Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system

INTRODUCTION: Few predictors of low capture threshold before the deployment of the Micra transcatheter pacing system (Micra TPS) have been determined. We aimed to identify fluoroscopic predictors of an acceptable capture threshold before Micra TPS deployment. METHODS: Sixty patients were successfull...

Descripción completa

Detalles Bibliográficos
Autores principales: Togashi, Ikuko, Sato, Toshiaki, Maeda , Akiko, Mohri, Takato, Katsume, Yumi, Tashiro, Mika, Momose, Yuichi, Nonoguchi, Noriko, Hoshida, Kyoko, Miwa, Yosuke, Ueda, Akiko, Soejima, Kyoko
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9315022/
https://www.ncbi.nlm.nih.gov/pubmed/35304791
http://dx.doi.org/10.1111/jce.15457
_version_ 1784754459963817984
author Togashi, Ikuko
Sato, Toshiaki
Maeda , Akiko
Mohri, Takato
Katsume, Yumi
Tashiro, Mika
Momose, Yuichi
Nonoguchi, Noriko
Hoshida, Kyoko
Miwa, Yosuke
Ueda, Akiko
Soejima, Kyoko
author_facet Togashi, Ikuko
Sato, Toshiaki
Maeda , Akiko
Mohri, Takato
Katsume, Yumi
Tashiro, Mika
Momose, Yuichi
Nonoguchi, Noriko
Hoshida, Kyoko
Miwa, Yosuke
Ueda, Akiko
Soejima, Kyoko
author_sort Togashi, Ikuko
collection PubMed
description INTRODUCTION: Few predictors of low capture threshold before the deployment of the Micra transcatheter pacing system (Micra TPS) have been determined. We aimed to identify fluoroscopic predictors of an acceptable capture threshold before Micra TPS deployment. METHODS: Sixty patients were successfully implanted with Micra TPS. Before deployment, gooseneck appearance of the catheter shaft was quantified using the angle between the tangent line of the shaft and the cup during diastole in the right anterior oblique (RAO) view. The direction of the device cup toward the ventricular septum was evaluated using the angle between the cup and the horizontal plane in the left anterior oblique (LAO) view. RESULTS: Of the 95 deployments we evaluated, 56 achieved an acceptable capture threshold of ≤2.0 V at 0.24 ms. In this acceptable threshold group, the deflection angle of the gooseneck shaft was significantly larger and the device cup was placed more horizontally with a lower elevation angle compared with those in the high threshold group. A deflection angle of ≥6° and an elevation angle of ≤30° were identified as the predictors of an acceptable capture threshold after deployment. An acceptable capture threshold was achieved in 24/31 (77.4%) patients in whom either angle criterion was satisfied at the first deployment. CONCLUSIONS: Diastolic gooseneck appearance of the delivery catheter in the RAO view or near‐horizontal direction in the LAO view predicts an acceptable capture threshold after deployment. The shape of the delivery catheter before deployment should be evaluated using multiple fluoroscopic views to ensure successful implantation of Micra TPS.
format Online
Article
Text
id pubmed-9315022
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-93150222022-07-30 Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system Togashi, Ikuko Sato, Toshiaki Maeda , Akiko Mohri, Takato Katsume, Yumi Tashiro, Mika Momose, Yuichi Nonoguchi, Noriko Hoshida, Kyoko Miwa, Yosuke Ueda, Akiko Soejima, Kyoko J Cardiovasc Electrophysiol Original Articles INTRODUCTION: Few predictors of low capture threshold before the deployment of the Micra transcatheter pacing system (Micra TPS) have been determined. We aimed to identify fluoroscopic predictors of an acceptable capture threshold before Micra TPS deployment. METHODS: Sixty patients were successfully implanted with Micra TPS. Before deployment, gooseneck appearance of the catheter shaft was quantified using the angle between the tangent line of the shaft and the cup during diastole in the right anterior oblique (RAO) view. The direction of the device cup toward the ventricular septum was evaluated using the angle between the cup and the horizontal plane in the left anterior oblique (LAO) view. RESULTS: Of the 95 deployments we evaluated, 56 achieved an acceptable capture threshold of ≤2.0 V at 0.24 ms. In this acceptable threshold group, the deflection angle of the gooseneck shaft was significantly larger and the device cup was placed more horizontally with a lower elevation angle compared with those in the high threshold group. A deflection angle of ≥6° and an elevation angle of ≤30° were identified as the predictors of an acceptable capture threshold after deployment. An acceptable capture threshold was achieved in 24/31 (77.4%) patients in whom either angle criterion was satisfied at the first deployment. CONCLUSIONS: Diastolic gooseneck appearance of the delivery catheter in the RAO view or near‐horizontal direction in the LAO view predicts an acceptable capture threshold after deployment. The shape of the delivery catheter before deployment should be evaluated using multiple fluoroscopic views to ensure successful implantation of Micra TPS. John Wiley and Sons Inc. 2022-03-25 2022-06 /pmc/articles/PMC9315022/ /pubmed/35304791 http://dx.doi.org/10.1111/jce.15457 Text en © 2022 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Original Articles
Togashi, Ikuko
Sato, Toshiaki
Maeda , Akiko
Mohri, Takato
Katsume, Yumi
Tashiro, Mika
Momose, Yuichi
Nonoguchi, Noriko
Hoshida, Kyoko
Miwa, Yosuke
Ueda, Akiko
Soejima, Kyoko
Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
title Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
title_full Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
title_fullStr Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
title_full_unstemmed Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
title_short Fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
title_sort fluoroscopic predictors of acceptable capture threshold during the implantation of the micra transcatheter pacing system
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9315022/
https://www.ncbi.nlm.nih.gov/pubmed/35304791
http://dx.doi.org/10.1111/jce.15457
work_keys_str_mv AT togashiikuko fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT satotoshiaki fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT maedaakiko fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT mohritakato fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT katsumeyumi fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT tashiromika fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT momoseyuichi fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT nonoguchinoriko fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT hoshidakyoko fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT miwayosuke fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT uedaakiko fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem
AT soejimakyoko fluoroscopicpredictorsofacceptablecapturethresholdduringtheimplantationofthemicratranscatheterpacingsystem