients at high risk for conventional surgery, avoiding high transvalvular gradients, which are associated with increased mortality. Some prostheses available in the market cannot be fractured. In an ex vivo test, the possibility of ring fracture of aortic valve bioprostheses produced in Brazil when submitted to radial force application using a high-pressure non-compliant balloon was evaluated. METHODS: One unit of each aortic valve bioprosthesis model, sizes 19 and 21 mm, produced by Brazilian companies (Braile Biomédica, Cardioprótese, and Labcor), was used. In the experiment, a non-compliant high-pressure balloon (Atlas®-Gold), 1 mm larger than the external diameter of the prosthesis, was positioned inside the valve annulus and inflated gradually aiming to fracture the prosthesis. Fracture pressures and photographic and radiological images of the prostheses before and after test were recorded. RESULTS: All prostheses were fractured. In the models with metal ring, the fracture pressures were between 23 and 25 atm. In the other prostheses, the rupture occurred between 10 and 13 atm. No deformations in the structure were observed, which could potentially damage the aortic root. CONCLUSION: All the Brazilian prostheses evaluated were fractured, although the presence of a metal ring in the prosthesis framework increases the pressure required for fracture. The information obtained helps in the planning of valve-in-valve procedures in patients with aortic valve bioprostheses.
format Online
Article
Text
id pubmed-9162417
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Sociedade Brasileira de Cirurgia Cardiovascular
record_format MEDLINE/PubMed
spelling pubmed-91624172022-06-13 Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test Mota, Andre Lupp Gaia, Diego Felipe da Fonseca, José Honório Palma Braz J Cardiovasc Surg Original Article INTRODUCTION: Aortic valve bioprostheses ring fracture in valve-in-valve procedures has shown low complication rates and presents as an option in the treatment of patients at high risk for conventional surgery, avoiding high transvalvular gradients, which are associated with increased mortality. Some prostheses available in the market cannot be fractured. In an ex vivo test, the possibility of ring fracture of aortic valve bioprostheses produced in Brazil when submitted to radial force application using a high-pressure non-compliant balloon was evaluated. METHODS: One unit of each aortic valve bioprosthesis model, sizes 19 and 21 mm, produced by Brazilian companies (Braile Biomédica, Cardioprótese, and Labcor), was used. In the experiment, a non-compliant high-pressure balloon (Atlas®-Gold), 1 mm larger than the external diameter of the prosthesis, was positioned inside the valve annulus and inflated gradually aiming to fracture the prosthesis. Fracture pressures and photographic and radiological images of the prostheses before and after test were recorded. RESULTS: All prostheses were fractured. In the models with metal ring, the fracture pressures were between 23 and 25 atm. In the other prostheses, the rupture occurred between 10 and 13 atm. No deformations in the structure were observed, which could potentially damage the aortic root. CONCLUSION: All the Brazilian prostheses evaluated were fractured, although the presence of a metal ring in the prosthesis framework increases the pressure required for fracture. The information obtained helps in the planning of valve-in-valve procedures in patients with aortic valve bioprostheses. Sociedade Brasileira de Cirurgia Cardiovascular 2022 /pmc/articles/PMC9162417/ /pubmed/35605215 http://dx.doi.org/10.21470/1678-9741-2021-0272 Text en https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Mota, Andre Lupp
Gaia, Diego Felipe
da Fonseca, José Honório Palma
Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test
title Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test
title_full Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test
title_fullStr Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test
title_full_unstemmed Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test
title_short Ring Fracture of Brazilian Aortic Valve Bioprostheses Using Non-Compliant High-Pressure Transcatheter Balloon, an Ex Vivo Test
title_sort ring fracture of brazilian aortic valve bioprostheses using non-compliant high-pressure transcatheter balloon, an ex vivo test
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9162417/
https://www.ncbi.nlm.nih.gov/pubmed/35605215
http://dx.doi.org/10.21470/1678-9741-2021-0272
work_keys_str_mv AT motaandrelupp ringfractureofbrazilianaorticvalvebioprosthesesusingnoncomplianthighpressuretranscatheterballoonanexvivotest
AT gaiadiegofelipe ringfractureofbrazilianaorticvalvebioprosthesesusingnoncomplianthighpressuretranscatheterballoonanexvivotest
AT dafonsecajosehonoriopalma ringfractureofbrazilianaorticvalvebioprosthesesusingnoncomplianthighpressuretranscatheterballoonanexvivotest