Accurate Signal Conditioning for Pulsed-Current Synchronous Measurements
This paper describes a compact electronic system employing a synchronous demodulation measurement method for the acquisition of pulsed-current signals. The fabricated prototype shows superior performance in terms of signal-to-noise ratio in comparison to conventional instrumentation performing free-...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9321935/ https://www.ncbi.nlm.nih.gov/pubmed/35891039 http://dx.doi.org/10.3390/s22145360 |
Sumario: | This paper describes a compact electronic system employing a synchronous demodulation measurement method for the acquisition of pulsed-current signals. The fabricated prototype shows superior performance in terms of signal-to-noise ratio in comparison to conventional instrumentation performing free-running measurements, especially when extremely narrow pulses are concerned. It shows a reading error around 0.1% independently of the signal duty cycle (D) in the investigated D = 10(−4)–10(−3) range. Conversely, high-precision electrometers display reading errors as high as 30% for a D = 10(−4), which reduces to less than 1% only for D > 3 × 10(−3). Field tests demonstrate that the developed front-end/readout electronics is particularly effective when coupled to dosimeters irradiated with the X-rays sourced by a medical linear accelerator. Therefore, it may surely be exploited for the real-time monitoring of the dosimeter output current, as required in modern radiotherapy techniques employing ultra-narrow pulses of high-energy photons or nuclear particles. |
---|