Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT

OBJECTIVE(S): Somatostatin receptor scintigraphy (SRS) using (111)In-pentetreotide has no established quantification method. The purpose of this study was to develop a new quantitative method to correct the partial volume effect (PVE) for individual energy peaks in (111)In-pentetreotide single-photo...

Descripción completa

Detalles Bibliográficos
Autores principales: Yamashita, Kosuke, Miyaji, Noriaki, Motegi, Kazuki, Terauchi, Takashi, Ito, Shigeki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Mashhad University of Medical Sciences 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9205847/
https://www.ncbi.nlm.nih.gov/pubmed/35800427
http://dx.doi.org/10.22038/AOJNMB.2022.61394.1430
_version_ 1784729215884591104
author Yamashita, Kosuke
Miyaji, Noriaki
Motegi, Kazuki
Terauchi, Takashi
Ito, Shigeki
author_facet Yamashita, Kosuke
Miyaji, Noriaki
Motegi, Kazuki
Terauchi, Takashi
Ito, Shigeki
author_sort Yamashita, Kosuke
collection PubMed
description OBJECTIVE(S): Somatostatin receptor scintigraphy (SRS) using (111)In-pentetreotide has no established quantification method. The purpose of this study was to develop a new quantitative method to correct the partial volume effect (PVE) for individual energy peaks in (111)In-pentetreotide single-photon emission computed tomography (SPECT). METHODS: Phantom experiments were performed to construct a new quantitative method. In the phantom experiments, a NEMA IEC body phantom was used. Acquisition was performed using two energy peaks (171 keV and 245 keV) on the SPECT/CT system. The volume of interest was set at each hot sphere and lung insert in the SPECT images of each energy peak, and the recovery coefficient (RC) was calculated to understand the PVE. A new quantitative index, the indium uptake index (IUI), was calculated using the RC to correct the PVE. The quantitative accuracy of the IUI in the hot sphere was confirmed. Case studies were performed to clarify the quantitative accuracy. In a case study, the relationship between the IUI and the Krenning score, which is used as a visual assessment, was evaluated for each lesion. RESULTS: The obtained RCs showed that the energy peak at 171 keV was faster in recovering the effect of PVE than that at 245 keV. The IUI in the 17-mm-diameter hot sphere was overestimated by 4.8% and 8.3% at 171 keV and 245 keV, respectively, compared to the actual IUIs. The relationship between IUI and Krenning score was rs=0.773 (p<0.005) at sum, rs=0.739 (p<0.005) at 171 keV, and rs=0.773 (p<0.005) at 245 keV. CONCLUSION: We have developed a new quantification method for (111)In-pentetreotide SPECT/CT using RC-based PVE correction for an individual energy peak of 171 keV. The quantitative accuracy of this method was high even for accumulations of less than 20 mm, and it showed a good relationship with the Krenning score; therefore, the clinical usefulness of IUI was demonstrated.
format Online
Article
Text
id pubmed-9205847
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Mashhad University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-92058472022-07-06 Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT Yamashita, Kosuke Miyaji, Noriaki Motegi, Kazuki Terauchi, Takashi Ito, Shigeki Asia Ocean J Nucl Med Biol Technical Note OBJECTIVE(S): Somatostatin receptor scintigraphy (SRS) using (111)In-pentetreotide has no established quantification method. The purpose of this study was to develop a new quantitative method to correct the partial volume effect (PVE) for individual energy peaks in (111)In-pentetreotide single-photon emission computed tomography (SPECT). METHODS: Phantom experiments were performed to construct a new quantitative method. In the phantom experiments, a NEMA IEC body phantom was used. Acquisition was performed using two energy peaks (171 keV and 245 keV) on the SPECT/CT system. The volume of interest was set at each hot sphere and lung insert in the SPECT images of each energy peak, and the recovery coefficient (RC) was calculated to understand the PVE. A new quantitative index, the indium uptake index (IUI), was calculated using the RC to correct the PVE. The quantitative accuracy of the IUI in the hot sphere was confirmed. Case studies were performed to clarify the quantitative accuracy. In a case study, the relationship between the IUI and the Krenning score, which is used as a visual assessment, was evaluated for each lesion. RESULTS: The obtained RCs showed that the energy peak at 171 keV was faster in recovering the effect of PVE than that at 245 keV. The IUI in the 17-mm-diameter hot sphere was overestimated by 4.8% and 8.3% at 171 keV and 245 keV, respectively, compared to the actual IUIs. The relationship between IUI and Krenning score was rs=0.773 (p<0.005) at sum, rs=0.739 (p<0.005) at 171 keV, and rs=0.773 (p<0.005) at 245 keV. CONCLUSION: We have developed a new quantification method for (111)In-pentetreotide SPECT/CT using RC-based PVE correction for an individual energy peak of 171 keV. The quantitative accuracy of this method was high even for accumulations of less than 20 mm, and it showed a good relationship with the Krenning score; therefore, the clinical usefulness of IUI was demonstrated. Mashhad University of Medical Sciences 2022 /pmc/articles/PMC9205847/ /pubmed/35800427 http://dx.doi.org/10.22038/AOJNMB.2022.61394.1430 Text en © 2022 mums.ac.ir All rights reserved https://creativecommons.org/licenses/by/3.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License, (http://creativecommons.org/licenses/by/3.0/ (https://creativecommons.org/licenses/by/3.0/) ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Note
Yamashita, Kosuke
Miyaji, Noriaki
Motegi, Kazuki
Terauchi, Takashi
Ito, Shigeki
Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT
title Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT
title_full Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT
title_fullStr Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT
title_full_unstemmed Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT
title_short Development of a new quantification method using partial volume effect correction for individual energy peaks in (111)In-pentetreotide SPECT/CT
title_sort development of a new quantification method using partial volume effect correction for individual energy peaks in (111)in-pentetreotide spect/ct
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9205847/
https://www.ncbi.nlm.nih.gov/pubmed/35800427
http://dx.doi.org/10.22038/AOJNMB.2022.61394.1430
work_keys_str_mv AT yamashitakosuke developmentofanewquantificationmethodusingpartialvolumeeffectcorrectionforindividualenergypeaksin111inpentetreotidespectct
AT miyajinoriaki developmentofanewquantificationmethodusingpartialvolumeeffectcorrectionforindividualenergypeaksin111inpentetreotidespectct
AT motegikazuki developmentofanewquantificationmethodusingpartialvolumeeffectcorrectionforindividualenergypeaksin111inpentetreotidespectct
AT terauchitakashi developmentofanewquantificationmethodusingpartialvolumeeffectcorrectionforindividualenergypeaksin111inpentetreotidespectct
AT itoshigeki developmentofanewquantificationmethodusingpartialvolumeeffectcorrectionforindividualenergypeaksin111inpentetreotidespectct