Cargando…

Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature

A solution revision prosthesis has a multilayer microporous Porocoat coating, and the availability of multiple stem body sizes ensures that the prosthesis is adapted to each patient’s anatomical structure so that there a firm attachment with the bone cortex in the middle of the femur. Therefore, the...

Descripción completa

Detalles Bibliográficos
Autores principales: Yuan, Long, Li, Sen, Li, Wanxiang, Bian, Jichao, Bao, Yahui, Zhou, Xiaopeng, Zhang, Yuanmin, Li, Wang, Wang, Guodong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9239620/
https://www.ncbi.nlm.nih.gov/pubmed/35777021
http://dx.doi.org/10.1097/MD.0000000000029811
_version_ 1784737339966226432
author Yuan, Long
Li, Sen
Li, Wanxiang
Bian, Jichao
Bao, Yahui
Zhou, Xiaopeng
Zhang, Yuanmin
Li, Wang
Wang, Guodong
author_facet Yuan, Long
Li, Sen
Li, Wanxiang
Bian, Jichao
Bao, Yahui
Zhou, Xiaopeng
Zhang, Yuanmin
Li, Wang
Wang, Guodong
author_sort Yuan, Long
collection PubMed
description A solution revision prosthesis has a multilayer microporous Porocoat coating, and the availability of multiple stem body sizes ensures that the prosthesis is adapted to each patient’s anatomical structure so that there a firm attachment with the bone cortex in the middle of the femur. Therefore, the Solution prosthesis is one of the most commonly used and most effective prostheses in total hip arthroplasty worldwide. PATIENT CONCERNS: We reported a case of a 54-year-old female patient with periprosthetic femoral fractures after hip arthroplasty. DIAGNOSIS: The case was identified as type B2 prosthesis loosening according to the Vancouver classification. INTERVENTIONS: We performed revision surgery on her using the Solution prosthesis. Seven months after the surgery, the patient developed a mid-femoral prosthesis fracture for no apparent reason. We performed a second revision surgery of the hip joint and allogeneic bone plate fixation. OUTCOMES: The patient was satisfied with the treatment. LESSONS: For patients with type B2 prosthesis loosening and prosthesis fracture, hip arthroplasty revision and an allogeneic bone plate could be used to ensure more stable support.
format Online
Article
Text
id pubmed-9239620
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-92396202022-06-30 Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature Yuan, Long Li, Sen Li, Wanxiang Bian, Jichao Bao, Yahui Zhou, Xiaopeng Zhang, Yuanmin Li, Wang Wang, Guodong Medicine (Baltimore) Research Article A solution revision prosthesis has a multilayer microporous Porocoat coating, and the availability of multiple stem body sizes ensures that the prosthesis is adapted to each patient’s anatomical structure so that there a firm attachment with the bone cortex in the middle of the femur. Therefore, the Solution prosthesis is one of the most commonly used and most effective prostheses in total hip arthroplasty worldwide. PATIENT CONCERNS: We reported a case of a 54-year-old female patient with periprosthetic femoral fractures after hip arthroplasty. DIAGNOSIS: The case was identified as type B2 prosthesis loosening according to the Vancouver classification. INTERVENTIONS: We performed revision surgery on her using the Solution prosthesis. Seven months after the surgery, the patient developed a mid-femoral prosthesis fracture for no apparent reason. We performed a second revision surgery of the hip joint and allogeneic bone plate fixation. OUTCOMES: The patient was satisfied with the treatment. LESSONS: For patients with type B2 prosthesis loosening and prosthesis fracture, hip arthroplasty revision and an allogeneic bone plate could be used to ensure more stable support. Lippincott Williams & Wilkins 2022-06-30 /pmc/articles/PMC9239620/ /pubmed/35777021 http://dx.doi.org/10.1097/MD.0000000000029811 Text en Copyright © 2022 the Author(s). Published by Wolters Kluwer Health, Inc. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License 4.0 (CCBY) (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Yuan, Long
Li, Sen
Li, Wanxiang
Bian, Jichao
Bao, Yahui
Zhou, Xiaopeng
Zhang, Yuanmin
Li, Wang
Wang, Guodong
Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature
title Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature
title_full Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature
title_fullStr Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature
title_full_unstemmed Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature
title_short Femoral prosthesis fracture after hip arthroplasty revision: A Case Report and Review of Literature
title_sort femoral prosthesis fracture after hip arthroplasty revision: a case report and review of literature
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9239620/
https://www.ncbi.nlm.nih.gov/pubmed/35777021
http://dx.doi.org/10.1097/MD.0000000000029811
work_keys_str_mv AT yuanlong femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT lisen femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT liwanxiang femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT bianjichao femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT baoyahui femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT zhouxiaopeng femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT zhangyuanmin femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT liwang femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature
AT wangguodong femoralprosthesisfractureafterhiparthroplastyrevisionacasereportandreviewofliterature