Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study

BACKGROUND: The aim of this in vitro study was to determine the accuracy in measuring the working length (WL) using Dentaport ZX, Rootor, and a newly introduced budget friendly electronic apex locator (EAL), E-Pex Pro in two commonly simulated clinical conditions (in the presence of irrigant and blo...

Descripción completa

Detalles Bibliográficos
Autores principales: Mohan, Malavika, Verma, Meenakshi R, Jain, Ashish K, Rao, Rahul D, Yadav, Priyanka, Agrawal, Sonal
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9200183/
https://www.ncbi.nlm.nih.gov/pubmed/35722068
http://dx.doi.org/10.4103/jcd.jcd_557_21
_version_ 1784728008383266816
author Mohan, Malavika
Verma, Meenakshi R
Jain, Ashish K
Rao, Rahul D
Yadav, Priyanka
Agrawal, Sonal
author_facet Mohan, Malavika
Verma, Meenakshi R
Jain, Ashish K
Rao, Rahul D
Yadav, Priyanka
Agrawal, Sonal
author_sort Mohan, Malavika
collection PubMed
description BACKGROUND: The aim of this in vitro study was to determine the accuracy in measuring the working length (WL) using Dentaport ZX, Rootor, and a newly introduced budget friendly electronic apex locator (EAL), E-Pex Pro in two commonly simulated clinical conditions (in the presence of irrigant and blood). MATERIALS AND METHODOLOGY: Eighty-eight single-rooted premolars were randomly assigned to two groups according to simulated clinical conditions: Group 1: Presence of irrigant (5% NaOCl) and Group 2: Presence of blood. WL was determined with all three EAL and was then compared with actual length (AL) of the tooth, which was measured using Vision Inspection System. The difference between the length measured by EAL and AL was tabulated and statistically analyzed using one-way analysis of variance (ANOVA) with post hoc Dunnett's test. All testing was done using two-sided tests at alpha 0.05 (95% confidence level). Thus, the criteria for rejecting the null hypothesis were “P < 0.05.” RESULTS: Measurement using Dentaport ZX, Rootor, and E-Pex Pro had an accuracy of 99.79%, 99.69%, and 99.64%, respectively, in Group 1 and 99.95%, 99.7%, and 99.74%, respectively, in Group 2. ANOVA revealed that the mean error value is least for Dentaport ZX followed by Rootor and then E-Pex pro EAL. CONCLUSION: Dentaport ZX gave better results both in the presence of NaOCl and blood followed by Rootor and E-Pex Pro EALs.
format Online
Article
Text
id pubmed-9200183
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-92001832022-06-16 Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study Mohan, Malavika Verma, Meenakshi R Jain, Ashish K Rao, Rahul D Yadav, Priyanka Agrawal, Sonal J Conserv Dent Original Article BACKGROUND: The aim of this in vitro study was to determine the accuracy in measuring the working length (WL) using Dentaport ZX, Rootor, and a newly introduced budget friendly electronic apex locator (EAL), E-Pex Pro in two commonly simulated clinical conditions (in the presence of irrigant and blood). MATERIALS AND METHODOLOGY: Eighty-eight single-rooted premolars were randomly assigned to two groups according to simulated clinical conditions: Group 1: Presence of irrigant (5% NaOCl) and Group 2: Presence of blood. WL was determined with all three EAL and was then compared with actual length (AL) of the tooth, which was measured using Vision Inspection System. The difference between the length measured by EAL and AL was tabulated and statistically analyzed using one-way analysis of variance (ANOVA) with post hoc Dunnett's test. All testing was done using two-sided tests at alpha 0.05 (95% confidence level). Thus, the criteria for rejecting the null hypothesis were “P < 0.05.” RESULTS: Measurement using Dentaport ZX, Rootor, and E-Pex Pro had an accuracy of 99.79%, 99.69%, and 99.64%, respectively, in Group 1 and 99.95%, 99.7%, and 99.74%, respectively, in Group 2. ANOVA revealed that the mean error value is least for Dentaport ZX followed by Rootor and then E-Pex pro EAL. CONCLUSION: Dentaport ZX gave better results both in the presence of NaOCl and blood followed by Rootor and E-Pex Pro EALs. Wolters Kluwer - Medknow 2022 2022-05-02 /pmc/articles/PMC9200183/ /pubmed/35722068 http://dx.doi.org/10.4103/jcd.jcd_557_21 Text en Copyright: © 2022 Journal of Conservative Dentistry https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Mohan, Malavika
Verma, Meenakshi R
Jain, Ashish K
Rao, Rahul D
Yadav, Priyanka
Agrawal, Sonal
Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study
title Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study
title_full Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study
title_fullStr Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study
title_full_unstemmed Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study
title_short Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An In Vitro Study
title_sort comparison of accuracy of dentaport zx, rootor and e-pex pro electronic apex locators in two simulated clinical conditions: an in vitro study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9200183/
https://www.ncbi.nlm.nih.gov/pubmed/35722068
http://dx.doi.org/10.4103/jcd.jcd_557_21
work_keys_str_mv AT mohanmalavika comparisonofaccuracyofdentaportzxrootorandepexproelectronicapexlocatorsintwosimulatedclinicalconditionsaninvitrostudy
AT vermameenakshir comparisonofaccuracyofdentaportzxrootorandepexproelectronicapexlocatorsintwosimulatedclinicalconditionsaninvitrostudy
AT jainashishk comparisonofaccuracyofdentaportzxrootorandepexproelectronicapexlocatorsintwosimulatedclinicalconditionsaninvitrostudy
AT raorahuld comparisonofaccuracyofdentaportzxrootorandepexproelectronicapexlocatorsintwosimulatedclinicalconditionsaninvitrostudy
AT yadavpriyanka comparisonofaccuracyofdentaportzxrootorandepexproelectronicapexlocatorsintwosimulatedclinicalconditionsaninvitrostudy
AT agrawalsonal comparisonofaccuracyofdentaportzxrootorandepexproelectronicapexlocatorsintwosimulatedclinicalconditionsaninvitrostudy