Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma

BACKGROUND: The present study aimed to investigate the dosimetric impact of metal stent for photon and proton treatment plans in hepatocellular carcinoma. METHODS: With computed tomography data of a water-equivalent solid phantom, dose perturbation caused by a metal stent included in the photon and...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Boram, Cho, Sungkoo, Park, Hee Chul, Kang, Sang-Won, Kim, Jae-Sung, Chung, Jin-Beom
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9288675/
https://www.ncbi.nlm.nih.gov/pubmed/35842636
http://dx.doi.org/10.1186/s13014-022-02100-8
_version_ 1784748500388413440
author Lee, Boram
Cho, Sungkoo
Park, Hee Chul
Kang, Sang-Won
Kim, Jae-Sung
Chung, Jin-Beom
author_facet Lee, Boram
Cho, Sungkoo
Park, Hee Chul
Kang, Sang-Won
Kim, Jae-Sung
Chung, Jin-Beom
author_sort Lee, Boram
collection PubMed
description BACKGROUND: The present study aimed to investigate the dosimetric impact of metal stent for photon and proton treatment plans in hepatocellular carcinoma. METHODS: With computed tomography data of a water-equivalent solid phantom, dose perturbation caused by a metal stent included in the photon and proton treatment of hepatocellular carcinoma was evaluated by comparing Eclipse and RayStation treatment planning system (TPS) to a Monte Carlo (MC) based dose calculator. Photon and proton plans were created with anterior–posterior/posterior-anterior (AP/PA) fields using a 6 MV beam and AP/PA fields of a wobbling beam using 150 MeV and a 10 cm ridge filter. The difference in dose distributions and dosimetric parameters were compared depending on the stent's positions (the bile duct (GB) and intestinal tract (GI)) and angles (0°, 45°, and 90°). Additionally, the dose variation in the target volume including the stent was comparatively evaluated through dose volume histogram (DVH) analysis. And the comparison of clinical cases was carried out in the same way. RESULTS: Percentage differences in the dosimetric parameters calculated by MC ranged from − 7.0 to 3.9% for the photon plan and − 33.7 to 4.3% for the proton plan, depending on the angle at which the GB and GI stents were placed, compared to those without the stent. The maximum difference was observed at the minimum dose (D(min)), which was observed in both photon and proton plans in the GB and GI stents deployed at a 90° incidence angle. The parameter differences were greater in the proton plan than in photon plan. The target volume showed various dose variations depending on positions and angles of stent for both beams. Compared with no-stent, the doses within the target volume containing the GI and GB stents for the photon beam were overestimated in the high-dose area at 0°, nearly equal within 1% at 45°, and underestimated at 90°. These doses to the proton beam were underestimated at all angles, and the amount of underdose to the target volume increased with an increase in the stent angle. However, the difference was significantly greater with the proton plan than the photon plan. CONCLUSIONS: Dose perturbations within the target volume due to the presence of the metal stent were not observed in the TPS calculations for photon and proton beams, but MC was used to confirm that there are dose variations within the target volume. The MC results found that delivery of the treatment beam avoiding the stent was the best method to prevent target volume underdose.
format Online
Article
Text
id pubmed-9288675
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-92886752022-07-18 Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma Lee, Boram Cho, Sungkoo Park, Hee Chul Kang, Sang-Won Kim, Jae-Sung Chung, Jin-Beom Radiat Oncol Research BACKGROUND: The present study aimed to investigate the dosimetric impact of metal stent for photon and proton treatment plans in hepatocellular carcinoma. METHODS: With computed tomography data of a water-equivalent solid phantom, dose perturbation caused by a metal stent included in the photon and proton treatment of hepatocellular carcinoma was evaluated by comparing Eclipse and RayStation treatment planning system (TPS) to a Monte Carlo (MC) based dose calculator. Photon and proton plans were created with anterior–posterior/posterior-anterior (AP/PA) fields using a 6 MV beam and AP/PA fields of a wobbling beam using 150 MeV and a 10 cm ridge filter. The difference in dose distributions and dosimetric parameters were compared depending on the stent's positions (the bile duct (GB) and intestinal tract (GI)) and angles (0°, 45°, and 90°). Additionally, the dose variation in the target volume including the stent was comparatively evaluated through dose volume histogram (DVH) analysis. And the comparison of clinical cases was carried out in the same way. RESULTS: Percentage differences in the dosimetric parameters calculated by MC ranged from − 7.0 to 3.9% for the photon plan and − 33.7 to 4.3% for the proton plan, depending on the angle at which the GB and GI stents were placed, compared to those without the stent. The maximum difference was observed at the minimum dose (D(min)), which was observed in both photon and proton plans in the GB and GI stents deployed at a 90° incidence angle. The parameter differences were greater in the proton plan than in photon plan. The target volume showed various dose variations depending on positions and angles of stent for both beams. Compared with no-stent, the doses within the target volume containing the GI and GB stents for the photon beam were overestimated in the high-dose area at 0°, nearly equal within 1% at 45°, and underestimated at 90°. These doses to the proton beam were underestimated at all angles, and the amount of underdose to the target volume increased with an increase in the stent angle. However, the difference was significantly greater with the proton plan than the photon plan. CONCLUSIONS: Dose perturbations within the target volume due to the presence of the metal stent were not observed in the TPS calculations for photon and proton beams, but MC was used to confirm that there are dose variations within the target volume. The MC results found that delivery of the treatment beam avoiding the stent was the best method to prevent target volume underdose. BioMed Central 2022-07-16 /pmc/articles/PMC9288675/ /pubmed/35842636 http://dx.doi.org/10.1186/s13014-022-02100-8 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Lee, Boram
Cho, Sungkoo
Park, Hee Chul
Kang, Sang-Won
Kim, Jae-Sung
Chung, Jin-Beom
Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
title Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
title_full Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
title_fullStr Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
title_full_unstemmed Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
title_short Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
title_sort assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9288675/
https://www.ncbi.nlm.nih.gov/pubmed/35842636
http://dx.doi.org/10.1186/s13014-022-02100-8
work_keys_str_mv AT leeboram assessmentofdoseperturbationsformetalstentinphotonandprotonradiotherapyplansforhepatocellularcarcinoma
AT chosungkoo assessmentofdoseperturbationsformetalstentinphotonandprotonradiotherapyplansforhepatocellularcarcinoma
AT parkheechul assessmentofdoseperturbationsformetalstentinphotonandprotonradiotherapyplansforhepatocellularcarcinoma
AT kangsangwon assessmentofdoseperturbationsformetalstentinphotonandprotonradiotherapyplansforhepatocellularcarcinoma
AT kimjaesung assessmentofdoseperturbationsformetalstentinphotonandprotonradiotherapyplansforhepatocellularcarcinoma
AT chungjinbeom assessmentofdoseperturbationsformetalstentinphotonandprotonradiotherapyplansforhepatocellularcarcinoma