Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice
Lymphedema affects 20% of women diagnosed with breast cancer. It is a pathology with no known cure. Animal models are essential to explore possible treatments to understand and potentially cure lymphedema. The rodent hindlimb lymphedema model is one of the most widely used. Different modalities have...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9293915/ https://www.ncbi.nlm.nih.gov/pubmed/35851094 http://dx.doi.org/10.1038/s41598-022-16311-2 |
_version_ | 1784749741579436032 |
---|---|
author | Bucan, Amar Wiinholt, Alexander Dalaei, Farima Gerke, Oke Hansen, Christian Rønn Sørensen, Jens Ahm |
author_facet | Bucan, Amar Wiinholt, Alexander Dalaei, Farima Gerke, Oke Hansen, Christian Rønn Sørensen, Jens Ahm |
author_sort | Bucan, Amar |
collection | PubMed |
description | Lymphedema affects 20% of women diagnosed with breast cancer. It is a pathology with no known cure. Animal models are essential to explore possible treatments to understand and potentially cure lymphedema. The rodent hindlimb lymphedema model is one of the most widely used. Different modalities have been used to measure lymphedema in the hindlimb of mice, and these are generally poorly assessed in terms of the interrater agreement; thus, there could be a risk of measuring bias and poor reproducibility. We examined the interrater agreement of µCT-scans, electronic caliper thickness of the paw and plethysmometer in the measurement of lymphedema in the hindlimb of mice. Three independent raters assessed 24 C57BL6 mice using these three modalities four times (week 1, 2, 4 and 8) with a total of 96 samples. The mean interrater differences were then calculated. The interrater agreement was highest in the µCT-scans, with an extremely low risk of measurement bias. The interrater agreement in the plethysmometer and electronic caliper was comparable with a low to moderate risk of measurement bias. The µCT-scanner should be used whenever possible. The electronic caliper should only be used if there is no µCT-scanner available. The plethysmometer should not be used in rodents of this size. |
format | Online Article Text |
id | pubmed-9293915 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Nature Publishing Group UK |
record_format | MEDLINE/PubMed |
spelling | pubmed-92939152022-07-20 Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice Bucan, Amar Wiinholt, Alexander Dalaei, Farima Gerke, Oke Hansen, Christian Rønn Sørensen, Jens Ahm Sci Rep Article Lymphedema affects 20% of women diagnosed with breast cancer. It is a pathology with no known cure. Animal models are essential to explore possible treatments to understand and potentially cure lymphedema. The rodent hindlimb lymphedema model is one of the most widely used. Different modalities have been used to measure lymphedema in the hindlimb of mice, and these are generally poorly assessed in terms of the interrater agreement; thus, there could be a risk of measuring bias and poor reproducibility. We examined the interrater agreement of µCT-scans, electronic caliper thickness of the paw and plethysmometer in the measurement of lymphedema in the hindlimb of mice. Three independent raters assessed 24 C57BL6 mice using these three modalities four times (week 1, 2, 4 and 8) with a total of 96 samples. The mean interrater differences were then calculated. The interrater agreement was highest in the µCT-scans, with an extremely low risk of measurement bias. The interrater agreement in the plethysmometer and electronic caliper was comparable with a low to moderate risk of measurement bias. The µCT-scanner should be used whenever possible. The electronic caliper should only be used if there is no µCT-scanner available. The plethysmometer should not be used in rodents of this size. Nature Publishing Group UK 2022-07-18 /pmc/articles/PMC9293915/ /pubmed/35851094 http://dx.doi.org/10.1038/s41598-022-16311-2 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Article Bucan, Amar Wiinholt, Alexander Dalaei, Farima Gerke, Oke Hansen, Christian Rønn Sørensen, Jens Ahm Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
title | Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
title_full | Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
title_fullStr | Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
title_full_unstemmed | Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
title_short | Microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
title_sort | microcomputed tomography versus plethysmometer and electronic caliper in the measurements of lymphedema in the hindlimb of mice |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9293915/ https://www.ncbi.nlm.nih.gov/pubmed/35851094 http://dx.doi.org/10.1038/s41598-022-16311-2 |
work_keys_str_mv | AT bucanamar microcomputedtomographyversusplethysmometerandelectroniccaliperinthemeasurementsoflymphedemainthehindlimbofmice AT wiinholtalexander microcomputedtomographyversusplethysmometerandelectroniccaliperinthemeasurementsoflymphedemainthehindlimbofmice AT dalaeifarima microcomputedtomographyversusplethysmometerandelectroniccaliperinthemeasurementsoflymphedemainthehindlimbofmice AT gerkeoke microcomputedtomographyversusplethysmometerandelectroniccaliperinthemeasurementsoflymphedemainthehindlimbofmice AT hansenchristianrønn microcomputedtomographyversusplethysmometerandelectroniccaliperinthemeasurementsoflymphedemainthehindlimbofmice AT sørensenjensahm microcomputedtomographyversusplethysmometerandelectroniccaliperinthemeasurementsoflymphedemainthehindlimbofmice |