The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study

Multi-Dynamic Multi-Echo (MDME) Sequence is a new method which can acquire various contrast-weighted images using quantitative relaxometric parameters measured from multicontrast images. The purpose of our study was to investigate the effect of scan parameters of MDME Sequence on measured T1, T2 val...

Descripción completa

Detalles Bibliográficos
Autores principales: Zheng, Zuofeng, Yang, Jiafei, Zhang, Dongpo, Ma, Jun, Yin, Hongxia, Liu, Yawen, Wang, Zhenchang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9239947/
https://www.ncbi.nlm.nih.gov/pubmed/35553390
http://dx.doi.org/10.1007/s13246-022-01128-0
_version_ 1784737427127009280
author Zheng, Zuofeng
Yang, Jiafei
Zhang, Dongpo
Ma, Jun
Yin, Hongxia
Liu, Yawen
Wang, Zhenchang
author_facet Zheng, Zuofeng
Yang, Jiafei
Zhang, Dongpo
Ma, Jun
Yin, Hongxia
Liu, Yawen
Wang, Zhenchang
author_sort Zheng, Zuofeng
collection PubMed
description Multi-Dynamic Multi-Echo (MDME) Sequence is a new method which can acquire various contrast-weighted images using quantitative relaxometric parameters measured from multicontrast images. The purpose of our study was to investigate the effect of scan parameters of MDME Sequence on measured T1, T2 values of phantoms at 3.0 T MRI scanner. Gray matter, white matter and cerebrospinal fluid simulation phantoms with different relaxation times (named GM, WM, CSF, respectively) were used in our study. All the phantoms were scanned 9 times on different days using MDME sequence with variations of echo train length, matrix, and acceleration factor. The T1, T2 measurements were acquired after each acquisition. The repeatability was characterized as the intragroup coefficient of variation (CV) of measured values over 9 times, and the discrepancies of measurements across different groups were characterized as intergroup CVs. The highest intragroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 1.36%, 1.75%, 0.74%, 1.41%, 1.70%, 7.79%, respectively. The highest intergroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 0.54%, 1.86%, 1.70%, 0.94%, 1.00%, 2.17%, respectively. Quantitative T1, T2 measurements of gray matter, white matter and cerebrospinal fluid simulation phantoms derived from the MDME sequence were not obviously affected by variations of scanning parameters, such as echo train length, matrix, and acceleration factor on 3T scanner.
format Online
Article
Text
id pubmed-9239947
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-92399472022-06-30 The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study Zheng, Zuofeng Yang, Jiafei Zhang, Dongpo Ma, Jun Yin, Hongxia Liu, Yawen Wang, Zhenchang Phys Eng Sci Med Scientific Note Multi-Dynamic Multi-Echo (MDME) Sequence is a new method which can acquire various contrast-weighted images using quantitative relaxometric parameters measured from multicontrast images. The purpose of our study was to investigate the effect of scan parameters of MDME Sequence on measured T1, T2 values of phantoms at 3.0 T MRI scanner. Gray matter, white matter and cerebrospinal fluid simulation phantoms with different relaxation times (named GM, WM, CSF, respectively) were used in our study. All the phantoms were scanned 9 times on different days using MDME sequence with variations of echo train length, matrix, and acceleration factor. The T1, T2 measurements were acquired after each acquisition. The repeatability was characterized as the intragroup coefficient of variation (CV) of measured values over 9 times, and the discrepancies of measurements across different groups were characterized as intergroup CVs. The highest intragroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 1.36%, 1.75%, 0.74%, 1.41%, 1.70%, 7.79%, respectively. The highest intergroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 0.54%, 1.86%, 1.70%, 0.94%, 1.00%, 2.17%, respectively. Quantitative T1, T2 measurements of gray matter, white matter and cerebrospinal fluid simulation phantoms derived from the MDME sequence were not obviously affected by variations of scanning parameters, such as echo train length, matrix, and acceleration factor on 3T scanner. Springer International Publishing 2022-05-13 2022 /pmc/articles/PMC9239947/ /pubmed/35553390 http://dx.doi.org/10.1007/s13246-022-01128-0 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Scientific Note
Zheng, Zuofeng
Yang, Jiafei
Zhang, Dongpo
Ma, Jun
Yin, Hongxia
Liu, Yawen
Wang, Zhenchang
The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
title The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
title_full The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
title_fullStr The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
title_full_unstemmed The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
title_short The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
title_sort effect of scan parameters on t1, t2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study
topic Scientific Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9239947/
https://www.ncbi.nlm.nih.gov/pubmed/35553390
http://dx.doi.org/10.1007/s13246-022-01128-0
work_keys_str_mv AT zhengzuofeng theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT yangjiafei theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT zhangdongpo theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT majun theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT yinhongxia theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT liuyawen theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT wangzhenchang theeffectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT zhengzuofeng effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT yangjiafei effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT zhangdongpo effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT majun effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT yinhongxia effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT liuyawen effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy
AT wangzhenchang effectofscanparametersont1t2relaxationtimesmeasuredwithmultidynamicmultiechosequenceaphantomstudy