Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT
BACKGROUND: To newly describe the vault measurement by using a widely used swept-source OCT-based optical biometer (IOLMaster700) and accessd the accuracy of vault measurement. METHODS: This was a retrospective, cross-sectional study. All patients underwent implantable Collamer lens (ICL) implantati...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9259877/ https://www.ncbi.nlm.nih.gov/pubmed/35814765 http://dx.doi.org/10.3389/fmed.2022.865719 |
_version_ | 1784741883960885248 |
---|---|
author | Du, Yali Jin, Chuang Yin, Shengjie Wang, Geng Ma, Qian Li, Yuancun Chen, Binyao Wang, Hongxi Qiu, Kunliang Zhang, Mingzhi |
author_facet | Du, Yali Jin, Chuang Yin, Shengjie Wang, Geng Ma, Qian Li, Yuancun Chen, Binyao Wang, Hongxi Qiu, Kunliang Zhang, Mingzhi |
author_sort | Du, Yali |
collection | PubMed |
description | BACKGROUND: To newly describe the vault measurement by using a widely used swept-source OCT-based optical biometer (IOLMaster700) and accessd the accuracy of vault measurement. METHODS: This was a retrospective, cross-sectional study. All patients underwent implantable Collamer lens (ICL) implantation surgery without complications. IOLMaster700 and AS-OCT analyses were conducted for each eye on the same day in the same condition. Measurements of anterior chamber depth (ACD), corneal-ICL (C-ICL), and vault values were made and recorded. The repeatability of the IOL Master700 measurements was quantified based upon intraclass correlation coefficient (ICC) values. Correlations between IOL Master700 and AS-OCT measurements made with these different analytical approaches were assessed. The agreement of instruments was evaluated using Bland-Altman plots. RESULTS: The IOLMaster700 instrument yielded highly reliable measurements of vault, C-ICL, and ACD (ICC = 0.996, 0.995, 0.995, respectively). Vault, C-ICL and ACD values as measured using the IOLMaster700, was slightly smaller than that measured via AS-OCT, but these differences were not significant (p = 0.652, p = 0.121 and p = 0.091, respectively). The vault, C-ICL, and ACD measurements by these two instruments were strongly correlated (r = 0.971, r = 0.944, and r = 0.963, respectively; all p < 0.001). The 95% limits of agreement for vault, C-ICL, and ACD measurements between the two devices were−0.08 to 0.08 mm,−0.14 to 0.11 mm, and−0.13 to 0.10 mm, respectively. CONCLUSIONS: The IOLMasrer700 can measure implanted ICL vault with a high degree of accuracy and repeatability. Good correlations and agreement were observed between IOLMaster700 and AS-OCT in measuring vault, C-ICL, and ACD measurements. |
format | Online Article Text |
id | pubmed-9259877 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-92598772022-07-08 Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT Du, Yali Jin, Chuang Yin, Shengjie Wang, Geng Ma, Qian Li, Yuancun Chen, Binyao Wang, Hongxi Qiu, Kunliang Zhang, Mingzhi Front Med (Lausanne) Medicine BACKGROUND: To newly describe the vault measurement by using a widely used swept-source OCT-based optical biometer (IOLMaster700) and accessd the accuracy of vault measurement. METHODS: This was a retrospective, cross-sectional study. All patients underwent implantable Collamer lens (ICL) implantation surgery without complications. IOLMaster700 and AS-OCT analyses were conducted for each eye on the same day in the same condition. Measurements of anterior chamber depth (ACD), corneal-ICL (C-ICL), and vault values were made and recorded. The repeatability of the IOL Master700 measurements was quantified based upon intraclass correlation coefficient (ICC) values. Correlations between IOL Master700 and AS-OCT measurements made with these different analytical approaches were assessed. The agreement of instruments was evaluated using Bland-Altman plots. RESULTS: The IOLMaster700 instrument yielded highly reliable measurements of vault, C-ICL, and ACD (ICC = 0.996, 0.995, 0.995, respectively). Vault, C-ICL and ACD values as measured using the IOLMaster700, was slightly smaller than that measured via AS-OCT, but these differences were not significant (p = 0.652, p = 0.121 and p = 0.091, respectively). The vault, C-ICL, and ACD measurements by these two instruments were strongly correlated (r = 0.971, r = 0.944, and r = 0.963, respectively; all p < 0.001). The 95% limits of agreement for vault, C-ICL, and ACD measurements between the two devices were−0.08 to 0.08 mm,−0.14 to 0.11 mm, and−0.13 to 0.10 mm, respectively. CONCLUSIONS: The IOLMasrer700 can measure implanted ICL vault with a high degree of accuracy and repeatability. Good correlations and agreement were observed between IOLMaster700 and AS-OCT in measuring vault, C-ICL, and ACD measurements. Frontiers Media S.A. 2022-06-23 /pmc/articles/PMC9259877/ /pubmed/35814765 http://dx.doi.org/10.3389/fmed.2022.865719 Text en Copyright © 2022 Du, Jin, Yin, Wang, Ma, Li, Chen, Wang, Qiu and Zhang. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Medicine Du, Yali Jin, Chuang Yin, Shengjie Wang, Geng Ma, Qian Li, Yuancun Chen, Binyao Wang, Hongxi Qiu, Kunliang Zhang, Mingzhi Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT |
title | Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT |
title_full | Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT |
title_fullStr | Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT |
title_full_unstemmed | Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT |
title_short | Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT |
title_sort | comparison of vault measurements using a swept-source oct-based optical biometer and anterior segment oct |
topic | Medicine |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9259877/ https://www.ncbi.nlm.nih.gov/pubmed/35814765 http://dx.doi.org/10.3389/fmed.2022.865719 |
work_keys_str_mv | AT duyali comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT jinchuang comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT yinshengjie comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT wanggeng comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT maqian comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT liyuancun comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT chenbinyao comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT wanghongxi comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT qiukunliang comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct AT zhangmingzhi comparisonofvaultmeasurementsusingasweptsourceoctbasedopticalbiometerandanteriorsegmentoct |