Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture

BACKGROUND: There are a variety of internal fixation methods for unstable femoral neck fractures (FNFs), but the best method is still unclear. Femoral neck system (FNS) is a dynamic angular stabilization system with cross screws, and is a new internal fixation implant designed for minimally invasive...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Zheng, Yang, Yong, Feng, Gangning, Guo, Haohui, Chen, Zhirong, Chen, Yaogeng, Jin, Qunhua
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9188240/
https://www.ncbi.nlm.nih.gov/pubmed/35689282
http://dx.doi.org/10.1186/s12938-022-01006-6
_version_ 1784725330956648448
author Wang, Zheng
Yang, Yong
Feng, Gangning
Guo, Haohui
Chen, Zhirong
Chen, Yaogeng
Jin, Qunhua
author_facet Wang, Zheng
Yang, Yong
Feng, Gangning
Guo, Haohui
Chen, Zhirong
Chen, Yaogeng
Jin, Qunhua
author_sort Wang, Zheng
collection PubMed
description BACKGROUND: There are a variety of internal fixation methods for unstable femoral neck fractures (FNFs), but the best method is still unclear. Femoral neck system (FNS) is a dynamic angular stabilization system with cross screws, and is a new internal fixation implant designed for minimally invasive fixation of FNFs. In this study, we conducted a biomechanical comparison of FNS, InterTan nail and three cannulated screws for the treatment of Pauwels III FNFs and investigate the biomechanical properties of FNS. METHODS: A total of 18 left artificial femurs were selected and randomly divide into Group A (fixation with FNS), Group B (fixation with InterTan nail) and Group C (fixation with three cannulated screws), with 6 specimens in each group. After creating Pauwels type III FNF models, the specimens in each were tested with non-destructive quasi-static tests, including torsion, A-P bending and axial compression tests. The average slope of the linear load–deformation curve obtained from quasi-static tests defines the initial torsional stiffness, A-P bending stiffness, and axial compression stiffness. After cyclic loading test was applied, the overall deformation of models and local deformation of implant holes in each group were assessed. The overall deformation was estimated as the displacement recorded by the software of the mechanical testing apparatus. Local deformation was defined as interfragmental displacement. Data were analyzed by one-way analysis of variance (ANOVA) followed by Bonferroni post hoc test using the SPSS software (version 24.0, IBM, New York, NY, USA). Correlation analysis was performed using Pearson’s correlation analysis. RESULTS: Group B exhibited significantly higher axial stiffness and A-P bending stiffness than the other two groups (P < 0.01), while Group A had significantly higher axial stiffness and A-P bending stiffness than Group C (P < 0.01). Groups A and B exhibited significantly higher torsional stiffness than Group C (P < 0.01), no statistical significance was observed between Groups A and B (P > 0.05). Group B exhibited significantly lower overall and local deformations than the other two groups (P < 0.01), while Group A had significantly lower overall and local deformations than Group C (P < 0.01). Correlation analysis revealed positive correlation between axial stiffness and A-P bending stiffness (r = 0.925, P < 0.01), torsional stiffness (r = 0.727, P < 0.01), between torsional stiffness and A-P bending stiffness; negative correlation between overall, local deformations and axial stiffness (r = − 0.889, − 0.901, respectively, both P < 0.01), and positive correlation between the two deformations (r = − 0.978, P < 0.01). CONCLUSION: For fixation of unstable FNFs, InterTan nail showed the highest axial stiffness and A-P bending stiffness, followed by FNS, and then three cannulated screws. Torsional stiffness of FNS was comparable to that of the InterTan nail. FNS, as a novel minimally invasive implant, can create good mechanical environment for the healing of unstable FNFs. Clinical studies are needed to confirm the potential advantages of FNS observed in this biomechanical study.
