Cargando…
Development and evaluation of a numerical simulation approach to predict metal artifacts from passive implants in MRI
OBJECTIVE: This study presents the development and evaluation of a numerical approach to simulate artifacts of metallic implants in an MR environment that can be applied to improve the testing procedure for MR image artifacts in medical implants according to ASTM F2119. METHODS: The numerical approa...
Autores principales: | Spronk, Tobias, Kraff, Oliver, Kreutner, Jakob, Schaefers, Gregor, Quick, Harald H. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9188622/ https://www.ncbi.nlm.nih.gov/pubmed/34655346 http://dx.doi.org/10.1007/s10334-021-00966-5 |
Ejemplares similares
-
Numerical approach to investigate MR imaging artifacts from orthopedic implants at different field strengths according to ASTM F2119
por: Spronk, Tobias, et al.
Publicado: (2023) -
Comparison of Cochlear Implant Magnets and Their MRI Artifact Size
por: Todt, I., et al.
Publicado: (2020) -
Simulation and Mitigation of the Wrap-Around Artifact in the MRI Image
por: Hu, Runze, et al.
Publicado: (2021) -
Influence of head orientation and implantation site of a novel transcutaneous bone conduction implant on MRI metal artifact reduction sequence
por: Talon, Emile, et al.
Publicado: (2022) -
Metal artifact correction strategies in MRI-based attenuation correction in PET/MRI
por: Schramm, Georg, et al.
Publicado: (2019)