Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT

BACKGROUND: To newly describe the vault measurement by using a widely used swept-source OCT-based optical biometer (IOLMaster700) and accessd the accuracy of vault measurement. METHODS: This was a retrospective, cross-sectional study. All patients underwent implantable Collamer lens (ICL) implantati...

Descripción completa

Detalles Bibliográficos
Autores principales: Du, Yali, Jin, Chuang, Yin, Shengjie, Wang, Geng, Ma, Qian, Li, Yuancun, Chen, Binyao, Wang, Hongxi, Qiu, Kunliang, Zhang, Mingzhi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9259877/
https://www.ncbi.nlm.nih.gov/pubmed/35814765
http://dx.doi.org/10.3389/fmed.2022.865719
_version_ 1784741883960885248
author Du, Yali
Jin, Chuang
Yin, Shengjie
Wang, Geng
Ma, Qian
Li, Yuancun
Chen, Binyao
Wang, Hongxi
Qiu, Kunliang
Zhang, Mingzhi
author_facet Du, Yali
Jin, Chuang
Yin, Shengjie
Wang, Geng
Ma, Qian
Li, Yuancun
Chen, Binyao
Wang, Hongxi
Qiu, Kunliang
Zhang, Mingzhi
author_sort Du, Yali
collection PubMed
description BACKGROUND: To newly describe the vault measurement by using a widely used swept-source OCT-based optical biometer (IOLMaster700) and accessd the accuracy of vault measurement. METHODS: This was a retrospective, cross-sectional study. All patients underwent implantable Collamer lens (ICL) implantation surgery without complications. IOLMaster700 and AS-OCT analyses were conducted for each eye on the same day in the same condition. Measurements of anterior chamber depth (ACD), corneal-ICL (C-ICL), and vault values were made and recorded. The repeatability of the IOL Master700 measurements was quantified based upon intraclass correlation coefficient (ICC) values. Correlations between IOL Master700 and AS-OCT measurements made with these different analytical approaches were assessed. The agreement of instruments was evaluated using Bland-Altman plots. RESULTS: The IOLMaster700 instrument yielded highly reliable measurements of vault, C-ICL, and ACD (ICC = 0.996, 0.995, 0.995, respectively). Vault, C-ICL and ACD values as measured using the IOLMaster700, was slightly smaller than that measured via AS-OCT, but these differences were not significant (p = 0.652, p = 0.121 and p = 0.091, respectively). The vault, C-ICL, and ACD measurements by these two instruments were strongly correlated (r = 0.971, r = 0.944, and r = 0.963, respectively; all p < 0.001). The 95% limits of agreement for vault, C-ICL, and ACD measurements between the two devices were−0.08 to 0.08 mm,−0.14 to 0.11 mm, and−0.13 to 0.10 mm, respectively. CONCLUSIONS: The IOLMasrer700 can measure implanted ICL vault with a high degree of accuracy and repeatability. Good correlations and agreement were observed between IOLMaster700 and AS-OCT in measuring vault, C-ICL, and ACD measurements.
format Online
Article
Text
id pubmed-9259877
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-92598772022-07-08 Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT Du, Yali Jin, Chuang Yin, Shengjie Wang, Geng Ma, Qian Li, Yuancun Chen, Binyao Wang, Hongxi Qiu, Kunliang Zhang, Mingzhi Front Med (Lausanne) Medicine BACKGROUND: To newly describe the vault measurement by using a widely used swept-source OCT-based optical biometer (IOLMaster700) and accessd the accuracy of vault measurement. METHODS: This was a retrospective, cross-sectional study. All patients underwent implantable Collamer lens (ICL) implantation surgery without complications. IOLMaster700 and AS-OCT analyses were conducted for each eye on the same day in the same condition. Measurements of anterior chamber depth (ACD), corneal-ICL (C-ICL), and vault values were made and recorded. The repeatability of the IOL Master700 measurements was quantified based upon intraclass correlation coefficient (ICC) values. Correlations between IOL Master700 and AS-OCT measurements made with these different analytical approaches were assessed. The agreement of instruments was evaluated using Bland-Altman plots. RESULTS: The IOLMaster700 instrument yielded highly reliable measurements of vault, C-ICL, and ACD (ICC = 0.996, 0.995, 0.995, respectively). Vault, C-ICL and ACD values as measured using the IOLMaster700, was slightly smaller than that measured via AS-OCT, but these differences were not significant (p = 0.652, p = 0.121 and p = 0.091, respectively). The vault, C-ICL, and ACD measurements by these two instruments were strongly correlated (r = 0.971, r = 0.944, and r = 0.963, respectively; all p < 0.001). The 95% limits of agreement for vault, C-ICL, and ACD measurements between the two devices were−0.08 to 0.08 mm,−0.14 to 0.11 mm, and−0.13 to 0.10 mm, respectively. CONCLUSIONS: The IOLMasrer700 can measure implanted ICL vault with a high degree of accuracy and repeatability. Good correlations and agreement were observed between IOLMaster700 and AS-OCT in measuring vault, C-ICL, and ACD measurements. Frontiers Media S.A. 2022-06-23 /pmc/articles/PMC9259877/ /pubmed/35814765 http://dx.doi.org/10.3389/fmed.2022.865719 Text en Copyright © 2022 Du, Jin, Yin, Wang, Ma, Li, Chen, Wang, Qiu and Zhang. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Medicine
Du, Yali
Jin, Chuang
Yin, Shengjie
Wang, Geng
Ma, Qian
Li, Yuancun
Chen, Binyao
Wang, Hongxi
Qiu, Kunliang
Zhang, Mingzhi
Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
title Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
title_full Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
title_fullStr Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
title_full_unstemmed Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
title_short Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
title_sort comparison of vault measurements using a swept-source oct-based optical biometer and anterior segment oct
topic Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9259877/
https://www.ncbi.nlm.nih.gov/pubmed/35814765
http://dx.doi.org/10.3389/fmed.2022.865719
work_keys_str_mv AT duyali comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT jinchuang comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT yinshengjie comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT wanggeng comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT maqian comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT liyuancun comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT chenbinyao comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT wanghongxi comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT qiukunliang comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct
AT zhangmingzhi comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct