Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900

BACKGROUND: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TA...

Descripción completa

Detalles Bibliográficos
Autores principales: Ruge, Hendrik, Alvarez-Covarrubias, Hector A., Deutsch, Oliver, Alalawi, Zahra, Vitanova, Keti, Lange, Rüdiger
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9238356/
https://www.ncbi.nlm.nih.gov/pubmed/35774373
http://dx.doi.org/10.3389/fcvm.2022.859088
_version_ 1784737032147304448
author Ruge, Hendrik
Alvarez-Covarrubias, Hector A.
Deutsch, Oliver
Alalawi, Zahra
Vitanova, Keti
Lange, Rüdiger
author_facet Ruge, Hendrik
Alvarez-Covarrubias, Hector A.
Deutsch, Oliver
Alalawi, Zahra
Vitanova, Keti
Lange, Rüdiger
author_sort Ruge, Hendrik
collection PubMed
description BACKGROUND: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR. METHODS: Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF. RESULTS: High-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect. CONCLUSION: Both 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics.
format Online
Article
Text
id pubmed-9238356
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-92383562022-06-29 Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 Ruge, Hendrik Alvarez-Covarrubias, Hector A. Deutsch, Oliver Alalawi, Zahra Vitanova, Keti Lange, Rüdiger Front Cardiovasc Med Cardiovascular Medicine BACKGROUND: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR. METHODS: Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF. RESULTS: High-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect. CONCLUSION: Both 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics. Frontiers Media S.A. 2022-06-02 /pmc/articles/PMC9238356/ /pubmed/35774373 http://dx.doi.org/10.3389/fcvm.2022.859088 Text en Copyright © 2022 Ruge, Alvarez-Covarrubias, Deutsch, Alalawi, Vitanova and Lange. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Cardiovascular Medicine
Ruge, Hendrik
Alvarez-Covarrubias, Hector A.
Deutsch, Oliver
Alalawi, Zahra
Vitanova, Keti
Lange, Rüdiger
Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
title Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
title_full Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
title_fullStr Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
title_full_unstemmed Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
title_short Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
title_sort bioprosthetic valve fracturing: in vitro testing of edwards perimount model p 2900
topic Cardiovascular Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9238356/
https://www.ncbi.nlm.nih.gov/pubmed/35774373
http://dx.doi.org/10.3389/fcvm.2022.859088
work_keys_str_mv AT rugehendrik bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900
AT alvarezcovarrubiashectora bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900
AT deutscholiver bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900
AT alalawizahra bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900
AT vitanovaketi bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900
AT langerudiger bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900