Cargando…
Effectiveness of Anchoring with Balloon Guide Catheter and Stent Retriever in Difficult Mechanical Thrombectomy for Large Vessel Occlusion
OBJECTIVE: A distal navigation of a large bore aspiration catheter during mechanical thrombectomy (MT) is important. However, delivering a large bore aspiration catheter is difficult to a tortuous or atherosclerotic artery. We report the experience of anchoring with balloon guide catheter (BGC) and...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Neurosurgical Society
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9271811/ https://www.ncbi.nlm.nih.gov/pubmed/35728980 http://dx.doi.org/10.3340/jkns.2021.0158 |
Sumario: | OBJECTIVE: A distal navigation of a large bore aspiration catheter during mechanical thrombectomy (MT) is important. However, delivering a large bore aspiration catheter is difficult to a tortuous or atherosclerotic artery. We report the experience of anchoring with balloon guide catheter (BGC) and stent retriever to facilitate the passage of an aspiration catheter in MT. METHODS: When navigating an aspiration catheter failed with a conventional co-axial microcatheter delivery, an anchoring technique was used. Two types of anchoring technique were applied to facilitate distal navigation of a large bore aspiration catheter during MT. First, a passage of aspiration catheter was attempted with a proximal BGC anchoring technique. If this technique also failed, another anchoring technique with distal stent retriever was tried. Consecutive patients who underwent MT with an anchoring technique were identified. Details of procedure, radiologic outcomes, and safety variables were evaluated. RESULTS: A total of 67 patients underwent MT with an anchoring technique. Initial trial of aspiration catheter passage with proximal BGC anchoring technique was successful for 35 patients (52.2%) and the second trial with distal stent retriever anchoring was successful for 32 patients (47.8%). Overall, navigation of a large bore aspiration catheter was successful for all patients (100%) without any procedure related complications. CONCLUSION: Our study showed the usefulness of anchoring technique with proximal BGC and distal stent retriever during MT, especially in those with an unfavorable anatomical structure. This technique could be an alternative option for delivering an of aspiration catheter to a distal location. |
---|