NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT

PURPOSE: The DMI PET/CT is a modular silicon photomultiplier–based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an addition...

Descripción completa

Detalles Bibliográficos
Autores principales: Zeimpekis, Konstantinos G., Kotasidis, Fotis A., Huellner, Martin, Nemirovsky, Alexandra, Kaufmann, Philipp A., Treyer, Valerie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9250480/
https://www.ncbi.nlm.nih.gov/pubmed/35284970
http://dx.doi.org/10.1007/s00259-022-05751-7
_version_ 1784739821861732352
author Zeimpekis, Konstantinos G.
Kotasidis, Fotis A.
Huellner, Martin
Nemirovsky, Alexandra
Kaufmann, Philipp A.
Treyer, Valerie
author_facet Zeimpekis, Konstantinos G.
Kotasidis, Fotis A.
Huellner, Martin
Nemirovsky, Alexandra
Kaufmann, Philipp A.
Treyer, Valerie
author_sort Zeimpekis, Konstantinos G.
collection PubMed
description PURPOSE: The DMI PET/CT is a modular silicon photomultiplier–based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade. METHODS: PET performance was evaluated using the NEMA NU 2–2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality. RESULTS: The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3–8.5% for 5 rings vs 1.9–6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions. CONCLUSIONS: The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels.
format Online
Article
Text
id pubmed-9250480
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-92504802022-07-04 NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT Zeimpekis, Konstantinos G. Kotasidis, Fotis A. Huellner, Martin Nemirovsky, Alexandra Kaufmann, Philipp A. Treyer, Valerie Eur J Nucl Med Mol Imaging Original Article PURPOSE: The DMI PET/CT is a modular silicon photomultiplier–based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade. METHODS: PET performance was evaluated using the NEMA NU 2–2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality. RESULTS: The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3–8.5% for 5 rings vs 1.9–6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions. CONCLUSIONS: The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels. Springer Berlin Heidelberg 2022-03-14 2022 /pmc/articles/PMC9250480/ /pubmed/35284970 http://dx.doi.org/10.1007/s00259-022-05751-7 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Zeimpekis, Konstantinos G.
Kotasidis, Fotis A.
Huellner, Martin
Nemirovsky, Alexandra
Kaufmann, Philipp A.
Treyer, Valerie
NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
title NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
title_full NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
title_fullStr NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
title_full_unstemmed NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
title_short NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT
title_sort nema nu 2–2018 performance evaluation of a new generation 30-cm axial field-of-view discovery mi pet/ct
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9250480/
https://www.ncbi.nlm.nih.gov/pubmed/35284970
http://dx.doi.org/10.1007/s00259-022-05751-7
work_keys_str_mv AT zeimpekiskonstantinosg nemanu22018performanceevaluationofanewgeneration30cmaxialfieldofviewdiscoverymipetct
AT kotasidisfotisa nemanu22018performanceevaluationofanewgeneration30cmaxialfieldofviewdiscoverymipetct
AT huellnermartin nemanu22018performanceevaluationofanewgeneration30cmaxialfieldofviewdiscoverymipetct
AT nemirovskyalexandra nemanu22018performanceevaluationofanewgeneration30cmaxialfieldofviewdiscoverymipetct
AT kaufmannphilippa nemanu22018performanceevaluationofanewgeneration30cmaxialfieldofviewdiscoverymipetct
AT treyervalerie nemanu22018performanceevaluationofanewgeneration30cmaxialfieldofviewdiscoverymipetct