format Online
Article
Text
id pubmed-9188240
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-91882402022-06-12 Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture Wang, Zheng Yang, Yong Feng, Gangning Guo, Haohui Chen, Zhirong Chen, Yaogeng Jin, Qunhua Biomed Eng Online Research BACKGROUND: There are a variety of internal fixation methods for unstable femoral neck fractures (FNFs), but the best method is still unclear. Femoral neck system (FNS) is a dynamic angular stabilization system with cross screws, and is a new internal fixation implant designed for minimally invasive fixation of FNFs. In this study, we conducted a biomechanical comparison of FNS, InterTan nail and three cannulated screws for the treatment of Pauwels III FNFs and investigate the biomechanical properties of FNS. METHODS: A total of 18 left artificial femurs were selected and randomly divide into Group A (fixation with FNS), Group B (fixation with InterTan nail) and Group C (fixation with three cannulated screws), with 6 specimens in each group. After creating Pauwels type III FNF models, the specimens in each were tested with non-destructive quasi-static tests, including torsion, A-P bending and axial compression tests. The average slope of the linear load–deformation curve obtained from quasi-static tests defines the initial torsional stiffness, A-P bending stiffness, and axial compression stiffness. After cyclic loading test was applied, the overall deformation of models and local deformation of implant holes in each group were assessed. The overall deformation was estimated as the displacement recorded by the software of the mechanical testing apparatus. Local deformation was defined as interfragmental displacement. Data were analyzed by one-way analysis of variance (ANOVA) followed by Bonferroni post hoc test using the SPSS software (version 24.0, IBM, New York, NY, USA). Correlation analysis was performed using Pearson’s correlation analysis. RESULTS: Group B exhibited significantly higher axial stiffness and A-P bending stiffness than the other two groups (P < 0.01), while Group A had significantly higher axial stiffness and A-P bending stiffness than Group C (P < 0.01). Groups A and B exhibited significantly higher torsional stiffness than Group C (P < 0.01), no statistical significance was observed between Groups A and B (P > 0.05). Group B exhibited significantly lower overall and local deformations than the other two groups (P < 0.01), while Group A had significantly lower overall and local deformations than Group C (P < 0.01). Correlation analysis revealed positive correlation between axial stiffness and A-P bending stiffness (r = 0.925, P < 0.01), torsional stiffness (r = 0.727, P < 0.01), between torsional stiffness and A-P bending stiffness; negative correlation between overall, local deformations and axial stiffness (r = − 0.889, − 0.901, respectively, both P < 0.01), and positive correlation between the two deformations (r = − 0.978, P < 0.01). CONCLUSION: For fixation of unstable FNFs, InterTan nail showed the highest axial stiffness and A-P bending stiffness, followed by FNS, and then three cannulated screws. Torsional stiffness of FNS was comparable to that of the InterTan nail. FNS, as a novel minimally invasive implant, can create good mechanical environment for the healing of unstable FNFs. Clinical studies are needed to confirm the potential advantages of FNS observed in this biomechanical study. BioMed Central 2022-06-10 /pmc/articles/PMC9188240/ /pubmed/35689282 http://dx.doi.org/10.1186/s12938-022-01006-6 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Wang, Zheng
Yang, Yong
Feng, Gangning
Guo, Haohui
Chen, Zhirong
Chen, Yaogeng
Jin, Qunhua
Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
title Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
title_full Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
title_fullStr Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
title_full_unstemmed Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
title_short Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
title_sort biomechanical comparison of the femoral neck system versus intertan nail and three cannulated screws for unstable pauwels type iii femoral neck fracture
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9188240/
https://www.ncbi.nlm.nih.gov/pubmed/35689282
http://dx.doi.org/10.1186/s12938-022-01006-6
work_keys_str_mv AT wangzheng biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture
AT yangyong biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture
AT fenggangning biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture
AT guohaohui biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture
AT chenzhirong biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture
AT chenyaogeng biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture
AT jinqunhua biomechanicalcomparisonofthefemoralnecksystemversusintertannailandthreecannulatedscrewsforunstablepauwelstypeiiifemoralneckfracture