Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation
PURPOSE: The purpose of this study was to evaluate three different methods of attachment of continuous loop suspensory cortical preparation of all soft tissue central quad tendon grafts compared to a bone block control for anterior cruciate ligament reconstruction on construct displacement and load...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9210367/ https://www.ncbi.nlm.nih.gov/pubmed/35747631 http://dx.doi.org/10.1016/j.asmr.2021.12.008 |
_version_ | 1784730153152151552 |
---|---|
author | Arakgi, Michelle E. Burkhart, Timothy A. Hoshino, Takashi Degen, Ryan Getgood, Alan |
author_facet | Arakgi, Michelle E. Burkhart, Timothy A. Hoshino, Takashi Degen, Ryan Getgood, Alan |
author_sort | Arakgi, Michelle E. |
collection | PubMed |
description | PURPOSE: The purpose of this study was to evaluate three different methods of attachment of continuous loop suspensory cortical preparation of all soft tissue central quad tendon grafts compared to a bone block control for anterior cruciate ligament reconstruction on construct displacement and load to failure. METHODS: Thirty-two cadaveric central quadriceps tendon (CQT) specimens were harvested, using three clinical techniques for graft fixation: cortical button alone (BTB EB), BTB cortical button with rip-stop suture (BTB RS), and continuous loop cortical button (BTB CL). A control group was also included that consisted of a bone block secured within testing clamps (BTB CON). Specimens were preloaded to 150 N. Tendons were then cyclically loaded between 50 N and 250 N for 1,000 cycles at .5 Hz. Displacement was measured at the point of fixation of the CQT after the 150 N preload, 250 N initial load and every 100th cycle. The specimens were loaded to failure after 1,000 cycles. RESULTS: There was a significant increase in displacement from .32 ± .56 mm for the BTB CON to 1.91 ± 1.13 mm for the BTB RS (P = .014) and 3.85 ± 2.32 mm for the BTB CL condition (P = .023). There was no significant increase in displacement for BTB EB (P = .182). Failure occurred for all of the BTB CL and 62.5% of the BTB EB specimens within the first 50 cycles. Twenty-five percent of the BTB CON specimens and 12.5% BTB RS failed at ∼400 and 500 cycles, respectively. Similar failure loads were observed for the BTB CON and the BTB RS (446.4 ± 151.46 N vs 505.74 ± 131.41 N; P = .99) Failure testing was not feasible for the BTB CL and BTB EB preparation methods. CONCLUSION: In response to cyclic loading, the three all-soft tissue suspensory conditions experienced significantly greater displacement compared to the bone block controls. None of the soft-tissue conditions appeared superior when compared to each other. CLINICAL RELEVANCE: It remains unknown which method of soft-tissue suspensory provides optimal fixation. As these autografts become more common, it is essential to evaluate which fixation methods provide superior outcomes |
format | Online Article Text |
id | pubmed-9210367 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-92103672022-06-22 Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation Arakgi, Michelle E. Burkhart, Timothy A. Hoshino, Takashi Degen, Ryan Getgood, Alan Arthrosc Sports Med Rehabil Original Article PURPOSE: The purpose of this study was to evaluate three different methods of attachment of continuous loop suspensory cortical preparation of all soft tissue central quad tendon grafts compared to a bone block control for anterior cruciate ligament reconstruction on construct displacement and load to failure. METHODS: Thirty-two cadaveric central quadriceps tendon (CQT) specimens were harvested, using three clinical techniques for graft fixation: cortical button alone (BTB EB), BTB cortical button with rip-stop suture (BTB RS), and continuous loop cortical button (BTB CL). A control group was also included that consisted of a bone block secured within testing clamps (BTB CON). Specimens were preloaded to 150 N. Tendons were then cyclically loaded between 50 N and 250 N for 1,000 cycles at .5 Hz. Displacement was measured at the point of fixation of the CQT after the 150 N preload, 250 N initial load and every 100th cycle. The specimens were loaded to failure after 1,000 cycles. RESULTS: There was a significant increase in displacement from .32 ± .56 mm for the BTB CON to 1.91 ± 1.13 mm for the BTB RS (P = .014) and 3.85 ± 2.32 mm for the BTB CL condition (P = .023). There was no significant increase in displacement for BTB EB (P = .182). Failure occurred for all of the BTB CL and 62.5% of the BTB EB specimens within the first 50 cycles. Twenty-five percent of the BTB CON specimens and 12.5% BTB RS failed at ∼400 and 500 cycles, respectively. Similar failure loads were observed for the BTB CON and the BTB RS (446.4 ± 151.46 N vs 505.74 ± 131.41 N; P = .99) Failure testing was not feasible for the BTB CL and BTB EB preparation methods. CONCLUSION: In response to cyclic loading, the three all-soft tissue suspensory conditions experienced significantly greater displacement compared to the bone block controls. None of the soft-tissue conditions appeared superior when compared to each other. CLINICAL RELEVANCE: It remains unknown which method of soft-tissue suspensory provides optimal fixation. As these autografts become more common, it is essential to evaluate which fixation methods provide superior outcomes Elsevier 2022-03-26 /pmc/articles/PMC9210367/ /pubmed/35747631 http://dx.doi.org/10.1016/j.asmr.2021.12.008 Text en © 2022 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Original Article Arakgi, Michelle E. Burkhart, Timothy A. Hoshino, Takashi Degen, Ryan Getgood, Alan Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation |
title | Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation |
title_full | Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation |
title_fullStr | Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation |
title_full_unstemmed | Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation |
title_short | Biomechanical Comparison of Three Suspensory Techniques for all Soft Tissue Central Quadriceps Tendon Graft Fixation |
title_sort | biomechanical comparison of three suspensory techniques for all soft tissue central quadriceps tendon graft fixation |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9210367/ https://www.ncbi.nlm.nih.gov/pubmed/35747631 http://dx.doi.org/10.1016/j.asmr.2021.12.008 |
work_keys_str_mv | AT arakgimichellee biomechanicalcomparisonofthreesuspensorytechniquesforallsofttissuecentralquadricepstendongraftfixation AT burkharttimothya biomechanicalcomparisonofthreesuspensorytechniquesforallsofttissuecentralquadricepstendongraftfixation AT hoshinotakashi biomechanicalcomparisonofthreesuspensorytechniquesforallsofttissuecentralquadricepstendongraftfixation AT degenryan biomechanicalcomparisonofthreesuspensorytechniquesforallsofttissuecentralquadricepstendongraftfixation AT getgoodalan biomechanicalcomparisonofthreesuspensorytechniquesforallsofttissuecentralquadricepstendongraftfixation |