Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results

Our aim was to compare the image quality and patient dose of contrast-enhanced oncologic chest-CT of a first-generation photon-counting detector (PCD-CT) and a second-generation dual-source dual-energy CT (DSCT). For this reason, one hundred consecutive oncologic patients (63 male, 65 ± 11 years, BM...

Descripción completa

Detalles Bibliográficos
Autores principales: Hagen, Florian, Walder, Lukas, Fritz, Jan, Gutjahr, Ralf, Schmidt, Bernhard, Faby, Sebastian, Bamberg, Fabian, Schoenberg, Stefan, Nikolaou, Konstantin, Horger, Marius
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9227736/
https://www.ncbi.nlm.nih.gov/pubmed/35736867
http://dx.doi.org/10.3390/tomography8030119
_version_ 1784734254965456896
author Hagen, Florian
Walder, Lukas
Fritz, Jan
Gutjahr, Ralf
Schmidt, Bernhard
Faby, Sebastian
Bamberg, Fabian
Schoenberg, Stefan
Nikolaou, Konstantin
Horger, Marius
author_facet Hagen, Florian
Walder, Lukas
Fritz, Jan
Gutjahr, Ralf
Schmidt, Bernhard
Faby, Sebastian
Bamberg, Fabian
Schoenberg, Stefan
Nikolaou, Konstantin
Horger, Marius
author_sort Hagen, Florian
collection PubMed
description Our aim was to compare the image quality and patient dose of contrast-enhanced oncologic chest-CT of a first-generation photon-counting detector (PCD-CT) and a second-generation dual-source dual-energy CT (DSCT). For this reason, one hundred consecutive oncologic patients (63 male, 65 ± 11 years, BMI: 16–42 kg/m(2)) were prospectively enrolled and evaluated. Clinically indicated contrast-enhanced chest-CT were obtained with PCD-CT and compared to previously obtained chest-DSCT in the same individuals. The median time interval between the scans was three months. The same contrast media protocol was used for both scans. PCD-CT was performed in QuantumPlus mode (obtaining full spectral information) at 120 kVp. DSCT was performed using 100 kV for Tube A and 140 kV for Tube B. “T3D” PCD-CT images were evaluated, which emulate conventional 120 keV polychromatic images. For DSCT, the convolution algorithm was set at I31f with class 1 iterative reconstruction, whereas comparable Br40 kernel and iterative reconstruction strengths (Q1 and Q3) were applied for PCD-CT. Two radiologists assessed image quality using a five-point Likert scale and performed measurements of vessels and lung parenchyma for signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and in the case of pulmonary metastases tumor-to-lung parenchyma contrast ratio. PCD-CT CNR(vessel) was significantly higher than DSCT CNR(vessel) (all, p < 0.05). Readers rated image contrast of mediastinum, vessels, and lung parenchyma significantly higher in PCD-CT than DSCT images (p < 0.001). Q3 PCD-CT CNR(lung_parenchyma) was significantly higher than DSCT CNR(lung_parenchyma) and Q1 PCD-CT CNR(lung_parenchyma) (p < 0.01). The tumor-to-lung parenchyma contrast ratio was significantly higher on PCD-CT than DSCT images (0.08 ± 0.04 vs. 0.03 ± 0.02, p < 0.001). CTDI, DLP, SSDE mean values for PCD-CT and DSCT were 4.17 ± 1.29 mGy vs. 7.21 ± 0.49 mGy, 151.01 ± 48.56 mGy * cm vs. 288.64 ± 31.17 mGy * cm and 4.23 ± 0.97 vs. 7.48 ± 1.09, respectively. PCD-CT enables oncologic chest-CT with a significantly reduced dose while maintaining image quality similar to a second-generation DSCT for comparable protocol settings.
format Online
Article
Text
id pubmed-9227736
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-92277362022-06-25 Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results Hagen, Florian Walder, Lukas Fritz, Jan Gutjahr, Ralf Schmidt, Bernhard Faby, Sebastian Bamberg, Fabian Schoenberg, Stefan Nikolaou, Konstantin Horger, Marius Tomography Article Our aim was to compare the image quality and patient dose of contrast-enhanced oncologic chest-CT of a first-generation photon-counting detector (PCD-CT) and a second-generation dual-source dual-energy CT (DSCT). For this reason, one hundred consecutive oncologic patients (63 male, 65 ± 11 years, BMI: 16–42 kg/m(2)) were prospectively enrolled and evaluated. Clinically indicated contrast-enhanced chest-CT were obtained with PCD-CT and compared to previously obtained chest-DSCT in the same individuals. The median time interval between the scans was three months. The same contrast media protocol was used for both scans. PCD-CT was performed in QuantumPlus mode (obtaining full spectral information) at 120 kVp. DSCT was performed using 100 kV for Tube A and 140 kV for Tube B. “T3D” PCD-CT images were evaluated, which emulate conventional 120 keV polychromatic images. For DSCT, the convolution algorithm was set at I31f with class 1 iterative reconstruction, whereas comparable Br40 kernel and iterative reconstruction strengths (Q1 and Q3) were applied for PCD-CT. Two radiologists assessed image quality using a five-point Likert scale and performed measurements of vessels and lung parenchyma for signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and in the case of pulmonary metastases tumor-to-lung parenchyma contrast ratio. PCD-CT CNR(vessel) was significantly higher than DSCT CNR(vessel) (all, p < 0.05). Readers rated image contrast of mediastinum, vessels, and lung parenchyma significantly higher in PCD-CT than DSCT images (p < 0.001). Q3 PCD-CT CNR(lung_parenchyma) was significantly higher than DSCT CNR(lung_parenchyma) and Q1 PCD-CT CNR(lung_parenchyma) (p < 0.01). The tumor-to-lung parenchyma contrast ratio was significantly higher on PCD-CT than DSCT images (0.08 ± 0.04 vs. 0.03 ± 0.02, p < 0.001). CTDI, DLP, SSDE mean values for PCD-CT and DSCT were 4.17 ± 1.29 mGy vs. 7.21 ± 0.49 mGy, 151.01 ± 48.56 mGy * cm vs. 288.64 ± 31.17 mGy * cm and 4.23 ± 0.97 vs. 7.48 ± 1.09, respectively. PCD-CT enables oncologic chest-CT with a significantly reduced dose while maintaining image quality similar to a second-generation DSCT for comparable protocol settings. MDPI 2022-06-03 /pmc/articles/PMC9227736/ /pubmed/35736867 http://dx.doi.org/10.3390/tomography8030119 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Hagen, Florian
Walder, Lukas
Fritz, Jan
Gutjahr, Ralf
Schmidt, Bernhard
Faby, Sebastian
Bamberg, Fabian
Schoenberg, Stefan
Nikolaou, Konstantin
Horger, Marius
Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
title Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
title_full Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
title_fullStr Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
title_full_unstemmed Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
title_short Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
title_sort image quality and radiation dose of contrast-enhanced chest-ct acquired on a clinical photon-counting detector ct vs. second-generation dual-source ct in an oncologic cohort: preliminary results
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9227736/
https://www.ncbi.nlm.nih.gov/pubmed/35736867
http://dx.doi.org/10.3390/tomography8030119
work_keys_str_mv AT hagenflorian imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT walderlukas imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT fritzjan imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT gutjahrralf imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT schmidtbernhard imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT fabysebastian imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT bambergfabian imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT schoenbergstefan imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT nikolaoukonstantin imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults
AT horgermarius imagequalityandradiationdoseofcontrastenhancedchestctacquiredonaclinicalphotoncountingdetectorctvssecondgenerationdualsourcectinanoncologiccohortpreliminaryresults