Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters

BACKGROUND: For wide-necked intracranial aneurysms, endo-saccular flow disruption can be a viable alternative to coiling or flow diverters. The Contour Neurovascular System is an intrasaccular flow diverter device targeting the neck of the aneurysm. Until now, the system had to be delivered through...

Descripción completa

Detalles Bibliográficos
Autores principales: Thormann, Maximilian, Mpotsaris, Anastasios, Behme, Daniel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9244739/
https://www.ncbi.nlm.nih.gov/pubmed/34423661
http://dx.doi.org/10.1177/19714009211041523
_version_ 1784738594503524352
author Thormann, Maximilian
Mpotsaris, Anastasios
Behme, Daniel
author_facet Thormann, Maximilian
Mpotsaris, Anastasios
Behme, Daniel
author_sort Thormann, Maximilian
collection PubMed
description BACKGROUND: For wide-necked intracranial aneurysms, endo-saccular flow disruption can be a viable alternative to coiling or flow diverters. The Contour Neurovascular System is an intrasaccular flow diverter device targeting the neck of the aneurysm. Until now, the system had to be delivered through a 0.027″ microcatheter. We report the first implantation and follow-up of the novel Contour 021 system compatible with 0.021″ microcatheters. Case presentation: A 54-year-old male patient presented with an unruptured right middle cerebral artery aneurysm at the right temporopolar branch. Existing medication included apixaban. An arteriogram showed a broad-based aneurysm. Due to its asymmetric geometry, neither the Woven EndoBridge nor stent-assisted coil embolisation were regarded as promising treatment strategies. To uphold the option of different treatment options, prasugrel 10 mg was initiated before treatment. Implantation was performed under general anaesthesia via femoral artery puncture. A 0.021″ Headway™ catheter was used for accessing the aneurysm. The Contour device was oversized to the equatorial plane. Deployment was successful with only one attempt without the need for re-sheathing. Follow-up catheter angiography was performed after three months, showing complete occlusion of the aneurysm. No procedure-related complications occurred. CONCLUSION: The 0.021 design of the Contour enlarges the subgroup of patients that can be treated with endo-saccular devices and will enable treatment of smaller and more distal aneurysms.
format Online
Article
Text
id pubmed-9244739
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-92447392022-07-01 Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters Thormann, Maximilian Mpotsaris, Anastasios Behme, Daniel Neuroradiol J Case Reports BACKGROUND: For wide-necked intracranial aneurysms, endo-saccular flow disruption can be a viable alternative to coiling or flow diverters. The Contour Neurovascular System is an intrasaccular flow diverter device targeting the neck of the aneurysm. Until now, the system had to be delivered through a 0.027″ microcatheter. We report the first implantation and follow-up of the novel Contour 021 system compatible with 0.021″ microcatheters. Case presentation: A 54-year-old male patient presented with an unruptured right middle cerebral artery aneurysm at the right temporopolar branch. Existing medication included apixaban. An arteriogram showed a broad-based aneurysm. Due to its asymmetric geometry, neither the Woven EndoBridge nor stent-assisted coil embolisation were regarded as promising treatment strategies. To uphold the option of different treatment options, prasugrel 10 mg was initiated before treatment. Implantation was performed under general anaesthesia via femoral artery puncture. A 0.021″ Headway™ catheter was used for accessing the aneurysm. The Contour device was oversized to the equatorial plane. Deployment was successful with only one attempt without the need for re-sheathing. Follow-up catheter angiography was performed after three months, showing complete occlusion of the aneurysm. No procedure-related complications occurred. CONCLUSION: The 0.021 design of the Contour enlarges the subgroup of patients that can be treated with endo-saccular devices and will enable treatment of smaller and more distal aneurysms. SAGE Publications 2021-08-23 2022-06 /pmc/articles/PMC9244739/ /pubmed/34423661 http://dx.doi.org/10.1177/19714009211041523 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/This article is distributed under the terms of the Creative Commons Attribution 4.0 License (https://creativecommons.org/licenses/by/4.0/) which permits any use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Case Reports
Thormann, Maximilian
Mpotsaris, Anastasios
Behme, Daniel
Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters
title Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters
title_full Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters
title_fullStr Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters
title_full_unstemmed Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters
title_short Treatment of a middle cerebral artery bifurcation aneurysm with the novel Contour Neurovascular System compatible with 0.021″ catheters
title_sort treatment of a middle cerebral artery bifurcation aneurysm with the novel contour neurovascular system compatible with 0.021″ catheters
topic Case Reports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9244739/
https://www.ncbi.nlm.nih.gov/pubmed/34423661
http://dx.doi.org/10.1177/19714009211041523
work_keys_str_mv AT thormannmaximilian treatmentofamiddlecerebralarterybifurcationaneurysmwiththenovelcontourneurovascularsystemcompatiblewith0021catheters
AT mpotsarisanastasios treatmentofamiddlecerebralarterybifurcationaneurysmwiththenovelcontourneurovascularsystemcompatiblewith0021catheters
AT behmedaniel treatmentofamiddlecerebralarterybifurcationaneurysmwiththenovelcontourneurovascularsystemcompatiblewith0021catheters