Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance

For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter pla...

Descripción completa

Detalles Bibliográficos
Autores principales: Goodall, Simon K., Norvill, Craig
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9278680/
https://www.ncbi.nlm.nih.gov/pubmed/35666629
http://dx.doi.org/10.1002/acm2.13661
_version_ 1784746234675724288
author Goodall, Simon K.
Norvill, Craig
author_facet Goodall, Simon K.
Norvill, Craig
author_sort Goodall, Simon K.
collection PubMed
description For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter plane. This is achieved by applying a pixel scaling factor (PSF). During this investigation, the dependence of the PSF at cardinal gantry angles was determined along with the resulting effects on the multi‐leaf collimator (MLC) quality assurance (QA) results for three linear accelerators (linacs). The PSF was found to vary by 0.0018–0.0022 mm/pixel during gantry rotation, which resulted in variations in the mean MLC reported error of up to 0.8 mm at 100 mm off‐axis with the gantry rotated to 180°. Measurement and application of a gantry angle–specific PSF is a simple process that can be implemented to improve the accuracy of EPID‐based MLC QA at cardinal gantry angles.
format Online
Article
Text
id pubmed-9278680
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-92786802022-07-15 Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance Goodall, Simon K. Norvill, Craig J Appl Clin Med Phys Technical Notes For Elekta Agility linear accelerators, the iViewGT electronic portal imaging device (EPID) is positioned at a nominal X‐Ray source‐to‐panel distance of 1600 mm. For display, image registration, and data processing purposes, the image pixels are scaled to spatial units at the treatment isocenter plane. This is achieved by applying a pixel scaling factor (PSF). During this investigation, the dependence of the PSF at cardinal gantry angles was determined along with the resulting effects on the multi‐leaf collimator (MLC) quality assurance (QA) results for three linear accelerators (linacs). The PSF was found to vary by 0.0018–0.0022 mm/pixel during gantry rotation, which resulted in variations in the mean MLC reported error of up to 0.8 mm at 100 mm off‐axis with the gantry rotated to 180°. Measurement and application of a gantry angle–specific PSF is a simple process that can be implemented to improve the accuracy of EPID‐based MLC QA at cardinal gantry angles. John Wiley and Sons Inc. 2022-06-06 /pmc/articles/PMC9278680/ /pubmed/35666629 http://dx.doi.org/10.1002/acm2.13661 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Goodall, Simon K.
Norvill, Craig
Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
title Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
title_full Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
title_fullStr Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
title_full_unstemmed Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
title_short Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
title_sort variation in elekta iview electronic portal imager pixel scale factor with gantry angle, and impact on multi‐leaf collimator quality assurance
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9278680/
https://www.ncbi.nlm.nih.gov/pubmed/35666629
http://dx.doi.org/10.1002/acm2.13661
work_keys_str_mv AT goodallsimonk variationinelektaiviewelectronicportalimagerpixelscalefactorwithgantryangleandimpactonmultileafcollimatorqualityassurance
AT norvillcraig variationinelektaiviewelectronicportalimagerpixelscalefactorwithgantryangleandimpactonmultileafcollimatorqualityassurance