Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser

INTRODUCTION: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI)...

Descripción completa

Detalles Bibliográficos
Autores principales: Buño Soto, Antonio, Peoc’h, Katell, Fasano, Tommaso, Diaz-Garzon, Jorge, González de la Presa, Bernardino, Chicha-Cattoir, Valerie, Canovi, Simone, de Pedro, Maria Sanz, Rico, Nayra, Robert, Tiphaine, Bonelli, Efrem, Calle, Pilar Fernández, Mira, Aurea, Lefevre, Guillaume, Vecchia, Luigi, Bedini, Jose Luis
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9195607/
https://www.ncbi.nlm.nih.gov/pubmed/35799986
http://dx.doi.org/10.11613/BM.2022.020709
_version_ 1784727002768474112
author Buño Soto, Antonio
Peoc’h, Katell
Fasano, Tommaso
Diaz-Garzon, Jorge
González de la Presa, Bernardino
Chicha-Cattoir, Valerie
Canovi, Simone
de Pedro, Maria Sanz
Rico, Nayra
Robert, Tiphaine
Bonelli, Efrem
Calle, Pilar Fernández
Mira, Aurea
Lefevre, Guillaume
Vecchia, Luigi
Bedini, Jose Luis
author_facet Buño Soto, Antonio
Peoc’h, Katell
Fasano, Tommaso
Diaz-Garzon, Jorge
González de la Presa, Bernardino
Chicha-Cattoir, Valerie
Canovi, Simone
de Pedro, Maria Sanz
Rico, Nayra
Robert, Tiphaine
Bonelli, Efrem
Calle, Pilar Fernández
Mira, Aurea
Lefevre, Guillaume
Vecchia, Luigi
Bedini, Jose Luis
author_sort Buño Soto, Antonio
collection PubMed
description INTRODUCTION: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe. MATERIALS AND METHODS: Precision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer’s cut-offs. Sample turnaround time (TAT) was examined. RESULTS: Imprecision repeatability CVs were 1.1–4.7% and within-lab imprecision were 1.8–7.6% (10.0–25,000 ng/L). The limit of blank (LoB), detection (LoD), and quantitation (LoQ) aligned with the manufacturer’s values of 0.5 ng/L, 1.6 ng/L, and 2.5 ng/L, respectively. Passing-Bablok regression demonstrated good correlations between Atellica IM analyser with other systems; some minor deviations were observed. All results in healthy volunteers fell below the 99th percentile URL, and greater than 50% of each sex demonstrated values above the LoD. No interference was observed for biotin (≤ 1500 µg/L), but a slight bias at 5.0 g/L haemoglobin and 50 ng/L Tn was observed. TAT from was fast (mean time = 10.9 minutes) and reproducible (6%CV). CONCLUSIONS: Real-world analytical and TAT performance of the hs-TnI assay on the Atellica IM analyser make this assay fit for routine use in clinical laboratories.
format Online
Article
Text
id pubmed-9195607
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-91956072022-07-06 Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser Buño Soto, Antonio Peoc’h, Katell Fasano, Tommaso Diaz-Garzon, Jorge González de la Presa, Bernardino Chicha-Cattoir, Valerie Canovi, Simone de Pedro, Maria Sanz Rico, Nayra Robert, Tiphaine Bonelli, Efrem Calle, Pilar Fernández Mira, Aurea Lefevre, Guillaume Vecchia, Luigi Bedini, Jose Luis Biochem Med (Zagreb) Original Articles INTRODUCTION: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe. MATERIALS AND METHODS: Precision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer’s cut-offs. Sample turnaround time (TAT) was examined. RESULTS: Imprecision repeatability CVs were 1.1–4.7% and within-lab imprecision were 1.8–7.6% (10.0–25,000 ng/L). The limit of blank (LoB), detection (LoD), and quantitation (LoQ) aligned with the manufacturer’s values of 0.5 ng/L, 1.6 ng/L, and 2.5 ng/L, respectively. Passing-Bablok regression demonstrated good correlations between Atellica IM analyser with other systems; some minor deviations were observed. All results in healthy volunteers fell below the 99th percentile URL, and greater than 50% of each sex demonstrated values above the LoD. No interference was observed for biotin (≤ 1500 µg/L), but a slight bias at 5.0 g/L haemoglobin and 50 ng/L Tn was observed. TAT from was fast (mean time = 10.9 minutes) and reproducible (6%CV). CONCLUSIONS: Real-world analytical and TAT performance of the hs-TnI assay on the Atellica IM analyser make this assay fit for routine use in clinical laboratories. Croatian Society of Medical Biochemistry and Laboratory Medicine 2022-06-15 2022-06-15 /pmc/articles/PMC9195607/ /pubmed/35799986 http://dx.doi.org/10.11613/BM.2022.020709 Text en Croatian Society of Medical Biochemistry and Laboratory Medicine. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Buño Soto, Antonio
Peoc’h, Katell
Fasano, Tommaso
Diaz-Garzon, Jorge
González de la Presa, Bernardino
Chicha-Cattoir, Valerie
Canovi, Simone
de Pedro, Maria Sanz
Rico, Nayra
Robert, Tiphaine
Bonelli, Efrem
Calle, Pilar Fernández
Mira, Aurea
Lefevre, Guillaume
Vecchia, Luigi
Bedini, Jose Luis
Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
title Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
title_full Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
title_fullStr Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
title_full_unstemmed Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
title_short Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
title_sort performance evaluation of the high sensitive troponin i assay on the atellica im analyser
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9195607/
https://www.ncbi.nlm.nih.gov/pubmed/35799986
http://dx.doi.org/10.11613/BM.2022.020709
work_keys_str_mv AT bunosotoantonio performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT peochkatell performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT fasanotommaso performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT diazgarzonjorge performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT gonzalezdelapresabernardino performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT chichacattoirvalerie performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT canovisimone performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT depedromariasanz performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT riconayra performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT roberttiphaine performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT bonelliefrem performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT callepilarfernandez performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT miraaurea performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT lefevreguillaume performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT vecchialuigi performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser
AT bedinijoseluis performanceevaluationofthehighsensitivetroponiniassayontheatellicaimanalyser