Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer

BACKGROUND AND AIMS: GEM Premier ChemSTAT is a new point-of-care system providing rapid creatinine, BUN and tCO(2) measurements together with electrolytes, metabolites, hematocrit, pH and pCO(2) from a single whole blood specimen in acute care settings such as emergency departments and intensive car...

Descripción completa

Detalles Bibliográficos
Autores principales: García-Fernández, Alba E., Barquín, Raquel, Martínez, Mariano, Ferrer, Roser, Casis, Ernesto, Xu, Clarke
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9307927/
https://www.ncbi.nlm.nih.gov/pubmed/35880117
http://dx.doi.org/10.1016/j.plabm.2022.e00297
_version_ 1784752871818919936
author García-Fernández, Alba E.
Barquín, Raquel
Martínez, Mariano
Ferrer, Roser
Casis, Ernesto
Xu, Clarke
author_facet García-Fernández, Alba E.
Barquín, Raquel
Martínez, Mariano
Ferrer, Roser
Casis, Ernesto
Xu, Clarke
author_sort García-Fernández, Alba E.
collection PubMed
description BACKGROUND AND AIMS: GEM Premier ChemSTAT is a new point-of-care system providing rapid creatinine, BUN and tCO(2) measurements together with electrolytes, metabolites, hematocrit, pH and pCO(2) from a single whole blood specimen in acute care settings such as emergency departments and intensive care units. Accurate measurements of whole blood creatinine can aid in the diagnosis and treatment of renal diseases. MATERIALS AND METHODS: Heparinized whole blood samples from different clinical locations were evaluated on the GEM Premier ChemSTAT and results compared to plasma from the same samples on the Beckman AU5800 or whole blood on the GEM Premier 4000. Precision studies were conducted with whole blood and quality control material. RESULTS: ChemSTAT correlated well with plasma samples on the AU5800 (regression slopes (S): 0.957–1.159, correlation coefficients (r)≥0.952) and with whole blood specimens on the GEM Premier 4000 (S: 0.9646–1.124, r ≥ 0.974). The repeatability was 0.1%–3.1% and QC precision were within lab and manufacturers’ specifications. CONCLUSION: ChemSTAT demonstrated strong correlation to the comparative methods and excellent precision. Combining with its continuous quality management, ChemSTAT is suitable for acute care settings to provide rapid, reliable results, which could minimize time-to-treatment and improve patient outcome.
format Online
Article
Text
id pubmed-9307927
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-93079272022-07-24 Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer García-Fernández, Alba E. Barquín, Raquel Martínez, Mariano Ferrer, Roser Casis, Ernesto Xu, Clarke Pract Lab Med Original Research Article BACKGROUND AND AIMS: GEM Premier ChemSTAT is a new point-of-care system providing rapid creatinine, BUN and tCO(2) measurements together with electrolytes, metabolites, hematocrit, pH and pCO(2) from a single whole blood specimen in acute care settings such as emergency departments and intensive care units. Accurate measurements of whole blood creatinine can aid in the diagnosis and treatment of renal diseases. MATERIALS AND METHODS: Heparinized whole blood samples from different clinical locations were evaluated on the GEM Premier ChemSTAT and results compared to plasma from the same samples on the Beckman AU5800 or whole blood on the GEM Premier 4000. Precision studies were conducted with whole blood and quality control material. RESULTS: ChemSTAT correlated well with plasma samples on the AU5800 (regression slopes (S): 0.957–1.159, correlation coefficients (r)≥0.952) and with whole blood specimens on the GEM Premier 4000 (S: 0.9646–1.124, r ≥ 0.974). The repeatability was 0.1%–3.1% and QC precision were within lab and manufacturers’ specifications. CONCLUSION: ChemSTAT demonstrated strong correlation to the comparative methods and excellent precision. Combining with its continuous quality management, ChemSTAT is suitable for acute care settings to provide rapid, reliable results, which could minimize time-to-treatment and improve patient outcome. Elsevier 2022-07-19 /pmc/articles/PMC9307927/ /pubmed/35880117 http://dx.doi.org/10.1016/j.plabm.2022.e00297 Text en © 2022 The Authors. Published by Elsevier B.V. https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Research Article
García-Fernández, Alba E.
Barquín, Raquel
Martínez, Mariano
Ferrer, Roser
Casis, Ernesto
Xu, Clarke
Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
title Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
title_full Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
title_fullStr Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
title_full_unstemmed Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
title_short Performance evaluation of a point of care cartridge of the new GEM Premier ChemSTAT analyzer
title_sort performance evaluation of a point of care cartridge of the new gem premier chemstat analyzer
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9307927/
https://www.ncbi.nlm.nih.gov/pubmed/35880117
http://dx.doi.org/10.1016/j.plabm.2022.e00297
work_keys_str_mv AT garciafernandezalbae performanceevaluationofapointofcarecartridgeofthenewgempremierchemstatanalyzer
AT barquinraquel performanceevaluationofapointofcarecartridgeofthenewgempremierchemstatanalyzer
AT martinezmariano performanceevaluationofapointofcarecartridgeofthenewgempremierchemstatanalyzer
AT ferrerroser performanceevaluationofapointofcarecartridgeofthenewgempremierchemstatanalyzer
AT casisernesto performanceevaluationofapointofcarecartridgeofthenewgempremierchemstatanalyzer
AT xuclarke performanceevaluationofapointofcarecartridgeofthenewgempremierchemstatanalyzer