Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation
PURPOSE: To investigate drug contamination of the working environment with paclitaxel drug-coated balloon (DCB) angioplasty due to loss of paclitaxel containing particles from the coating during DCB preparation, insertion, and inflation. MATERIAL AND METHODS: In an experimetal laboratory setting, dr...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9307540/ https://www.ncbi.nlm.nih.gov/pubmed/35689119 http://dx.doi.org/10.1007/s00270-022-03164-5 |
_version_ | 1784752785484414976 |
---|---|
author | Faenger, Bernd Heinrich, Andreas Hilger, Ingrid Teichgräber, Ulf |
author_facet | Faenger, Bernd Heinrich, Andreas Hilger, Ingrid Teichgräber, Ulf |
author_sort | Faenger, Bernd |
collection | PubMed |
description | PURPOSE: To investigate drug contamination of the working environment with paclitaxel drug-coated balloon (DCB) angioplasty due to loss of paclitaxel containing particles from the coating during DCB preparation, insertion, and inflation. MATERIAL AND METHODS: In an experimetal laboratory setting, drug loss during removal of the protective cover and insertion of the DCB through the hemostatic valve of the introducer sheath and after inflation was examined. In seven DCB types of different manufacturers, semi-quantitative image analysis was performed during five standardized tests cycles. Additionally, every DCB type passed one cycle of a wipe test and one cycle of air sampling. RESULTS: By removing the protective cover, the paclitaxel-covered balloon surface was significantly reduced in 3 out of 7 products (P = 0.043). Overall, extend of decline ranged from 0.4 to 12%. In 6 of 7 products, powdered paclitaxel clusters dropped down upon removal of the protective cover (0.099 ng/cm(2) up to approx. 22 ng/cm(2)). Contamination of the air was detected in none of the DCB types. When pushed through the vascular sheath, none of the investigated DCB types showed a significant loss of paclitaxel from the coated balloon surface. After balloon inflation, the paclitaxel-coated surface area varied between manufacturers ranging from 25.9 to 97.8%. CONCLUSION: In some DCB types, the removal of the protective cover already leads to a significant loss of paclitaxel and paclitaxel-coated surfaces. As a result, there will be a contamination of the workplace and a reduction in the therapeutic dose. LEVEL OF EVIDENCE: No level of evidence. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00270-022-03164-5. |
format | Online Article Text |
id | pubmed-9307540 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Springer US |
record_format | MEDLINE/PubMed |
spelling | pubmed-93075402022-07-24 Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation Faenger, Bernd Heinrich, Andreas Hilger, Ingrid Teichgräber, Ulf Cardiovasc Intervent Radiol Laboratory Investigation PURPOSE: To investigate drug contamination of the working environment with paclitaxel drug-coated balloon (DCB) angioplasty due to loss of paclitaxel containing particles from the coating during DCB preparation, insertion, and inflation. MATERIAL AND METHODS: In an experimetal laboratory setting, drug loss during removal of the protective cover and insertion of the DCB through the hemostatic valve of the introducer sheath and after inflation was examined. In seven DCB types of different manufacturers, semi-quantitative image analysis was performed during five standardized tests cycles. Additionally, every DCB type passed one cycle of a wipe test and one cycle of air sampling. RESULTS: By removing the protective cover, the paclitaxel-covered balloon surface was significantly reduced in 3 out of 7 products (P = 0.043). Overall, extend of decline ranged from 0.4 to 12%. In 6 of 7 products, powdered paclitaxel clusters dropped down upon removal of the protective cover (0.099 ng/cm(2) up to approx. 22 ng/cm(2)). Contamination of the air was detected in none of the DCB types. When pushed through the vascular sheath, none of the investigated DCB types showed a significant loss of paclitaxel from the coated balloon surface. After balloon inflation, the paclitaxel-coated surface area varied between manufacturers ranging from 25.9 to 97.8%. CONCLUSION: In some DCB types, the removal of the protective cover already leads to a significant loss of paclitaxel and paclitaxel-coated surfaces. As a result, there will be a contamination of the workplace and a reduction in the therapeutic dose. LEVEL OF EVIDENCE: No level of evidence. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00270-022-03164-5. Springer US 2022-06-10 2022 /pmc/articles/PMC9307540/ /pubmed/35689119 http://dx.doi.org/10.1007/s00270-022-03164-5 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Laboratory Investigation Faenger, Bernd Heinrich, Andreas Hilger, Ingrid Teichgräber, Ulf Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation |
title | Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation |
title_full | Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation |
title_fullStr | Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation |
title_full_unstemmed | Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation |
title_short | Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation |
title_sort | drug loss from paclitaxel-coated balloons during preparation, insertion and inflation for angioplasty: a laboratory investigation |
topic | Laboratory Investigation |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9307540/ https://www.ncbi.nlm.nih.gov/pubmed/35689119 http://dx.doi.org/10.1007/s00270-022-03164-5 |
work_keys_str_mv | AT faengerbernd druglossfrompaclitaxelcoatedballoonsduringpreparationinsertionandinflationforangioplastyalaboratoryinvestigation AT heinrichandreas druglossfrompaclitaxelcoatedballoonsduringpreparationinsertionandinflationforangioplastyalaboratoryinvestigation AT hilgeringrid druglossfrompaclitaxelcoatedballoonsduringpreparationinsertionandinflationforangioplastyalaboratoryinvestigation AT teichgraberulf druglossfrompaclitaxelcoatedballoonsduringpreparationinsertionandinflationforangioplastyalaboratoryinvestigation |