Cargando…
Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy
BACKGROUND: Hypofractionation in prostate radiotherapy is of increasing interest. Steep dose gradients and a large weight on each individual fraction emphasize the need for motion management. Real‐time motion management techniques such as multileaf collimator (MLC) tracking or couch tracking typical...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9322296/ https://www.ncbi.nlm.nih.gov/pubmed/35395104 http://dx.doi.org/10.1002/mp.15660 |
_version_ | 1784756263456866304 |
---|---|
author | Muurholm, Casper Gammelmark Ravkilde, Thomas De Roover, Robin Skouboe, Simon Hansen, Rune Crijns, Wouter Depuydt, Tom Poulsen, Per R. |
author_facet | Muurholm, Casper Gammelmark Ravkilde, Thomas De Roover, Robin Skouboe, Simon Hansen, Rune Crijns, Wouter Depuydt, Tom Poulsen, Per R. |
author_sort | Muurholm, Casper Gammelmark |
collection | PubMed |
description | BACKGROUND: Hypofractionation in prostate radiotherapy is of increasing interest. Steep dose gradients and a large weight on each individual fraction emphasize the need for motion management. Real‐time motion management techniques such as multileaf collimator (MLC) tracking or couch tracking typically adjust for translational motion while rotations remain uncompensated with unknown dosimetric impact. PURPOSE: The purpose of this study is to demonstrate and validate dynamic real‐time rotation‐including dose reconstruction during radiotherapy experiments with and without MLC and couch tracking. METHODS: Real‐time dose reconstruction was performed using the in‐house developed software DoseTracker. DoseTracker receives streamed target positions and accelerator parameters during treatment delivery and uses a pencil beam algorithm with water density assumption to reconstruct the dose in a moving target. DoseTracker's ability to reconstruct motion‐induced dose errors in a dynamically rotating and translating target was investigated during three different scenarios: (1) no motion compensation and translational motion correction with (2) MLC tracking and (3) couch tracking. In each scenario, dose reconstruction was performed online and in real time during delivery of two dual‐arc volumetric‐modulated arc therapy prostate plans with a prescribed fraction dose of 7 Gy to the prostate and simultaneous intraprostatic lesion boosts with doses of at least 8 Gy, but up to 10 Gy as long as the organs at risk dose constraints were fulfilled. The plans were delivered to a pelvis phantom that replicated three patient‐measured motion traces using a rotational insert with 21 layers of EBT3 film spaced 2.5 mm apart. DoseTracker repeatedly calculated the actual motion‐including dose increment and the planned static dose increment since the last calculation in 84 500 points in the film stack. The experiments were performed with a TrueBeam accelerator with MLC and couch tracking based on electromagnetic transponders embedded in the film stack. The motion‐induced dose error was quantified as the difference between the final cumulative dose with motion and without motion using the 2D 2%/2 mm γ‐failure rate and the difference in dose to 95% of the clinical target volume (CTV ΔD (95%)) and the gross target volume (GTV ΔD (95%)) as well as the difference in dose to 0.1 cm(3) of the urethra, bladder, and rectum (ΔD (0.1CC)). The motion‐induced errors were compared between dose reconstructions and film measurements. RESULTS: The dose was reconstructed in all calculation points at a mean frequency of 4.7 Hz. The root‐mean‐square difference between real‐time reconstructed and film‐measured motion‐induced errors was 3.1%‐points (γ‐failure rate), 0.13 Gy (CTV ΔD (95%)), 0.23 Gy (GTV ΔD (95%)), 0.19 Gy (urethra ΔD (0.1CC)), 0.09 Gy (bladder ΔD (0.1CC)), and 0.07 Gy (rectum ΔD (0.1CC)). CONCLUSIONS: In a series of phantom experiments, online real‐time rotation‐including dose reconstruction was performed for the first time. The calculated motion‐induced errors agreed well with film measurements. The dose reconstruction provides a valuable tool for monitoring dose delivery and investigating the efficacy of advanced motion‐compensation techniques in the presence of translational and rotational motion. |
format | Online Article Text |
