Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation

BACKGROUND: Left ventricular assist devices (LVADs) are implantable pumps that act as a life support therapy for patients with severe heart failure. Despite improving the survival rate, LVAD therapy can carry major complications. Particularly, the flow distortion introduced by the LVAD in the left v...

Descripción completa

Detalles Bibliográficos
Autores principales: Santiago, Alfonso, Butakoff, Constantine, Eguzkitza, Beatriz, Gray, Richard A., May-Newman, Karen, Pathmanathan, Pras, Vu, Vi, Vázquez, Mariano
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9232142/
https://www.ncbi.nlm.nih.gov/pubmed/35696442
http://dx.doi.org/10.1371/journal.pcbi.1010141
_version_ 1784735507816644608
author Santiago, Alfonso
Butakoff, Constantine
Eguzkitza, Beatriz
Gray, Richard A.
May-Newman, Karen
Pathmanathan, Pras
Vu, Vi
Vázquez, Mariano
author_facet Santiago, Alfonso
Butakoff, Constantine
Eguzkitza, Beatriz
Gray, Richard A.
May-Newman, Karen
Pathmanathan, Pras
Vu, Vi
Vázquez, Mariano
author_sort Santiago, Alfonso
collection PubMed
description BACKGROUND: Left ventricular assist devices (LVADs) are implantable pumps that act as a life support therapy for patients with severe heart failure. Despite improving the survival rate, LVAD therapy can carry major complications. Particularly, the flow distortion introduced by the LVAD in the left ventricle (LV) may induce thrombus formation. While previous works have used numerical models to study the impact of multiple variables in the intra-LV stagnation regions, a comprehensive validation analysis has never been executed. The main goal of this work is to present a model of the LV-LVAD system and to design and follow a verification, validation and uncertainty quantification (VVUQ) plan based on the ASME V&V40 and V&V20 standards to ensure credible predictions. METHODS: The experiment used to validate the simulation is the SDSU cardiac simulator, a bench mock-up of the cardiovascular system that allows mimicking multiple operation conditions for the heart-LVAD system. The numerical model is based on Alya, the BSC’s in-house platform for numerical modelling. Alya solves the Navier-Stokes equation with an Arbitrary Lagrangian-Eulerian (ALE) formulation in a deformable ventricle and includes pressure-driven valves, a 0D Windkessel model for the arterial output and a LVAD boundary condition modeled through a dynamic pressure-flow performance curve. The designed VVUQ plan involves: (a) a risk analysis and the associated credibility goals; (b) a verification stage to ensure correctness in the numerical solution procedure; (c) a sensitivity analysis to quantify the impact of the inputs on the four quantities of interest (QoIs) (average aortic root flow [Image: see text] , maximum aortic root flow [Image: see text] , average LVAD flow [Image: see text] , and maximum LVAD flow [Image: see text] ); (d) an uncertainty quantification using six validation experiments that include extreme operating conditions. RESULTS: Numerical code verification tests ensured correctness of the solution procedure and numerical calculation verification showed a grid convergence index (GCI)(95%) <3.3%. The total Sobol indices obtained during the sensitivity analysis demonstrated that the ejection fraction, the heart rate, and the pump performance curve coefficients are the most impactful inputs for the analysed QoIs. The Minkowski norm is used as validation metric for the uncertainty quantification. It shows that the midpoint cases have more accurate results when compared to the extreme cases. The total computational cost of the simulations was above 100 [core-years] executed in around three weeks time span in Marenostrum IV supercomputer. CONCLUSIONS: This work details a novel numerical model for the LV-LVAD system, that is supported by the design and execution of a VVUQ plan created following recognised international standards. We present a methodology demonstrating that stringent VVUQ according to ASME standards is feasible but computationally expensive.
format Online
Article
Text
id pubmed-9232142
