Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus

BACKGROUND: The accuracy of low-value platelet (PLT) testing is a key reference for clinical decision-making regarding PLT transfusion or surgery. This study aimed to evaluate the capability to detect low-value PLT of the 8 times optical platelet counting (PLT-8X) mode of the BC-6800Plus auto hemato...

Descripción completa

Detalles Bibliográficos
Autores principales: Shen, Yimin, Liu, Dongmei, Wang, Yong, Cao, Jun, Zhang, Shuaishuai, Wen, Hui, Dong, Qiuqiu, Zheng, Dong, Qiu, Jun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: AME Publishing Company 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9279826/
https://www.ncbi.nlm.nih.gov/pubmed/35845540
http://dx.doi.org/10.21037/atm-22-1501
_version_ 1784746488624054272
author Shen, Yimin
Liu, Dongmei
Wang, Yong
Cao, Jun
Zhang, Shuaishuai
Wen, Hui
Dong, Qiuqiu
Zheng, Dong
Qiu, Jun
author_facet Shen, Yimin
Liu, Dongmei
Wang, Yong
Cao, Jun
Zhang, Shuaishuai
Wen, Hui
Dong, Qiuqiu
Zheng, Dong
Qiu, Jun
author_sort Shen, Yimin
collection PubMed
description BACKGROUND: The accuracy of low-value platelet (PLT) testing is a key reference for clinical decision-making regarding PLT transfusion or surgery. This study aimed to evaluate the capability to detect low-value PLT of the 8 times optical platelet counting (PLT-8X) mode of the BC-6800Plus auto hematology analyzer. METHODS: Totally 40 fresh anticoagulated venous whole blood samples with PLT ≤50×10(9)/L were collected from the clinical laboratory at the First Affiliated Hospital of Soochow University to evaluate the precision of low-value PLT in PLT-8X mode. Moreover, 67 samples with PLT <100×10(9)/L were collected, and a methodological comparison was conducted between the results of PLT-8X and those obtained by the reference method [red blood cell (RBC)/PLT ratio method] recommended by the International Committee for Standardization of Hematology (ICSH). In addition, fresh whole blood and PLT-free plasma were used to prepare samples at a series of concentrations within the low-value ranges to evaluate the linearity of PLT and finally investigate the limit of blank (LoB) and limit of detection (LoD) for low-value PLT in PLT-8X mode. RESULTS: The precision low-value PLT in PLT-8X mode was significantly superior to that in PLT-I (P<0.0001); and the correlation coefficient of PLT-8X mode compared with the reference method of flow cytometer was 0.992. The expected bias at the low-value medical decision levels (5×10(9)/L, 10×10(9)/L and 20×10(9)/L) of PLT fell within the range of ±1.0×10(9)/L, and that at 50×10(9)/L fell within the range of ±1.5×10(9)/L. In addition, the correlation coefficient of PLT-8X in the low-value ranges (0 to 30×10(9)/L and 0 to 100×10(9)/L) of PLT was greater than 0.99, and its nonlinear coefficients were not significantly different from 0 (P>0.05). Finally, the LoB and LoD of PLT-8X mode were 0.33×10(9)/L and 0.89×10(9)/L, respectively. CONCLUSIONS: The PLT-8X mode of the BC-6800Plus auto hematology analyzer has extremely low detection sensitivity for low-value PLTs in anticoagulated venous whole blood samples. With good linearity, its precision and accuracy can sufficiently meet the needs for hematology laboratory use, and it can effectively help clinicians to accurately diagnose and treat patients with thrombopenia-related diseases.
format Online
Article
Text
id pubmed-9279826
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher AME Publishing Company
record_format MEDLINE/PubMed
spelling pubmed-92798262022-07-15 Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus Shen, Yimin Liu, Dongmei Wang, Yong Cao, Jun Zhang, Shuaishuai Wen, Hui Dong, Qiuqiu Zheng, Dong Qiu, Jun Ann Transl Med Original Article BACKGROUND: The accuracy of low-value platelet (PLT) testing is a key reference for clinical decision-making regarding PLT transfusion or surgery. This study aimed to evaluate the capability to detect low-value PLT of the 8 times optical platelet counting (PLT-8X) mode of the BC-6800Plus auto hematology analyzer. METHODS: Totally 40 fresh anticoagulated venous whole blood samples with PLT ≤50×10(9)/L were collected from the clinical laboratory at the First Affiliated Hospital of Soochow University to evaluate the precision of low-value PLT in PLT-8X mode. Moreover, 67 samples with PLT <100×10(9)/L were collected, and a methodological comparison was conducted between the results of PLT-8X and those obtained by the reference method [red blood cell (RBC)/PLT ratio method] recommended by the International Committee for Standardization of Hematology (ICSH). In addition, fresh whole blood and PLT-free plasma were used to prepare samples at a series of concentrations within the low-value ranges to evaluate the linearity of PLT and finally investigate the limit of blank (LoB) and limit of detection (LoD) for low-value PLT in PLT-8X mode. RESULTS: The precision low-value PLT in PLT-8X mode was significantly superior to that in PLT-I (P<0.0001); and the correlation coefficient of PLT-8X mode compared with the reference method of flow cytometer was 0.992. The expected bias at the low-value medical decision levels (5×10(9)/L, 10×10(9)/L and 20×10(9)/L) of PLT fell within the range of ±1.0×10(9)/L, and that at 50×10(9)/L fell within the range of ±1.5×10(9)/L. In addition, the correlation coefficient of PLT-8X in the low-value ranges (0 to 30×10(9)/L and 0 to 100×10(9)/L) of PLT was greater than 0.99, and its nonlinear coefficients were not significantly different from 0 (P>0.05). Finally, the LoB and LoD of PLT-8X mode were 0.33×10(9)/L and 0.89×10(9)/L, respectively. CONCLUSIONS: The PLT-8X mode of the BC-6800Plus auto hematology analyzer has extremely low detection sensitivity for low-value PLTs in anticoagulated venous whole blood samples. With good linearity, its precision and accuracy can sufficiently meet the needs for hematology laboratory use, and it can effectively help clinicians to accurately diagnose and treat patients with thrombopenia-related diseases. AME Publishing Company 2022-06 /pmc/articles/PMC9279826/ /pubmed/35845540 http://dx.doi.org/10.21037/atm-22-1501 Text en 2022 Annals of Translational Medicine. All rights reserved. https://creativecommons.org/licenses/by-nc-nd/4.0/Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) .
spellingShingle Original Article
Shen, Yimin
Liu, Dongmei
Wang, Yong
Cao, Jun
Zhang, Shuaishuai
Wen, Hui
Dong, Qiuqiu
Zheng, Dong
Qiu, Jun
Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus
title Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus
title_full Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus
title_fullStr Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus
title_full_unstemmed Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus
title_short Evaluation on clinical performance of the new low-value platelet measurement mode (PLT-8X) of BC-6800Plus
title_sort evaluation on clinical performance of the new low-value platelet measurement mode (plt-8x) of bc-6800plus
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9279826/
https://www.ncbi.nlm.nih.gov/pubmed/35845540
http://dx.doi.org/10.21037/atm-22-1501
work_keys_str_mv AT shenyimin evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT liudongmei evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT wangyong evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT caojun evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT zhangshuaishuai evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT wenhui evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT dongqiuqiu evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT zhengdong evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus
AT qiujun evaluationonclinicalperformanceofthenewlowvalueplateletmeasurementmodeplt8xofbc6800plus