Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter pla...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9278680/ https://www.ncbi.nlm.nih.gov/pubmed/35666629 http://dx.doi.org/10.1002/acm2.13661 |
_version_ | 1784746234675724288 |
---|---|
author | Goodall, Simon K. Norvill, Craig |
author_facet | Goodall, Simon K. Norvill, Craig |
author_sort | Goodall, Simon K. |
collection | PubMed |
description | For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter plane. This is achieved by applying a pixel scaling factor (PSF). During this investigation, the dependence of the PSF at cardinal gantry angles was determined along with the resulting effects on the multi‐leaf collimator (MLC) quality assurance (QA) results for three linear accelerators (linacs). The PSF was found to vary by 0.0018–0.0022 mm/pixel during gantry rotation, which resulted in variations in the mean MLC reported error of up to 0.8 mm at 100 mm off‐axis with the gantry rotated to 180°. Measurement and application of a gantry angle–specific PSF is a simple process that can be implemented to improve the accuracy of EPID‐based MLC QA at cardinal gantry angles. |
format | Online Article Text |
id | pubmed-9278680 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-92786802022-07-15 Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance Goodall, Simon K. Norvill, Craig J Appl Clin Med Phys Technical Notes For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter plane. This is achieved by applying a pixel scaling factor (PSF). During this investigation, the dependence of the PSF at cardinal gantry angles was determined along with the resulting effects on the multi‐leaf collimator (MLC) quality assurance (QA) results for three linear accelerators (linacs). The PSF was found to vary by 0.0018–0.0022 mm/pixel during gantry rotation, which resulted in variations in the mean MLC reported error of up to 0.8 mm at 100 mm off‐axis with the gantry rotated to 180°. Measurement and application of a gantry angle–specific PSF is a simple process that can be implemented to improve the accuracy of EPID‐based MLC QA at cardinal gantry angles. John Wiley and Sons Inc. 2022-06-06 /pmc/articles/PMC9278680/ /pubmed/35666629 http://dx.doi.org/10.1002/acm2.13661 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Notes Goodall, Simon K. Norvill, Craig Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
title | Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
title_full | Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
title_fullStr | Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
title_full_unstemmed | Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
title_short | Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
title_sort | variation in elekta iview electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance |
topic | Technical Notes |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9278680/ https://www.ncbi.nlm.nih.gov/pubmed/35666629 http://dx.doi.org/10.1002/acm2.13661 |
work_keys_str_mv | AT goodallsimonk variationinelektaiviewelectronicportalimagerpixelscalefactorwithgantryangleandimpactonmultileafcollimatorqualityassurance AT norvillcraig variationinelektaiviewelectronicportalimagerpixelscalefactorwithgantryangleandimpactonmultileafcollimatorqualityassurance |