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-92321422022-06-25 Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation Santiago, Alfonso Butakoff, Constantine Eguzkitza, Beatriz Gray, Richard A. May-Newman, Karen Pathmanathan, Pras Vu, Vi Vázquez, Mariano PLoS Comput Biol Research Article BACKGROUND: Left ventricular assist devices (LVADs) are implantable pumps that act as a life support therapy for patients with severe heart failure. Despite improving the survival rate, LVAD therapy can carry major complications. Particularly, the flow distortion introduced by the LVAD in the left ventricle (LV) may induce thrombus formation. While previous works have used numerical models to study the impact of multiple variables in the intra-LV stagnation regions, a comprehensive validation analysis has never been executed. The main goal of this work is to present a model of the LV-LVAD system and to design and follow a verification, validation and uncertainty quantification (VVUQ) plan based on the ASME V&V40 and V&V20 standards to ensure credible predictions. METHODS: The experiment used to validate the simulation is the SDSU cardiac simulator, a bench mock-up of the cardiovascular system that allows mimicking multiple operation conditions for the heart-LVAD system. The numerical model is based on Alya, the BSC’s in-house platform for numerical modelling. Alya solves the Navier-Stokes equation with an Arbitrary Lagrangian-Eulerian (ALE) formulation in a deformable ventricle and includes pressure-driven valves, a 0D Windkessel model for the arterial output and a LVAD boundary condition modeled through a dynamic pressure-flow performance curve. The designed VVUQ plan involves: (a) a risk analysis and the associated credibility goals; (b) a verification stage to ensure correctness in the numerical solution procedure; (c) a sensitivity analysis to quantify the impact of the inputs on the four quantities of interest (QoIs) (average aortic root flow [Image: see text] , maximum aortic root flow [Image: see text] , average LVAD flow [Image: see text] , and maximum LVAD flow [Image: see text] ); (d) an uncertainty quantification using six validation experiments that include extreme operating conditions. RESULTS: Numerical code verification tests ensured correctness of the solution procedure and numerical calculation verification showed a grid convergence index (GCI)(95%) <3.3%. The total Sobol indices obtained during the sensitivity analysis demonstrated that the ejection fraction, the heart rate, and the pump performance curve coefficients are the most impactful inputs for the analysed QoIs. The Minkowski norm is used as validation metric for the uncertainty quantification. It shows that the midpoint cases have more accurate results when compared to the extreme cases. The total computational cost of the simulations was above 100 [core-years] executed in around three weeks time span in Marenostrum IV supercomputer. CONCLUSIONS: This work details a novel numerical model for the LV-LVAD system, that is supported by the design and execution of a VVUQ plan created following recognised international standards. We present a methodology demonstrating that stringent VVUQ according to ASME standards is feasible but computationally expensive. Public Library of Science 2022-06-13 /pmc/articles/PMC9232142/ /pubmed/35696442 http://dx.doi.org/10.1371/journal.pcbi.1010141 Text en https://creativecommons.org/publicdomain/zero/1.0/This is an open access article, free of all copyright, and may be freely reproduced, distributed, transmitted, modified, built upon, or otherwise used by anyone for any lawful purpose. The work is made available under the Creative Commons CC0 (https://creativecommons.org/publicdomain/zero/1.0/) public domain dedication.
spellingShingle Research Article
Santiago, Alfonso
Butakoff, Constantine
Eguzkitza, Beatriz
Gray, Richard A.
May-Newman, Karen
Pathmanathan, Pras
Vu, Vi
Vázquez, Mariano
Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation
title Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation
title_full Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation
title_fullStr Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation
title_full_unstemmed Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation
title_short Design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after LVAD implantation
title_sort design and execution of a verification, validation, and uncertainty quantification plan for a numerical model of left ventricular flow after lvad implantation
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9232142/
https://www.ncbi.nlm.nih.gov/pubmed/35696442
http://dx.doi.org/10.1371/journal.pcbi.1010141
work_keys_str_mv AT santiagoalfonso designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT butakoffconstantine designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT eguzkitzabeatriz designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT grayricharda designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT maynewmankaren designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT pathmanathanpras designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT vuvi designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation
AT vazquezmariano designandexecutionofaverificationvalidationanduncertaintyquantificationplanforanumericalmodelofleftventricularflowafterlvadimplantation