Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories
INTRODUCTION: The modern approach to quality control (QC) in medical laboratories implies the development of a risk-based control plan. This paper aims to develop a risk-based QC plan for a laboratory with a small daily testing volume and to integrate the already optimized moving average (MA) contro...
Autores principales: | Lukić, Vera, Ignjatović, Svetlana |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Croatian Society of Medical Biochemistry and Laboratory Medicine
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9195605/ https://www.ncbi.nlm.nih.gov/pubmed/35799981 http://dx.doi.org/10.11613/BM.2022.020711 |
Ejemplares similares
-
Optimizing moving average control procedures for small-volume laboratories: can it be done?
por: Lukić, Vera, et al.
Publicado: (2019) -
Moving average procedures as an additional tool for real-time analytical quality control: challenges and opportunities of implementation in small-volume medical laboratories
por: Lukić, Vera, et al.
Publicado: (2021) -
Performance of new nonparametric Tukey modified exponentially weighted moving average—Moving average control chart
por: Talordphop, Khanittha, et al.
Publicado: (2022) -
Patient-based quality control for glucometers: using the moving sum of positive patient results and moving average
por: Lim, Chun Yee, et al.
Publicado: (2020) -
Exponentially weighted moving average—Moving average charts for monitoring the process mean
por: Sukparungsee, Saowanit, et al.
Publicado: (2020)