Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation

PURPOSE: To describe and evaluate relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) for ocular surface sensitivity measurement. METHODS: Characteristics of Liquid Jet (LJ) droplets (consisting of isotonic saline solution) were analysed: vertical and...

Descripción completa

Detalles Bibliográficos
Autores principales: Nosch, Daniela S, Oscity, Matthias, Steigmeier, Peter, Käser, Emanuele, Loepfe, Markus, Joos, Roland E
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9305237/
https://www.ncbi.nlm.nih.gov/pubmed/35156726
http://dx.doi.org/10.1111/opo.12962
_version_ 1784752277540569088
author Nosch, Daniela S
Oscity, Matthias
Steigmeier, Peter
Käser, Emanuele
Loepfe, Markus
Joos, Roland E
author_facet Nosch, Daniela S
Oscity, Matthias
Steigmeier, Peter
Käser, Emanuele
Loepfe, Markus
Joos, Roland E
author_sort Nosch, Daniela S
collection PubMed
description PURPOSE: To describe and evaluate relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) for ocular surface sensitivity measurement. METHODS: Characteristics of Liquid Jet (LJ) droplets (consisting of isotonic saline solution) were analysed: vertical and horizontal displacement and speed of LJ droplets were recorded with the aid of the High Speed Photron FASTCAM NOVA S6 camera (stimulus duration: 40 ms). Stimulus mass was assessed for 20 sets of 10 LJs with aid of a microbalance (pressure range of 100–1500 mbar). RESULTS: Because continuous flow LJ disintegrated into droplets in the lower pressure range (<700 mbar), pulsed stimuli were applied in order to obtain similar stimulus characteristics across the applied pressure range. For all measurements, very little variability was observed. Vertical and horizontal displacement did not exceed 0.13 mm in either direction. The mass per shot showed an unexpected cubic dependency on pressure. Up to approximately 700 mbar, LJ speed showed an almost linear relationship. For the pressure range of >700–1500 mbar, variability increased and speed decreased compared to the expected in a linear manner. However, this may be caused by the difficulty of identifying pattern changes of LJ droplets from one high speed image frame to the next with increasing stimulus speed, when determining LJ speed via pixel count. CONCLUSIONS: Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity was shown to deliver fine droplets with a pulsed stimulus mode, in a repeatable manner with precise localisation to the ocular surface. Very little variability was observed in LJ speed and mass for the typical pressure range required for clinical sensitivity measurements.
format Online
Article
Text
id pubmed-9305237
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-93052372022-07-28 Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation Nosch, Daniela S Oscity, Matthias Steigmeier, Peter Käser, Emanuele Loepfe, Markus Joos, Roland E Ophthalmic Physiol Opt Regular Articles PURPOSE: To describe and evaluate relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) for ocular surface sensitivity measurement. METHODS: Characteristics of Liquid Jet (LJ) droplets (consisting of isotonic saline solution) were analysed: vertical and horizontal displacement and speed of LJ droplets were recorded with the aid of the High Speed Photron FASTCAM NOVA S6 camera (stimulus duration: 40 ms). Stimulus mass was assessed for 20 sets of 10 LJs with aid of a microbalance (pressure range of 100–1500 mbar). RESULTS: Because continuous flow LJ disintegrated into droplets in the lower pressure range (<700 mbar), pulsed stimuli were applied in order to obtain similar stimulus characteristics across the applied pressure range. For all measurements, very little variability was observed. Vertical and horizontal displacement did not exceed 0.13 mm in either direction. The mass per shot showed an unexpected cubic dependency on pressure. Up to approximately 700 mbar, LJ speed showed an almost linear relationship. For the pressure range of >700–1500 mbar, variability increased and speed decreased compared to the expected in a linear manner. However, this may be caused by the difficulty of identifying pattern changes of LJ droplets from one high speed image frame to the next with increasing stimulus speed, when determining LJ speed via pixel count. CONCLUSIONS: Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity was shown to deliver fine droplets with a pulsed stimulus mode, in a repeatable manner with precise localisation to the ocular surface. Very little variability was observed in LJ speed and mass for the typical pressure range required for clinical sensitivity measurements. John Wiley and Sons Inc. 2022-02-14 2022-05 /pmc/articles/PMC9305237/ /pubmed/35156726 http://dx.doi.org/10.1111/opo.12962 Text en © 2022 The Authors. Ophthalmic and Physiological Optics published by John Wiley & Sons Ltd on behalf of College of Optometrists https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Regular Articles
Nosch, Daniela S
Oscity, Matthias
Steigmeier, Peter
Käser, Emanuele
Loepfe, Markus
Joos, Roland E
Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation
title Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation
title_full Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation
title_fullStr Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation
title_full_unstemmed Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation
title_short Working principle and relevant physical properties of the Swiss Liquid Jet Aesthesiometer for Corneal Sensitivity (SLACS) evaluation
title_sort working principle and relevant physical properties of the swiss liquid jet aesthesiometer for corneal sensitivity (slacs) evaluation
topic Regular Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9305237/
https://www.ncbi.nlm.nih.gov/pubmed/35156726
http://dx.doi.org/10.1111/opo.12962
work_keys_str_mv AT noschdanielas workingprincipleandrelevantphysicalpropertiesoftheswissliquidjetaesthesiometerforcornealsensitivityslacsevaluation
AT oscitymatthias workingprincipleandrelevantphysicalpropertiesoftheswissliquidjetaesthesiometerforcornealsensitivityslacsevaluation
AT steigmeierpeter workingprincipleandrelevantphysicalpropertiesoftheswissliquidjetaesthesiometerforcornealsensitivityslacsevaluation
AT kaseremanuele workingprincipleandrelevantphysicalpropertiesoftheswissliquidjetaesthesiometerforcornealsensitivityslacsevaluation
AT loepfemarkus workingprincipleandrelevantphysicalpropertiesoftheswissliquidjetaesthesiometerforcornealsensitivityslacsevaluation
AT joosrolande workingprincipleandrelevantphysicalpropertiesoftheswissliquidjetaesthesiometerforcornealsensitivityslacsevaluation