Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy
PURPOSE: The Local Effect Model version one (LEM I) is applied clinically across Europe to quantify the relative biological effectiveness (RBE) of carbon ion beams. It requires the full particle fluence spectrum differential in energy in each voxel as input parameter. Treatment planning systems (TPS...
Autores principales: | , , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9321194/ https://www.ncbi.nlm.nih.gov/pubmed/35416302 http://dx.doi.org/10.1002/mp.15652 |
_version_ | 1784755978398334976 |
---|---|
author | Resch, Andreas Franz Schafasand, Mansure Lackner, Niklas Niessen, Tom Beck, Staffan Elia, Alessio Boersma, David Grevillot, Loïc Fossati, Piero Glimelius, Lars Stock, Markus Georg, Dietmar Carlino, Antonio |
author_facet | Resch, Andreas Franz Schafasand, Mansure Lackner, Niklas Niessen, Tom Beck, Staffan Elia, Alessio Boersma, David Grevillot, Loïc Fossati, Piero Glimelius, Lars Stock, Markus Georg, Dietmar Carlino, Antonio |
author_sort | Resch, Andreas Franz |
collection | PubMed |
description | PURPOSE: The Local Effect Model version one (LEM I) is applied clinically across Europe to quantify the relative biological effectiveness (RBE) of carbon ion beams. It requires the full particle fluence spectrum differential in energy in each voxel as input parameter. Treatment planning systems (TPSs) use beamline‐specific look‐up tables generated with Monte Carlo (MC) codes. In this study, the changes in RBE weighted dose were quantified using different levels of details in the simulation or different MC codes. METHODS: The particle fluence differential in energy was simulated with FLUKA and Geant4 at 500 depths in water in 1‐mm steps for 58 initial carbon ion energies (between 120.0 and 402.8 MeV/u). A dedicated beam model was applied, including the full description of the Nozzle using GATE‐RTionV1.0 (Geant4.10.03p03). In addition, two tables generated with FLUKA were compared. The starting points of the FLUKA simulations were phase space (PhS) files from, firstly, the Geant4 nozzle simulations, and secondly, a clinical beam model where an analytic approach was used to mimic the beamline. Treatment plans (TPs) were generated with RayStation 8B (RaySearch Laboratories AB, Sweden) for cubic targets in water and 10 clinical patient cases using the clinical beam model. Subsequently, the RBE weighted dose was re‐computed using the two other fluence tables (FLUKA PhS or Geant4). RESULTS: The fluence spectra of the primary and secondary particles simulated with Geant4 and FLUKA generally agreed well for the primary particles. Differences were mainly observed for the secondary particles. Interchanging the two energy spectra (FLUKA vs. GEANT4) to calculate the RBE weighted dose distributions resulted in average deviations of less than 1% in the entrance up to the end of the target region, with a maximum local deviation at the distal edge of the target. In the fragment tail, larger discrepancies of up to 5% on average were found for deep‐seated targets. The patient and water phantom cases demonstrated similar results. CONCLUSION: RBE weighted doses agreed well within all tested setups, confirming the clinical beam model provided by the TPS vendor. Furthermore, the results showed that the open source and generally available MC code Geant4 (in particular using GATE or GATE‐RTion) can also be used to generate basic beam data required for RBE calculation in carbon ion therapy. |
format | Online Article Text |
id | pubmed-9321194 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-93211942022-07-30 Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy Resch, Andreas Franz Schafasand, Mansure Lackner, Niklas Niessen, Tom Beck, Staffan Elia, Alessio Boersma, David Grevillot, Loïc Fossati, Piero Glimelius, Lars Stock, Markus Georg, Dietmar Carlino, Antonio Med Phys COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY PURPOSE: The Local Effect Model version one (LEM I) is applied clinically across Europe to quantify the relative biological effectiveness (RBE) of carbon ion beams. It requires the full particle fluence spectrum differential in energy in each voxel as input parameter. Treatment planning systems (TPSs) use beamline‐specific look‐up tables generated with Monte Carlo (MC) codes. In this study, the changes in RBE weighted dose were quantified using different levels of details in the simulation or different MC codes. METHODS: The particle fluence differential in energy was simulated with FLUKA and Geant4 at 500 depths in water in 1‐mm steps for 58 initial carbon ion energies (between 120.0 and 402.8 MeV/u). A dedicated beam model was applied, including the full description of the Nozzle using GATE‐RTionV1.0 (Geant4.10.03p03). In addition, two tables generated with FLUKA were compared. The starting points of the FLUKA simulations were phase space (PhS) files from, firstly, the Geant4 nozzle simulations, and secondly, a clinical beam model where an analytic approach was used to mimic the beamline. Treatment plans (TPs) were generated with RayStation 8B (RaySearch Laboratories AB, Sweden) for cubic targets in water and 10 clinical patient cases using the clinical beam model. Subsequently, the RBE weighted dose was re‐computed using the two other fluence tables (FLUKA PhS or Geant4). RESULTS: The fluence spectra of the primary and secondary particles simulated with Geant4 and FLUKA generally agreed well for the primary particles. Differences were mainly observed for the secondary particles. Interchanging the two energy spectra (FLUKA vs. GEANT4) to calculate the RBE weighted dose distributions resulted in average deviations of less than 1% in the entrance up to the end of the target region, with a maximum local deviation at the distal edge of the target. In the fragment tail, larger discrepancies of up to 5% on average were found for deep‐seated targets. The patient and water phantom cases demonstrated similar results. CONCLUSION: RBE weighted doses agreed well within all tested setups, confirming the clinical beam model provided by the TPS vendor. Furthermore, the results showed that the open source and generally available MC code Geant4 (in particular using GATE or GATE‐RTion) can also be used to generate basic beam data required for RBE calculation in carbon ion therapy. John Wiley and Sons Inc. 2022-04-27 2022-06 /pmc/articles/PMC9321194/ /pubmed/35416302 http://dx.doi.org/10.1002/mp.15652 Text en © 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY Resch, Andreas Franz Schafasand, Mansure Lackner, Niklas Niessen, Tom Beck, Staffan Elia, Alessio Boersma, David Grevillot, Loïc Fossati, Piero Glimelius, Lars Stock, Markus Georg, Dietmar Carlino, Antonio Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
title | Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
title_full | Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
title_fullStr | Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
title_full_unstemmed | Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
title_short | Technical note: Impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
title_sort | technical note: impact of beamline‐specific particle energy spectra on clinical plans in carbon ion beam therapy |
topic | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9321194/ https://www.ncbi.nlm.nih.gov/pubmed/35416302 http://dx.doi.org/10.1002/mp.15652 |
work_keys_str_mv | AT reschandreasfranz technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT schafasandmansure technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT lacknerniklas technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT niessentom technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT beckstaffan technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT eliaalessio technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT boersmadavid technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT grevillotloic technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT fossatipiero technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT glimeliuslars technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT stockmarkus technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT georgdietmar technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy AT carlinoantonio technicalnoteimpactofbeamlinespecificparticleenergyspectraonclinicalplansincarbonionbeamtherapy |