Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures

The aim of this study is to investigate the effectiveness of dose reducing software (ClarityIQ) on patient and staff dose during fluoroscopically guided cardiac procedures. Dose measurements were collected in a room without dose reducing software (n = 157) and compared with similar procedures perfor...

Descripción completa

Detalles Bibliográficos
Autores principales: Wilson-Stewart, Kelly S., Fontanarosa, Davide, Malacova, Eva, Trapp, Jamie V.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9239962/
https://www.ncbi.nlm.nih.gov/pubmed/35532868
http://dx.doi.org/10.1007/s13246-022-01126-2
_version_ 1784737430646030336
author Wilson-Stewart, Kelly S.
Fontanarosa, Davide
Malacova, Eva
Trapp, Jamie V.
author_facet Wilson-Stewart, Kelly S.
Fontanarosa, Davide
Malacova, Eva
Trapp, Jamie V.
author_sort Wilson-Stewart, Kelly S.
collection PubMed
description The aim of this study is to investigate the effectiveness of dose reducing software (ClarityIQ) on patient and staff dose during fluoroscopically guided cardiac procedures. Dose measurements were collected in a room without dose reducing software (n = 157) and compared with similar procedures performed in two rooms with the software (n = 1141). Procedures included diagnostic coronary angiography, percutaneous coronary intervention, deployment of cardiac closure devices (for occlusion of atrial septal defect, patent foramen ovale, and atrial appendage) and insertion of permanent pacemakers. The dose reducing software was found to be effective in reducing patient and staff dose by approximately 50%. This study has added to the limited literature reporting on the capability of dose reducing software to decrease radiation exposure during the implantation of cardiac closure devices, as well as demonstrating a reduction in dose to the cardiologist and nursing staff. Administrators should ensure timely upgrades to angiographic equipment to safeguard patients and staff against the potentially adverse effects of radiation exposure. Regardless of the use of dose reducing software, the mean occupational dose during closure devices was in descending order scout > scrub > cardiologist. Scrub nurse dose was found to be higher than the cardiologist during closure devices (0.98/0.26 μSv) and diagnostic coronary angiograms (1.51/0.82 μSv). Nursing staff should be aware that their levels of radiation dose during some cardiac procedures may come close to or even exceed that of the cardiologist.
format Online
Article
Text
id pubmed-9239962
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-92399622022-06-30 Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures Wilson-Stewart, Kelly S. Fontanarosa, Davide Malacova, Eva Trapp, Jamie V. Phys Eng Sci Med Scientific Paper The aim of this study is to investigate the effectiveness of dose reducing software (ClarityIQ) on patient and staff dose during fluoroscopically guided cardiac procedures. Dose measurements were collected in a room without dose reducing software (n = 157) and compared with similar procedures performed in two rooms with the software (n = 1141). Procedures included diagnostic coronary angiography, percutaneous coronary intervention, deployment of cardiac closure devices (for occlusion of atrial septal defect, patent foramen ovale, and atrial appendage) and insertion of permanent pacemakers. The dose reducing software was found to be effective in reducing patient and staff dose by approximately 50%. This study has added to the limited literature reporting on the capability of dose reducing software to decrease radiation exposure during the implantation of cardiac closure devices, as well as demonstrating a reduction in dose to the cardiologist and nursing staff. Administrators should ensure timely upgrades to angiographic equipment to safeguard patients and staff against the potentially adverse effects of radiation exposure. Regardless of the use of dose reducing software, the mean occupational dose during closure devices was in descending order scout > scrub > cardiologist. Scrub nurse dose was found to be higher than the cardiologist during closure devices (0.98/0.26 μSv) and diagnostic coronary angiograms (1.51/0.82 μSv). Nursing staff should be aware that their levels of radiation dose during some cardiac procedures may come close to or even exceed that of the cardiologist. Springer International Publishing 2022-05-09 2022 /pmc/articles/PMC9239962/ /pubmed/35532868 http://dx.doi.org/10.1007/s13246-022-01126-2 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visithttp://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Scientific Paper
Wilson-Stewart, Kelly S.
Fontanarosa, Davide
Malacova, Eva
Trapp, Jamie V.
Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
title Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
title_full Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
title_fullStr Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
title_full_unstemmed Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
title_short Impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
title_sort impact of dose reducing software on patient and staff temple dose during fluoroscopically guided pacemaker insertion, closure devices implantation and coronary angiography procedures
topic Scientific Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9239962/
https://www.ncbi.nlm.nih.gov/pubmed/35532868
http://dx.doi.org/10.1007/s13246-022-01126-2
work_keys_str_mv AT wilsonstewartkellys impactofdosereducingsoftwareonpatientandstafftempledoseduringfluoroscopicallyguidedpacemakerinsertionclosuredevicesimplantationandcoronaryangiographyprocedures
AT fontanarosadavide impactofdosereducingsoftwareonpatientandstafftempledoseduringfluoroscopicallyguidedpacemakerinsertionclosuredevicesimplantationandcoronaryangiographyprocedures
AT malacovaeva impactofdosereducingsoftwareonpatientandstafftempledoseduringfluoroscopicallyguidedpacemakerinsertionclosuredevicesimplantationandcoronaryangiographyprocedures
AT trappjamiev impactofdosereducingsoftwareonpatientandstafftempledoseduringfluoroscopicallyguidedpacemakerinsertionclosuredevicesimplantationandcoronaryangiographyprocedures