Cargando…
Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories
INTRODUCTION: The modern approach to quality control (QC) in medical laboratories implies the development of a risk-based control plan. This paper aims to develop a risk-based QC plan for a laboratory with a small daily testing volume and to integrate the already optimized moving average (MA) contro...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9195605/ https://www.ncbi.nlm.nih.gov/pubmed/35799981 http://dx.doi.org/10.11613/BM.2022.020711 |
_version_ | 1784727002227408896 |
---|---|
author | Lukić, Vera Ignjatović, Svetlana |
author_facet | Lukić, Vera Ignjatović, Svetlana |
author_sort | Lukić, Vera |
collection | PubMed |
description | INTRODUCTION: The modern approach to quality control (QC) in medical laboratories implies the development of a risk-based control plan. This paper aims to develop a risk-based QC plan for a laboratory with a small daily testing volume and to integrate the already optimized moving average (MA) control procedures into this plan. MATERIALS AND METHODS: A multistage bracketed QC plan for ten clinical chemistry analytes was made using a Westgard QC frequency calculator. Previously, MA procedures were optimized by the bias detection simulation method. RESULTS: Aspartate aminotransferase, HDL-cholesterol and potassium had patient-risk sigma metrics greater than 6, albumin and cholesterol greater than 5, creatinine, chlorides, calcium and total proteins between 4 and 5, and sodium less than 4. Based on the calculated run sizes and characteristics of optimized MA procedures, for 6 tests, it was possible to replace the monitoring QC procedure with an MA procedure. For the remaining 4 tests, it was necessary to keep the monitoring QC procedure and introduce MA control for added security. CONCLUSION: This study showed that even in a laboratory with a small volume of daily testing, it is possible to make a risk-based QC plan and integrate MA control procedures into that plan. |
format | Online Article Text |
id | pubmed-9195605 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Croatian Society of Medical Biochemistry and Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-91956052022-07-06 Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories Lukić, Vera Ignjatović, Svetlana Biochem Med (Zagreb) Original Articles INTRODUCTION: The modern approach to quality control (QC) in medical laboratories implies the development of a risk-based control plan. This paper aims to develop a risk-based QC plan for a laboratory with a small daily testing volume and to integrate the already optimized moving average (MA) control procedures into this plan. MATERIALS AND METHODS: A multistage bracketed QC plan for ten clinical chemistry analytes was made using a Westgard QC frequency calculator. Previously, MA procedures were optimized by the bias detection simulation method. RESULTS: Aspartate aminotransferase, HDL-cholesterol and potassium had patient-risk sigma metrics greater than 6, albumin and cholesterol greater than 5, creatinine, chlorides, calcium and total proteins between 4 and 5, and sodium less than 4. Based on the calculated run sizes and characteristics of optimized MA procedures, for 6 tests, it was possible to replace the monitoring QC procedure with an MA procedure. For the remaining 4 tests, it was necessary to keep the monitoring QC procedure and introduce MA control for added security. CONCLUSION: This study showed that even in a laboratory with a small volume of daily testing, it is possible to make a risk-based QC plan and integrate MA control procedures into that plan. Croatian Society of Medical Biochemistry and Laboratory Medicine 2022-06-15 2022-06-15 /pmc/articles/PMC9195605/ /pubmed/35799981 http://dx.doi.org/10.11613/BM.2022.020711 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Articles Lukić, Vera Ignjatović, Svetlana Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
title | Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
title_full | Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
title_fullStr | Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
title_full_unstemmed | Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
title_short | Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
title_sort | integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9195605/ https://www.ncbi.nlm.nih.gov/pubmed/35799981 http://dx.doi.org/10.11613/BM.2022.020711 |
work_keys_str_mv | AT lukicvera integratingmovingaveragecontrolproceduresintotheriskbasedqualitycontrolplaninsmallvolumemedicallaboratories AT ignjatovicsvetlana integratingmovingaveragecontrolproceduresintotheriskbasedqualitycontrolplaninsmallvolumemedicallaboratories |