Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram
Background: Holter electrocardiogram (ECG) is the gold standard for ambulatory monitoring of atrial fibrillation (AF) but it is insufficient because of its limited recording time. Although several consumer ECG devices provide longer recording time, they generally do not undergo the regulatory proces...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
F1000 Research Limited
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9263574/ https://www.ncbi.nlm.nih.gov/pubmed/35860478 http://dx.doi.org/10.12688/f1000research.75712.2 |
_version_ | 1784742764594855936 |
---|---|
author | Machino, Takeshi Aonuma, Kazutaka Komatsu, Yuki Yamasaki, Hiro Igarashi, Miyako Nogami, Akihiko Ieda, Masaki |
author_facet | Machino, Takeshi Aonuma, Kazutaka Komatsu, Yuki Yamasaki, Hiro Igarashi, Miyako Nogami, Akihiko Ieda, Masaki |
author_sort | Machino, Takeshi |
collection | PubMed |
description | Background: Holter electrocardiogram (ECG) is the gold standard for ambulatory monitoring of atrial fibrillation (AF) but it is insufficient because of its limited recording time. Although several consumer ECG devices provide longer recording time, they generally do not undergo the regulatory process for medical use. Furthermore, current medical-grade devices for longer ECG monitoring are not continuous or too invasive for AF monitoring. A wearable ECG with a medical-grade dry textile electrode is a promising technology to remedy this limitation. This pilot study aimed to simultaneously compare the wearable and Holter ECGs for ambulatory monitoring in a clinical setting. Methods: This prospective observational study enrolled 18 patients who underwent AF ablation. One day after AF ablation, ambulatory ECG was obtained for three hours simultaneously using both the wearable and Holter ECG devices. Automatic ECG interpretations between devices were compared with correlation and agreement analyses. Results: Simultaneous ECG monitoring demonstrated a comparable analysis time and total heart beats between the two devices. Almost complete correlation and agreement were also demonstrated in all clinically relevant testing aspects except in R-wave amplitude (r = 0.743, p < .001). AF was detected in three patients. AF duration was the same in both ECG devices in two patients with continuous AF. In the remaining patient with intermittent AF, AF duration was shortened by 0.6% with the wearable ECG as compared to that with the Holter ECG. Conclusions: Simultaneous ECG comparison revealed a high consistency between the wearable and Holter ECG devices. The results of this study warrant further clinical studies for long-term monitoring of ambulatory ECG after AF ablation. |
format | Online Article Text |
id | pubmed-9263574 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | F1000 Research Limited |
record_format | MEDLINE/PubMed |
spelling | pubmed-92635742022-07-19 Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram Machino, Takeshi Aonuma, Kazutaka Komatsu, Yuki Yamasaki, Hiro Igarashi, Miyako Nogami, Akihiko Ieda, Masaki F1000Res Research Article Background: Holter electrocardiogram (ECG) is the gold standard for ambulatory monitoring of atrial fibrillation (AF) but it is insufficient because of its limited recording time. Although several consumer ECG devices provide longer recording time, they generally do not undergo the regulatory process for medical use. Furthermore, current medical-grade devices for longer ECG monitoring are not continuous or too invasive for AF monitoring. A wearable ECG with a medical-grade dry textile electrode is a promising technology to remedy this limitation. This pilot study aimed to simultaneously compare the wearable and Holter ECGs for ambulatory monitoring in a clinical setting. Methods: This prospective observational study enrolled 18 patients who underwent AF ablation. One day after AF ablation, ambulatory ECG was obtained for three hours simultaneously using both the wearable and Holter ECG devices. Automatic ECG interpretations between devices were compared with correlation and agreement analyses. Results: Simultaneous ECG monitoring demonstrated a comparable analysis time and total heart beats between the two devices. Almost complete correlation and agreement were also demonstrated in all clinically relevant testing aspects except in R-wave amplitude (r = 0.743, p < .001). AF was detected in three patients. AF duration was the same in both ECG devices in two patients with continuous AF. In the remaining patient with intermittent AF, AF duration was shortened by 0.6% with the wearable ECG as compared to that with the Holter ECG. Conclusions: Simultaneous ECG comparison revealed a high consistency between the wearable and Holter ECG devices. The results of this study warrant further clinical studies for long-term monitoring of ambulatory ECG after AF ablation. F1000 Research Limited 2022-06-30 /pmc/articles/PMC9263574/ /pubmed/35860478 http://dx.doi.org/10.12688/f1000research.75712.2 Text en Copyright: © 2022 Machino T et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Machino, Takeshi Aonuma, Kazutaka Komatsu, Yuki Yamasaki, Hiro Igarashi, Miyako Nogami, Akihiko Ieda, Masaki Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram |
title | Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram |
title_full | Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram |
title_fullStr | Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram |
title_full_unstemmed | Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram |
title_short | Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram |
title_sort | dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: a pilot study of simultaneous comparison to the holter electrocardiogram |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9263574/ https://www.ncbi.nlm.nih.gov/pubmed/35860478 http://dx.doi.org/10.12688/f1000research.75712.2 |
work_keys_str_mv | AT machinotakeshi drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram AT aonumakazutaka drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram AT komatsuyuki drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram AT yamasakihiro drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram AT igarashimiyako drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram AT nogamiakihiko drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram AT iedamasaki drytextileelectrodeforambulatorymonitoringaftercatheterablationofatrialfibrillationapilotstudyofsimultaneouscomparisontotheholterelectrocardiogram |