Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics

The article considers the potential applicability of C12880MA and C11708MA Hamamatsu microspectrometers, which are characterized by an extremely compact design, occupying a small volume of several cubic centimeters, in portable spectrometric equipment with spatial resolution for monitoring the optic...

Descripción completa

Detalles Bibliográficos
Autores principales: Hotra, Oleksandra, Firago, Vladimir, Levkovich, Nikolay, Shuliko, Konstantin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9185238/
https://www.ncbi.nlm.nih.gov/pubmed/35684813
http://dx.doi.org/10.3390/s22114195
_version_ 1784724674240839680
author Hotra, Oleksandra
Firago, Vladimir
Levkovich, Nikolay
Shuliko, Konstantin
author_facet Hotra, Oleksandra
Firago, Vladimir
Levkovich, Nikolay
Shuliko, Konstantin
author_sort Hotra, Oleksandra
collection PubMed
description The article considers the potential applicability of C12880MA and C11708MA Hamamatsu microspectrometers, which are characterized by an extremely compact design, occupying a small volume of several cubic centimeters, in portable spectrometric equipment with spatial resolution for monitoring the optical properties of condensed scattering media. The development of methods for determining the reduced scattering and absorption spectral coefficients of radiation from various scattering materials and products allows us to speak about the possibility of real-time control of the volume concentration of optically active components included in them, for example, fat and water in dairy products. For this, it is necessary to provide sufficiently accurate spectra of diffusely reflected broadband light radiation at different distances between the points of radiation entrance and registration. The aim of the manuscript is to assess the possibility of using the considered microspectrometers in compact devices for optical diagnostics and control of the optical properties of condensed scattering media. The features of the connection diagram of these microspectrometers and the necessary methods for correcting the initially obtained spectral dependencies of diffusive reflection, which will be of interest to developers of spectral diagnostic equipment, are considered in detail. The need to eliminate the influence of the inhomogeneity of dark counts of a CMOS photodiode array is shown. The hardware functions of the C12880MA and C11708MA Hamammatsu microspectrometers, as well as the AvaSpec 2048L fiber-optic spectrometer, were experimentally measured and compared. Methods for correcting the nonlinearity of their reading scales and light characteristics, as well as improving their equivalent spectral resolution using digital Wiener filtering, are described. It is shown that the equivalent spectral resolution of C12880MA and C11708MA microspectrometers can be improved by about 40% when recording smooth spectra, subject to the condition that the resulting side oscillations are small. It is pointed out that in order to reduce the level of side oscillations in the corrected spectra with improved resolution, it is necessary to ensure the smoothness of the original spectra and a good signal-to-noise ratio. A conclusion is made about the possibility of using the considered microspectrometers in portable spectrometric equipment with careful consideration of their characteristics, the features of their switching circuit, and the necessary software.
format Online
Article
Text
id pubmed-9185238
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-91852382022-06-11 Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics Hotra, Oleksandra Firago, Vladimir Levkovich, Nikolay Shuliko, Konstantin Sensors (Basel) Article The article considers the potential applicability of C12880MA and C11708MA Hamamatsu microspectrometers, which are characterized by an extremely compact design, occupying a small volume of several cubic centimeters, in portable spectrometric equipment with spatial resolution for monitoring the optical properties of condensed scattering media. The development of methods for determining the reduced scattering and absorption spectral coefficients of radiation from various scattering materials and products allows us to speak about the possibility of real-time control of the volume concentration of optically active components included in them, for example, fat and water in dairy products. For this, it is necessary to provide sufficiently accurate spectra of diffusely reflected broadband light radiation at different distances between the points of radiation entrance and registration. The aim of the manuscript is to assess the possibility of using the considered microspectrometers in compact devices for optical diagnostics and control of the optical properties of condensed scattering media. The features of the connection diagram of these microspectrometers and the necessary methods for correcting the initially obtained spectral dependencies of diffusive reflection, which will be of interest to developers of spectral diagnostic equipment, are considered in detail. The need to eliminate the influence of the inhomogeneity of dark counts of a CMOS photodiode array is shown. The hardware functions of the C12880MA and C11708MA Hamammatsu microspectrometers, as well as the AvaSpec 2048L fiber-optic spectrometer, were experimentally measured and compared. Methods for correcting the nonlinearity of their reading scales and light characteristics, as well as improving their equivalent spectral resolution using digital Wiener filtering, are described. It is shown that the equivalent spectral resolution of C12880MA and C11708MA microspectrometers can be improved by about 40% when recording smooth spectra, subject to the condition that the resulting side oscillations are small. It is pointed out that in order to reduce the level of side oscillations in the corrected spectra with improved resolution, it is necessary to ensure the smoothness of the original spectra and a good signal-to-noise ratio. A conclusion is made about the possibility of using the considered microspectrometers in portable spectrometric equipment with careful consideration of their characteristics, the features of their switching circuit, and the necessary software. MDPI 2022-05-31 /pmc/articles/PMC9185238/ /pubmed/35684813 http://dx.doi.org/10.3390/s22114195 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Hotra, Oleksandra
Firago, Vladimir
Levkovich, Nikolay
Shuliko, Konstantin
Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics
title Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics
title_full Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics
title_fullStr Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics
title_full_unstemmed Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics
title_short Investigation of the Possibility of Using Microspectrometers Based on CMOS Photodiode Arrays in Small-Sized Devices for Optical Diagnostics
title_sort investigation of the possibility of using microspectrometers based on cmos photodiode arrays in small-sized devices for optical diagnostics
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9185238/
https://www.ncbi.nlm.nih.gov/pubmed/35684813
http://dx.doi.org/10.3390/s22114195
work_keys_str_mv AT hotraoleksandra investigationofthepossibilityofusingmicrospectrometersbasedoncmosphotodiodearraysinsmallsizeddevicesforopticaldiagnostics
AT firagovladimir investigationofthepossibilityofusingmicrospectrometersbasedoncmosphotodiodearraysinsmallsizeddevicesforopticaldiagnostics
AT levkovichnikolay investigationofthepossibilityofusingmicrospectrometersbasedoncmosphotodiodearraysinsmallsizeddevicesforopticaldiagnostics
AT shulikokonstantin investigationofthepossibilityofusingmicrospectrometersbasedoncmosphotodiodearraysinsmallsizeddevicesforopticaldiagnostics