Validation of cardiac diffusion tensor imaging sequences: A multicentre test–retest phantom study

Cardiac diffusion tensor imaging (DTI) is an emerging technique for the in vivo characterisation of myocardial microstructure, and there is a growing need for its validation and standardisation. We sought to establish the accuracy, precision, repeatability and reproducibility of state‐of‐the‐art pul...

Descripción completa

Detalles Bibliográficos
Autores principales: Teh, Irvin, Romero R., William A., Boyle, Jordan, Coll‐Font, Jaume, Dall'Armellina, Erica, Ennis, Daniel B., Ferreira, Pedro F., Kalra, Prateek, Kolipaka, Arunark, Kozerke, Sebastian, Lohr, David, Mongeon, François‐Pierre, Moulin, Kévin, Nguyen, Christopher, Nielles‐Vallespin, Sonia, Raterman, Brian, Schreiber, Laura M., Scott, Andrew D., Sosnovik, David E., Stoeck, Christian T., Tous, Cyril, Tunnicliffe, Elizabeth M., Weng, Andreas M., Croisille, Pierre, Viallon, Magalie, Schneider, Jürgen E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9285553/
https://www.ncbi.nlm.nih.gov/pubmed/34967060
http://dx.doi.org/10.1002/nbm.4685
Descripción
Sumario:Cardiac diffusion tensor imaging (DTI) is an emerging technique for the in vivo characterisation of myocardial microstructure, and there is a growing need for its validation and standardisation. We sought to establish the accuracy, precision, repeatability and reproducibility of state‐of‐the‐art pulse sequences for cardiac DTI among 10 centres internationally. Phantoms comprising 0%–20% polyvinylpyrrolidone (PVP) were scanned with DTI using a product pulsed gradient spin echo (PGSE; N = 10 sites) sequence, and a custom motion‐compensated spin echo (SE; N = 5) or stimulated echo acquisition mode (STEAM; N = 5) sequence suitable for cardiac DTI in vivo. A second identical scan was performed 1–9 days later, and the data were analysed centrally. The average mean diffusivities (MDs) in 0% PVP were (1.124, 1.130, 1.113) x 10(−3) mm(2)/s for PGSE, SE and STEAM, respectively, and accurate to within 1.5% of reference data from the literature. The coefficients of variation in MDs across sites were 2.6%, 3.1% and 2.1% for PGSE, SE and STEAM, respectively, and were similar to previous studies using only PGSE. Reproducibility in MD was excellent, with mean differences in PGSE, SE and STEAM of (0.3 ± 2.3, 0.24 ± 0.95, 0.52 ± 0.58) x 10(−5) mm(2)/s (mean ± 1.96 SD). We show that custom sequences for cardiac DTI provide accurate, precise, repeatable and reproducible measurements. Further work in anisotropic and/or deforming phantoms is warranted.