An active fixation quadripolar left ventricular lead for cardiac resynchronization therapy with reduced postoperative complication rates

BACKGROUND: The rate of left ventricular (LV) lead displacement after cardiac resynchronization therapy (CRT) remains high despite improvements in lead technology. In 2017, a novel quadripolar lead with active fixation technology became available in the UK. METHODS: This was a retrospective, observa...

Descripción completa

Detalles Bibliográficos
Autores principales: Robertson, Calum, Duffey, Owen, Tang, Pok‐Tin, Fairhurst, Natalie, Monteiro, Cristiana, Green, Peregrine, Grogono, Joanna, Davies, Mark, Lewis, Andrew, Wijesurendra, Rohan, Ormerod, Julian, Gamble, James, Ginks, Matthew, Rajappan, Kim, Bashir, Yaver, Betts, Tim R., Herring, Neil
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9304298/
https://www.ncbi.nlm.nih.gov/pubmed/34968010
http://dx.doi.org/10.1111/jce.15346
Descripción
Sumario:BACKGROUND: The rate of left ventricular (LV) lead displacement after cardiac resynchronization therapy (CRT) remains high despite improvements in lead technology. In 2017, a novel quadripolar lead with active fixation technology became available in the UK. METHODS: This was a retrospective, observational study analyzing device complications in 476 consecutive patients undergoing successful first‐time implantation of a CRT device at a tertiary center from 2017 to 2020. RESULTS: Both active (n = 135) and passive fixation (n = 341) quadripolar leads had similar success rates for implantation (99.3% vs. 98.8%, p = 1.00), although the pacing threshold (0.89 [0.60–1.25] vs. 1.00 [0.70–1.60] V, p = .01) and lead impedance (632 [552–794] vs. 730 [636–862] Ohms, p < .0001) were significantly lower for the active fixation lead. Patients receiving an active fixation lead had a reduced incidence of lead displacement at 6 months (0.74% vs. 4.69%, p = .036). There was no significant difference in the rate of right atrial (RA) and right ventricular (RV) lead displacement between the two groups (RA: 1.48% vs. 1.17%, p = .68; RV: 2.22% vs. 1.76%, p = .72). Reprogramming the LV lead after displacement was unsuccessful in most cases (successful reprogramming: Active fix = 0/1, Passive fix = 1/16) therefore nearly all patients required a repeat procedure. As a result, the rate of intervention within 6 months for lead displacement was significantly lower when patients were implanted with the active fixation lead (0.74% vs. 4.40%, p = .049). CONCLUSION: The novel active fixation lead in our study has a lower incidence of lead displacement and re‐intervention compared to conventional quadripolar leads for CRT.