Influence of Sex, BMI, and Skin Color on the Accuracy of Non-Invasive Cuffless Photoplethysmography-Based Blood Pressure Measurements

Vital signs obtained by photoplethysmography-based devices might be influenced by subcutaneous fat and skin color. This observational comparison study aimed to test the accuracy of blood pressure (BP) measurements between a photoplethysmography-based device and cuff-based BP device in ambulatory ind...

Descripción completa

Detalles Bibliográficos
Autores principales: Nachman, Dean, Eisenkraft, Arik, Goldstein, Nir, Ben-Ishay, Arik, Fons, Meir, Merin, Roei, Gepner, Yftach
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9277111/
https://www.ncbi.nlm.nih.gov/pubmed/35846008
http://dx.doi.org/10.3389/fphys.2022.911544
Descripción
Sumario:Vital signs obtained by photoplethysmography-based devices might be influenced by subcutaneous fat and skin color. This observational comparison study aimed to test the accuracy of blood pressure (BP) measurements between a photoplethysmography-based device and cuff-based BP device in ambulatory individuals, coming for a routine BP checkup. Systolic BP (SBP) and diastolic BP (DBP) measurements were stratified based on sex, BMI (<25; 25 ≤BMI<30; 30 ≤kg/m(2)), and skin color (types 1–3 and 4–6 by the Fitzpatrick scale). A total of 1548 measurements were analyzed. Correlations of SBP and DBP between the devices among males/females were between 0.914–0.987 (p < 0.001), and Bland-Altman analysis showed a bias of less than 0.5 mmHg for both sexes. Correlations of SBP and DBP between the devices among BMI groups were between 0.931–0.991 (p < 0.001), and Bland-Altman analysis showed a bias of less than 1 mmHg for all. Correlations of SBP and DBP between the devices among the skin color groups were between 0.936–0.983 (p < 0.001), and Bland-Altman analysis showed a bias of less than 1 mmHg for all. This study shows similar and high agreements between BP measurements obtained using a PPG-based non-invasive cuffless BP device and a cuff-based BP device across sex, BMI, and skin color groups.