Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard
The volumetric primary standard with a constant volume (pVTt) determines the flow rate from the rate of change of gas density in the defined measuring volume. The key element of the measuring system that uses the flying start–stop method is a diverter. This paper presents a time correction model tha...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9182970/ https://www.ncbi.nlm.nih.gov/pubmed/35684623 http://dx.doi.org/10.3390/s22114001 |
_version_ | 1784724171723374592 |
---|---|
author | Žibret, Primož Bobovnik, Gregor Kutin, Jože |
author_facet | Žibret, Primož Bobovnik, Gregor Kutin, Jože |
author_sort | Žibret, Primož |
collection | PubMed |
description | The volumetric primary standard with a constant volume (pVTt) determines the flow rate from the rate of change of gas density in the defined measuring volume. The key element of the measuring system that uses the flying start–stop method is a diverter. This paper presents a time correction model that adapts the correction according to the diverting speed implemented in our pVTt system developed for flow rates smaller than [Formula: see text]. A detailed study of the diverter’s effects on the effective collection time and two methods for determining the time correction are presented. One is based on the ISO 4185 standard and the other on measurements of the dynamic pressure upstream of the diverter. A set of correction time measurements were made at different diversion speeds to define the correction model at a flow rate of [Formula: see text]. The results show very good agreement between the measurements with both methods and the defined correction model. Additional measurements were made at smaller flow rates and the results indicate the independence of the time correction from the measured flow rate. |
format | Online Article Text |
id | pubmed-9182970 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-91829702022-06-10 Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard Žibret, Primož Bobovnik, Gregor Kutin, Jože Sensors (Basel) Article The volumetric primary standard with a constant volume (pVTt) determines the flow rate from the rate of change of gas density in the defined measuring volume. The key element of the measuring system that uses the flying start–stop method is a diverter. This paper presents a time correction model that adapts the correction according to the diverting speed implemented in our pVTt system developed for flow rates smaller than [Formula: see text]. A detailed study of the diverter’s effects on the effective collection time and two methods for determining the time correction are presented. One is based on the ISO 4185 standard and the other on measurements of the dynamic pressure upstream of the diverter. A set of correction time measurements were made at different diversion speeds to define the correction model at a flow rate of [Formula: see text]. The results show very good agreement between the measurements with both methods and the defined correction model. Additional measurements were made at smaller flow rates and the results indicate the independence of the time correction from the measured flow rate. MDPI 2022-05-25 /pmc/articles/PMC9182970/ /pubmed/35684623 http://dx.doi.org/10.3390/s22114001 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Žibret, Primož Bobovnik, Gregor Kutin, Jože Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard |
title | Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard |
title_full | Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard |
title_fullStr | Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard |
title_full_unstemmed | Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard |
title_short | Time-Correction Model Based on Diverter Speed for a pVTt Gas Flow Primary Standard |
title_sort | time-correction model based on diverter speed for a pvtt gas flow primary standard |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9182970/ https://www.ncbi.nlm.nih.gov/pubmed/35684623 http://dx.doi.org/10.3390/s22114001 |
work_keys_str_mv | AT zibretprimoz timecorrectionmodelbasedondiverterspeedforapvttgasflowprimarystandard AT bobovnikgregor timecorrectionmodelbasedondiverterspeedforapvttgasflowprimarystandard AT kutinjoze timecorrectionmodelbasedondiverterspeedforapvttgasflowprimarystandard |