An autopsy case of retrograde in situ branched stent grafting for a complex aortic arch aneurysm

Although retrograde in situ branched stent grafting (RIBS) is one possible treatment option for thoracic aortic aneurysms, concerns exist regarding the durability of the stent graft (SG) at the junction between the main SG and the branched SG. We report on the autopsy results of a patient treated wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Shukuzawa, Kota, Baba, Takeshi, Nishie, Ryosuke, Ozawa, Hirotsugu, Omori, Makiko, Hara, Masayuki, Tachihara, Hiromasa, Ohki, Takao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9260453/
https://www.ncbi.nlm.nih.gov/pubmed/35812121
http://dx.doi.org/10.1016/j.jvscit.2022.04.007
Descripción
Sumario:Although retrograde in situ branched stent grafting (RIBS) is one possible treatment option for thoracic aortic aneurysms, concerns exist regarding the durability of the stent graft (SG) at the junction between the main SG and the branched SG. We report on the autopsy results of a patient treated with RIBS for a complex aortic arch aneurysm. The patient had died of a nonaortic cause 14 months after the RIBS procedure. On computed tomography analysis and a leak test, the harvested SG was intact without any stent fracture, stenosis, or junctional leak at 1 atm water pressure (760 mm Hg).