Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report
BACKGROUND: Inadvertent lead malposition (ILM) in the left ventricle (LV) via the subclavian artery is a rare complication during the insertion of cardiac implantable electronic devices (CIED). If not identified, there is a risk of systemic thromboembolism. Transarterial pacing lead extraction often...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9240413/ https://www.ncbi.nlm.nih.gov/pubmed/35775017 http://dx.doi.org/10.1093/ehjcr/ytac234 |
_version_ | 1784737546959323136 |
---|---|
author | Ng, Primero Paul, Vince Shetty, Sharad Lambert, James |
author_facet | Ng, Primero Paul, Vince Shetty, Sharad Lambert, James |
author_sort | Ng, Primero |
collection | PubMed |
description | BACKGROUND: Inadvertent lead malposition (ILM) in the left ventricle (LV) via the subclavian artery is a rare complication during the insertion of cardiac implantable electronic devices (CIED). If not identified, there is a risk of systemic thromboembolism. Transarterial pacing lead extraction often requires surgical removal and carries high risks of bleeding and thromboembolism, but percutaneous extraction has also been previously described. CASE SUMMARY: A 71-year-old female presented with left homonymous hemianopia on Day 1 post-insertion of a dual-chamber permanent pacemaker (PPM). A computed tomography (CT) angiogram of the brain and aortic arch revealed an acute occlusion of a branch of the right posterior circulating artery (PCA) and a malpositioned pacing lead in the left subclavian artery. Urgent percutaneous removal of the transarterial lead using the retained wire technique was successfully performed. DISCUSSION: Inadvertent lead malposition in the arterial system is rare and often requires lead extraction due to systemic thromboembolic complications. The retained wire technique has been previously described for percutaneous transvenous lead extraction and exchange, but to our knowledge, we are the first to report utilizing this technique for transarterial lead extraction. Using a case report, we highlight the utility, safety, and effectiveness of the retained wire technique in extracting a malposition lead in the subclavian artery and LV. |
format | Online Article Text |
id | pubmed-9240413 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-92404132022-06-29 Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report Ng, Primero Paul, Vince Shetty, Sharad Lambert, James Eur Heart J Case Rep Case Report BACKGROUND: Inadvertent lead malposition (ILM) in the left ventricle (LV) via the subclavian artery is a rare complication during the insertion of cardiac implantable electronic devices (CIED). If not identified, there is a risk of systemic thromboembolism. Transarterial pacing lead extraction often requires surgical removal and carries high risks of bleeding and thromboembolism, but percutaneous extraction has also been previously described. CASE SUMMARY: A 71-year-old female presented with left homonymous hemianopia on Day 1 post-insertion of a dual-chamber permanent pacemaker (PPM). A computed tomography (CT) angiogram of the brain and aortic arch revealed an acute occlusion of a branch of the right posterior circulating artery (PCA) and a malpositioned pacing lead in the left subclavian artery. Urgent percutaneous removal of the transarterial lead using the retained wire technique was successfully performed. DISCUSSION: Inadvertent lead malposition in the arterial system is rare and often requires lead extraction due to systemic thromboembolic complications. The retained wire technique has been previously described for percutaneous transvenous lead extraction and exchange, but to our knowledge, we are the first to report utilizing this technique for transarterial lead extraction. Using a case report, we highlight the utility, safety, and effectiveness of the retained wire technique in extracting a malposition lead in the subclavian artery and LV. Oxford University Press 2022-06-10 /pmc/articles/PMC9240413/ /pubmed/35775017 http://dx.doi.org/10.1093/ehjcr/ytac234 Text en © The Author(s) 2022. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | Case Report Ng, Primero Paul, Vince Shetty, Sharad Lambert, James Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
title | Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
title_full | Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
title_fullStr | Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
title_full_unstemmed | Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
title_short | Percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
title_sort | percutaneous extraction of a malpositioned subclavian arterial pacing lead using the retained wire technique and a vascular closure device: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9240413/ https://www.ncbi.nlm.nih.gov/pubmed/35775017 http://dx.doi.org/10.1093/ehjcr/ytac234 |
work_keys_str_mv | AT ngprimero percutaneousextractionofamalpositionedsubclavianarterialpacingleadusingtheretainedwiretechniqueandavascularclosuredeviceacasereport AT paulvince percutaneousextractionofamalpositionedsubclavianarterialpacingleadusingtheretainedwiretechniqueandavascularclosuredeviceacasereport AT shettysharad percutaneousextractionofamalpositionedsubclavianarterialpacingleadusingtheretainedwiretechniqueandavascularclosuredeviceacasereport AT lambertjames percutaneousextractionofamalpositionedsubclavianarterialpacingleadusingtheretainedwiretechniqueandavascularclosuredeviceacasereport |