Electric-Field-Based Guidance for Percutaneous Catheter Vessel Crossing
Percutaneous procedures to divert blood flow from one blood vessel to another can be performed with intravascular catheters but demand a method to align a crossing needle from one vessel to another. Fluoroscopic imaging alone is not adequate, and it is preferable to have a sensor on one catheter tha...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9269779/ https://www.ncbi.nlm.nih.gov/pubmed/35808416 http://dx.doi.org/10.3390/s22134928 |
Sumario: | Percutaneous procedures to divert blood flow from one blood vessel to another can be performed with intravascular catheters but demand a method to align a crossing needle from one vessel to another. Fluoroscopic imaging alone is not adequate, and it is preferable to have a sensor on one catheter that detects the correct alignment of an incoming needle. This can be implemented by generating dipole electric fields from the crossing catheter which are detected by a receiving catheter in the target vessel and, thus, can calculate and display the degree of alignment, permitting the operator to rotate the crossing catheter to guarantee alignment when deploying a crossing needle. Catheters were built using this concept and evaluated in vitro. The results show that accurate alignment is achieved, and a successful crossing can be made. The concept is being further developed for further clinical evaluation. |
---|