A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements
Corrosion and wear are commonly found at the taper‐trunnion connection of modular total hip arthroplasty (THA) explanted devices. While metal/metal (M/M) modular taper‐trunnion connections exhibit more wear/corrosion than ceramic/metal (C/M) modular taper‐trunnion connections, damage is present in b...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley & Sons, Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9292311/ https://www.ncbi.nlm.nih.gov/pubmed/34164932 http://dx.doi.org/10.1002/jbm.b.34897 |
_version_ | 1784749339690663936 |
---|---|
author | Cadel, Eileen S. Topoleski, L.D. Timmie Vesnovsky, Oleg Anderson, Charles R. Hopper, Robert H. Engh, Charles A. Di Prima, Matthew A. |
author_facet | Cadel, Eileen S. Topoleski, L.D. Timmie Vesnovsky, Oleg Anderson, Charles R. Hopper, Robert H. Engh, Charles A. Di Prima, Matthew A. |
author_sort | Cadel, Eileen S. |
collection | PubMed |
description | Corrosion and wear are commonly found at the taper‐trunnion connection of modular total hip arthroplasty (THA) explanted devices. While metal/metal (M/M) modular taper‐trunnion connections exhibit more wear/corrosion than ceramic/metal (C/M) modular taper‐trunnion connections, damage is present in both, regardless of material. This study used a combination of assessment techniques including clinical data, visual scoring assessment, optical imaging, profilometry, and x‐ray photoelectron microscopy (XPS), to investigate wear mechanisms and damage features at the modular taper‐trunnion connection of 10 M/M and 8 C/M explanted THAs. No correlation was found between any demographic variable and corrosion wear and assessment scores. All assessment techniques demonstrated that the stem trunnions had more damage than head tapers for both explant groups and agreed that C/M explants had less corrosion and wear compared to M/M explants. However, visual assessment scores differed between assessment techniques when evaluating the tapers and trunnions within the two groups. Profilometry showed an increase (p <.05) in surface roughness for stem trunnions compared to head tapers for both explant groups. X‐ray photoelectron spectroscopy performed on deposits from two M/M explants found chromium and molybdenum carbides beneath the surface while chromium sulfate and aged bone mineral were found on the surface suggesting that the debris is a result of corrosion rather than wear. These results indicate that taper‐trunnion damage is more prevalent for M/M explants, but C/M explants are still susceptible to damage. More comprehensive analysis of damage is necessary to better understand the origins of taper‐trunnion damage. |
format | Online Article Text |
id | pubmed-9292311 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | John Wiley & Sons, Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-92923112022-07-20 A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements Cadel, Eileen S. Topoleski, L.D. Timmie Vesnovsky, Oleg Anderson, Charles R. Hopper, Robert H. Engh, Charles A. Di Prima, Matthew A. J Biomed Mater Res B Appl Biomater Research Articles Corrosion and wear are commonly found at the taper‐trunnion connection of modular total hip arthroplasty (THA) explanted devices. While metal/metal (M/M) modular taper‐trunnion connections exhibit more wear/corrosion than ceramic/metal (C/M) modular taper‐trunnion connections, damage is present in both, regardless of material. This study used a combination of assessment techniques including clinical data, visual scoring assessment, optical imaging, profilometry, and x‐ray photoelectron microscopy (XPS), to investigate wear mechanisms and damage features at the modular taper‐trunnion connection of 10 M/M and 8 C/M explanted THAs. No correlation was found between any demographic variable and corrosion wear and assessment scores. All assessment techniques demonstrated that the stem trunnions had more damage than head tapers for both explant groups and agreed that C/M explants had less corrosion and wear compared to M/M explants. However, visual assessment scores differed between assessment techniques when evaluating the tapers and trunnions within the two groups. Profilometry showed an increase (p <.05) in surface roughness for stem trunnions compared to head tapers for both explant groups. X‐ray photoelectron spectroscopy performed on deposits from two M/M explants found chromium and molybdenum carbides beneath the surface while chromium sulfate and aged bone mineral were found on the surface suggesting that the debris is a result of corrosion rather than wear. These results indicate that taper‐trunnion damage is more prevalent for M/M explants, but C/M explants are still susceptible to damage. More comprehensive analysis of damage is necessary to better understand the origins of taper‐trunnion damage. John Wiley & Sons, Inc. 2021-06-23 2022-01 /pmc/articles/PMC9292311/ /pubmed/34164932 http://dx.doi.org/10.1002/jbm.b.34897 Text en © 2021 The Authors. Journal of Biomedical Materials Research Part B: Applied Biomaterials published by Wiley Periodicals LLC. This article is a U.S. Government work and is in the public domain in the USA. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Articles Cadel, Eileen S. Topoleski, L.D. Timmie Vesnovsky, Oleg Anderson, Charles R. Hopper, Robert H. Engh, Charles A. Di Prima, Matthew A. A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
title | A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
title_full | A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
title_fullStr | A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
title_full_unstemmed | A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
title_short | A comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
title_sort | comparison of metal/metal and ceramic/metal taper‐trunnion modular connections in explanted total hip replacements |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9292311/ https://www.ncbi.nlm.nih.gov/pubmed/34164932 http://dx.doi.org/10.1002/jbm.b.34897 |
work_keys_str_mv | AT cadeleileens acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT topoleskildtimmie acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT vesnovskyoleg acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT andersoncharlesr acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT hopperroberth acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT enghcharlesa acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT diprimamatthewa acomparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT cadeleileens comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT topoleskildtimmie comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT vesnovskyoleg comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT andersoncharlesr comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT hopperroberth comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT enghcharlesa comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements AT diprimamatthewa comparisonofmetalmetalandceramicmetaltapertrunnionmodularconnectionsinexplantedtotalhipreplacements |