Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector
BACKGROUND: Electronic brachytherapy (eBT) is considered a safe treatment with good outcomes. However, eBT lacks standardized and independent dose verification, which could impede future use. PURPOSE: To validate the 3D dose‐to‐water distribution of an electronic brachytherapy (eBT) source using a s...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9314913/ https://www.ncbi.nlm.nih.gov/pubmed/35196404 http://dx.doi.org/10.1002/mp.15568 |
_version_ | 1784754433011220480 |
---|---|
author | Georgi, Peter Kertzscher, Gustavo Nyvang, Lars Šolc, Jaroslav Schneider, Thorsten Tanderup, Kari Johansen, Jacob Graversen |
author_facet | Georgi, Peter Kertzscher, Gustavo Nyvang, Lars Šolc, Jaroslav Schneider, Thorsten Tanderup, Kari Johansen, Jacob Graversen |
author_sort | Georgi, Peter |
collection | PubMed |
description | BACKGROUND: Electronic brachytherapy (eBT) is considered a safe treatment with good outcomes. However, eBT lacks standardized and independent dose verification, which could impede future use. PURPOSE: To validate the 3D dose‐to‐water distribution of an electronic brachytherapy (eBT) source using a small‐volume plastic scintillation detector (PSD). METHODS: The relative dose distribution of a Papillon 50 (P50) (Ariane Medical Systems, UK) eBT source was measured in water with a PSD consisting of a cylindrical scintillating BCF‐12 fiber (length: 0.5 mm, Ø: 1 mm) coupled to a photodetector via an optical fiber. The measurements were performed with the PSD mounted on a motorized stage in a water phantom (MP3) (PTW, Germany). This allowed the sensitive volume of the PSD to be moved to predetermined positions relative to the P50 applicator, which pointed vertically downward while just breaching the water surface. The percentage depth‐dose (PDD) was measured from 0 to 50 mm source‐to‐detector distance (SDD) in 1–3 mm steps. Dose profiles were measured along two perpendicular axes at five different SDDs with step sizes down to 0.5 mm. Characterization of the PSD consisted of determining the energy correction through Monte Carlo (MC) simulation and by measuring the stability and dose rate linearity using a well‐type ionization chamber as a reference. The measured PDD and profiles were validated with corresponding MC simulations. RESULTS: The measured and simulated PDD curves agreed within 2% (except at 0 mm and 43 mm depth) after the PSD measurements were corrected for energy dependency. The absorbed dose decreased by a factor of 2 at 7 mm depth and by a factor of 10 at 26 mm depth. The measured dose profiles showed dose gradients at the profile edges of more than 50%/mm at 5 mm depth and 15%/mm at 50 mm depth. The measured profile widths increased 0.66 mm per 1 mm depth, while the simulated profile widths increased 0.74 mm per 1 mm depth. An azimuthal dependency of > 10% was observed in the dose at 10 mm distance from the beam center. The total uncertainty of the measured relative dose is < 2.5% with a positional uncertainty of 0.4 mm. The measurements for a full 3D dose characterization (PDD and profiles) can be carried out within 8 h, the limiting factor being cooling of the P50. CONCLUSION: The PSD and MP3 water phantoms provided a method to independently verify the relative 3D dose distribution in water of an eBT source. |
format | Online Article Text |
id | pubmed-9314913 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-93149132022-07-30 Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector Georgi, Peter Kertzscher, Gustavo Nyvang, Lars Šolc, Jaroslav Schneider, Thorsten Tanderup, Kari Johansen, Jacob Graversen Med Phys COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY BACKGROUND: Electronic brachytherapy (eBT) is considered a safe treatment with good outcomes. However, eBT lacks standardized and independent dose verification, which could impede future use. PURPOSE: To validate the 3D dose‐to‐water distribution of an electronic brachytherapy (eBT) source using a small‐volume plastic scintillation detector (PSD). METHODS: The relative dose distribution of a Papillon 50 (P50) (Ariane Medical Systems, UK) eBT source was measured in water with a PSD consisting of a cylindrical scintillating BCF‐12 fiber (length: 0.5 mm, Ø: 1 mm) coupled to a photodetector via an optical fiber. The measurements were performed with the PSD mounted on a motorized stage in a water phantom (MP3) (PTW, Germany). This allowed the sensitive volume of the PSD to be moved to predetermined positions relative to the P50 applicator, which pointed vertically downward while just breaching the water surface. The percentage depth‐dose (PDD) was measured from 0 to 50 mm source‐to‐detector distance (SDD) in 1–3 mm steps. Dose profiles were measured along two perpendicular axes at five different SDDs with step sizes down to 0.5 mm. Characterization of the PSD consisted of determining the energy correction through Monte Carlo (MC) simulation and by measuring the stability and dose rate linearity using a well‐type ionization chamber as a reference. The measured PDD and profiles were validated with corresponding MC simulations. RESULTS: The measured and simulated PDD curves agreed within 2% (except at 0 mm and 43 mm depth) after the PSD measurements were corrected for energy dependency. The absorbed dose decreased by a factor of 2 at 7 mm depth and by a factor of 10 at 26 mm depth. The measured dose profiles showed dose gradients at the profile edges of more than 50%/mm at 5 mm depth and 15%/mm at 50 mm depth. The measured profile widths increased 0.66 mm per 1 mm depth, while the simulated profile widths increased 0.74 mm per 1 mm depth. An azimuthal dependency of > 10% was observed in the dose at 10 mm distance from the beam center. The total uncertainty of the measured relative dose is < 2.5% with a positional uncertainty of 0.4 mm. The measurements for a full 3D dose characterization (PDD and profiles) can be carried out within 8 h, the limiting factor being cooling of the P50. CONCLUSION: The PSD and MP3 water phantoms provided a method to independently verify the relative 3D dose distribution in water of an eBT source. John Wiley and Sons Inc. 2022-03-03 2022-05 /pmc/articles/PMC9314913/ /pubmed/35196404 http://dx.doi.org/10.1002/mp.15568 Text en © 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY Georgi, Peter Kertzscher, Gustavo Nyvang, Lars Šolc, Jaroslav Schneider, Thorsten Tanderup, Kari Johansen, Jacob Graversen Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector |
title | Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector |
title_full | Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector |
title_fullStr | Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector |
title_full_unstemmed | Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector |
title_short | Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector |
title_sort | toward 3d dose verification of an electronic brachytherapy source with a plastic scintillation detector |
topic | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9314913/ https://www.ncbi.nlm.nih.gov/pubmed/35196404 http://dx.doi.org/10.1002/mp.15568 |
work_keys_str_mv | AT georgipeter toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector AT kertzschergustavo toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector AT nyvanglars toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector AT solcjaroslav toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector AT schneiderthorsten toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector AT tanderupkari toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector AT johansenjacobgraversen toward3ddoseverificationofanelectronicbrachytherapysourcewithaplasticscintillationdetector |