Complications of Transcutaneous Protheses – A Systematic Review of Publications Over the Past 10 Years

Introduction  Bone conduction implants, responsible for transmitting sound from an external processor to the inner ear, can be divided into active and passive, depending on the vibratory stimulus location. The use of transcutaneous device has increased, given its aesthetic appeal, the complications...

Descripción completa

Detalles Bibliográficos
Autores principales: Souza, Marcos Antonio de, Vallejos Riart, Santiago Luis, Souza, Stephanie Rugeri de, de Brito, Rubens, Bento, Ricardo Ferreira
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Thieme Revinter Publicações Ltda. 2022
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9282975/
https://www.ncbi.nlm.nih.gov/pubmed/35846823
http://dx.doi.org/10.1055/s-0042-1742352
Descripción
Sumario:Introduction  Bone conduction implants, responsible for transmitting sound from an external processor to the inner ear, can be divided into active and passive, depending on the vibratory stimulus location. The use of transcutaneous device has increased, given its aesthetic appeal, the complications and limitations of percutaneous devices, and patient's treatment adherence, focusing mainly on efficacy. However, various complications are associated with the use of transcutaneous prosthesis, which can often be serious. Objective  To approach the literature on complications involving transcutaneous bone-anchored prostheses through a systematic review of articles published in the past 10 years (2011–2021). Data Synthesis  The MEDLINE, EMBASE, Scopus, and Cochrane Library databases were searched. All articles written in English reporting on currently available transcutaneous prosthesis implantation and its complications were selected. Studies on both children and adults were included. The data on complications were extracted, and complications were classified as minor or major and associated to each device used. Thirty-seven articles were included in the study, of which 14 were prospective cohort studies, 22 were retrospective case series, and 1 was a case report. Most studies (18) included both adults and children. Moreover, 901 implantations were performed, of which 552 implanted Baha Attract (Cochlear Ltd., Sydney, Australia), 244 implanted BoneBridge (MED-EL, Innsbruck, Austria), and 105 implanted Sophono (Sophono Inc., Boulder, CO, USA]). Furthermore, 192 adverse events were reported (total complication rate, 21.3%), with 161 minor complications (84.3%) and 31 major complications (16.1%). Conclusion  Transcutaneous prosthesis is an audiological alternative with fewer complications than percutaneous prosthesis. However, its indication should be judicious because complications are common, and although most complications are minor, serious infections requiring explantation may develop.