Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900
BACKGROUND: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TA...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9238356/ https://www.ncbi.nlm.nih.gov/pubmed/35774373 http://dx.doi.org/10.3389/fcvm.2022.859088 |
_version_ | 1784737032147304448 |
---|---|
author | Ruge, Hendrik Alvarez-Covarrubias, Hector A. Deutsch, Oliver Alalawi, Zahra Vitanova, Keti Lange, Rüdiger |
author_facet | Ruge, Hendrik Alvarez-Covarrubias, Hector A. Deutsch, Oliver Alalawi, Zahra Vitanova, Keti Lange, Rüdiger |
author_sort | Ruge, Hendrik |
collection | PubMed |
description | BACKGROUND: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR. METHODS: Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF. RESULTS: High-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect. CONCLUSION: Both 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics. |
format | Online Article Text |
id | pubmed-9238356 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-92383562022-06-29 Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 Ruge, Hendrik Alvarez-Covarrubias, Hector A. Deutsch, Oliver Alalawi, Zahra Vitanova, Keti Lange, Rüdiger Front Cardiovasc Med Cardiovascular Medicine BACKGROUND: Bioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR. METHODS: Using four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF. RESULTS: High-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect. CONCLUSION: Both 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics. Frontiers Media S.A. 2022-06-02 /pmc/articles/PMC9238356/ /pubmed/35774373 http://dx.doi.org/10.3389/fcvm.2022.859088 Text en Copyright © 2022 Ruge, Alvarez-Covarrubias, Deutsch, Alalawi, Vitanova and Lange. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Cardiovascular Medicine Ruge, Hendrik Alvarez-Covarrubias, Hector A. Deutsch, Oliver Alalawi, Zahra Vitanova, Keti Lange, Rüdiger Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 |
title | Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 |
title_full | Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 |
title_fullStr | Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 |
title_full_unstemmed | Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 |
title_short | Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900 |
title_sort | bioprosthetic valve fracturing: in vitro testing of edwards perimount model p 2900 |
topic | Cardiovascular Medicine |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9238356/ https://www.ncbi.nlm.nih.gov/pubmed/35774373 http://dx.doi.org/10.3389/fcvm.2022.859088 |
work_keys_str_mv | AT rugehendrik bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900 AT alvarezcovarrubiashectora bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900 AT deutscholiver bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900 AT alalawizahra bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900 AT vitanovaketi bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900 AT langerudiger bioprostheticvalvefracturinginvitrotestingofedwardsperimountmodelp2900 |