id | pubmed-9322296 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-93222962022-07-30 Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy Muurholm, Casper Gammelmark Ravkilde, Thomas De Roover, Robin Skouboe, Simon Hansen, Rune Crijns, Wouter Depuydt, Tom Poulsen, Per R. Med Phys THERAPEUTIC INTERVENTIONS BACKGROUND: Hypofractionation in prostate radiotherapy is of increasing interest. Steep dose gradients and a large weight on each individual fraction emphasize the need for motion management. Real‐time motion management techniques such as multileaf collimator (MLC) tracking or couch tracking typically adjust for translational motion while rotations remain uncompensated with unknown dosimetric impact. PURPOSE: The purpose of this study is to demonstrate and validate dynamic real‐time rotation‐including dose reconstruction during radiotherapy experiments with and without MLC and couch tracking. METHODS: Real‐time dose reconstruction was performed using the in‐house developed software DoseTracker. DoseTracker receives streamed target positions and accelerator parameters during treatment delivery and uses a pencil beam algorithm with water density assumption to reconstruct the dose in a moving target. DoseTracker's ability to reconstruct motion‐induced dose errors in a dynamically rotating and translating target was investigated during three different scenarios: (1) no motion compensation and translational motion correction with (2) MLC tracking and (3) couch tracking. In each scenario, dose reconstruction was performed online and in real time during delivery of two dual‐arc volumetric‐modulated arc therapy prostate plans with a prescribed fraction dose of 7 Gy to the prostate and simultaneous intraprostatic lesion boosts with doses of at least 8 Gy, but up to 10 Gy as long as the organs at risk dose constraints were fulfilled. The plans were delivered to a pelvis phantom that replicated three patient‐measured motion traces using a rotational insert with 21 layers of EBT3 film spaced 2.5 mm apart. DoseTracker repeatedly calculated the actual motion‐including dose increment and the planned static dose increment since the last calculation in 84 500 points in the film stack. The experiments were performed with a TrueBeam accelerator with MLC and couch tracking based on electromagnetic transponders embedded in the film stack. The motion‐induced dose error was quantified as the difference between the final cumulative dose with motion and without motion using the 2D 2%/2 mm γ‐failure rate and the difference in dose to 95% of the clinical target volume (CTV ΔD (95%)) and the gross target volume (GTV ΔD (95%)) as well as the difference in dose to 0.1 cm(3) of the urethra, bladder, and rectum (ΔD (0.1CC)). The motion‐induced errors were compared between dose reconstructions and film measurements. RESULTS: The dose was reconstructed in all calculation points at a mean frequency of 4.7 Hz. The root‐mean‐square difference between real‐time reconstructed and film‐measured motion‐induced errors was 3.1%‐points (γ‐failure rate), 0.13 Gy (CTV ΔD (95%)), 0.23 Gy (GTV ΔD (95%)), 0.19 Gy (urethra ΔD (0.1CC)), 0.09 Gy (bladder ΔD (0.1CC)), and 0.07 Gy (rectum ΔD (0.1CC)). CONCLUSIONS: In a series of phantom experiments, online real‐time rotation‐including dose reconstruction was performed for the first time. The calculated motion‐induced errors agreed well with film measurements. The dose reconstruction provides a valuable tool for monitoring dose delivery and investigating the efficacy of advanced motion‐compensation techniques in the presence of translational and rotational motion. John Wiley and Sons Inc. 2022-04-25 2022-06 /pmc/articles/PMC9322296/ /pubmed/35395104 http://dx.doi.org/10.1002/mp.15660 Text en © 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | THERAPEUTIC INTERVENTIONS Muurholm, Casper Gammelmark Ravkilde, Thomas De Roover, Robin Skouboe, Simon Hansen, Rune Crijns, Wouter Depuydt, Tom Poulsen, Per R. Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
title | Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
title_full | Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
title_fullStr | Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
title_full_unstemmed | Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
title_short | Experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
title_sort | experimental investigation of dynamic real‐time rotation‐including dose reconstruction during prostate tracking radiotherapy |
topic | THERAPEUTIC INTERVENTIONS |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9322296/ https://www.ncbi.nlm.nih.gov/pubmed/35395104 http://dx.doi.org/10.1002/mp.15660 |
work_keys_str_mv | AT muurholmcaspergammelmark experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT ravkildethomas experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT derooverrobin experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT skouboesimon experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT hansenrune experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT crijnswouter experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT depuydttom experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy AT poulsenperr experimentalinvestigationofdynamicrealtimerotationincludingdosereconstructionduringprostatetrackingradiotherapy |