Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation

INTRODUCTION: During radiofrequency ablation (RFA) using conventional RFA catheters (RFC), ~90% of the energy dissipates into the bloodstream/surrounding tissue. We hypothesized that a novel insulated‐tip ablation catheter (SMT) capable of blocking the radiofrequency path may focus most of the energ...

Descripción completa

Detalles Bibliográficos
Autores principales: Aryana, Arash, Irastorza, Ramiro M., Berjano, Enrique, Cohen, Richard J., Kraus, Jeffrey, Haghighi‐Mood, Ali, Reddy, Vivek Y., d'Avila, André
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9325397/
https://www.ncbi.nlm.nih.gov/pubmed/35322477
http://dx.doi.org/10.1111/jce.15461
_version_ 1784757041686904832
author Aryana, Arash
Irastorza, Ramiro M.
Berjano, Enrique
Cohen, Richard J.
Kraus, Jeffrey
Haghighi‐Mood, Ali
Reddy, Vivek Y.
d'Avila, André
author_facet Aryana, Arash
Irastorza, Ramiro M.
Berjano, Enrique
Cohen, Richard J.
Kraus, Jeffrey
Haghighi‐Mood, Ali
Reddy, Vivek Y.
d'Avila, André
author_sort Aryana, Arash
collection PubMed
description INTRODUCTION: During radiofrequency ablation (RFA) using conventional RFA catheters (RFC), ~90% of the energy dissipates into the bloodstream/surrounding tissue. We hypothesized that a novel insulated‐tip ablation catheter (SMT) capable of blocking the radiofrequency path may focus most of the energy into the targeted tissue while utilizing reduced power and irrigation. METHODS: This study evaluated the outcomes of RFA using SMT versus an RFC in silico, ex vivo, and in vivo. Radiofrequency applications were delivered over porcine myocardium (ex vivo) and porcine thigh muscle preparations superfused with heparinized blood (in vivo). Altogether, 274 radiofrequency applications were delivered using SMT (4–15 W, 2 or 20 ml/min) and 74 applications using RFC (30 W, 30 ml/min). RESULTS: RFA using SMT proved capable of directing 66.8% of the radiofrequency energy into the targeted tissue. Accordingly, low power–low irrigation RFA using SMT (8–12 W, 2 ml/min) yielded lesion sizes comparable with RFC, whereas high power–high irrigation (15 W, 20 ml/min) RFA with SMT yielded lesions larger than RFC (p < .05). Although SMT was associated with greater impedance drops ex vivo and in vivo, ablation using RFC was associated with increased charring/steam pop/tissue cavitation (p < .05). Lastly, lesions created with SMT were more homogeneous than RFC (p < .001). CONCLUSION: Low power–low irrigation (8–12 W, 2 ml/min) RFA using the novel SMT ablation catheter can create more uniform, but comparable‐sized lesions as RFC with reduced charring/steam pop/tissue cavitation. High power–high irrigation (15 W, 20 ml/min) RFA with SMT yields lesions larger than RFC.
format Online
Article
Text
id pubmed-9325397
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-93253972022-07-30 Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation Aryana, Arash Irastorza, Ramiro M. Berjano, Enrique Cohen, Richard J. Kraus, Jeffrey Haghighi‐Mood, Ali Reddy, Vivek Y. d'Avila, André J Cardiovasc Electrophysiol Original Articles INTRODUCTION: During radiofrequency ablation (RFA) using conventional RFA catheters (RFC), ~90% of the energy dissipates into the bloodstream/surrounding tissue. We hypothesized that a novel insulated‐tip ablation catheter (SMT) capable of blocking the radiofrequency path may focus most of the energy into the targeted tissue while utilizing reduced power and irrigation. METHODS: This study evaluated the outcomes of RFA using SMT versus an RFC in silico, ex vivo, and in vivo. Radiofrequency applications were delivered over porcine myocardium (ex vivo) and porcine thigh muscle preparations superfused with heparinized blood (in vivo). Altogether, 274 radiofrequency applications were delivered using SMT (4–15 W, 2 or 20 ml/min) and 74 applications using RFC (30 W, 30 ml/min). RESULTS: RFA using SMT proved capable of directing 66.8% of the radiofrequency energy into the targeted tissue. Accordingly, low power–low irrigation RFA using SMT (8–12 W, 2 ml/min) yielded lesion sizes comparable with RFC, whereas high power–high irrigation (15 W, 20 ml/min) RFA with SMT yielded lesions larger than RFC (p < .05). Although SMT was associated with greater impedance drops ex vivo and in vivo, ablation using RFC was associated with increased charring/steam pop/tissue cavitation (p < .05). Lastly, lesions created with SMT were more homogeneous than RFC (p < .001). CONCLUSION: Low power–low irrigation (8–12 W, 2 ml/min) RFA using the novel SMT ablation catheter can create more uniform, but comparable‐sized lesions as RFC with reduced charring/steam pop/tissue cavitation. High power–high irrigation (15 W, 20 ml/min) RFA with SMT yields lesions larger than RFC. John Wiley and Sons Inc. 2022-04-12 2022-06 /pmc/articles/PMC9325397/ /pubmed/35322477 http://dx.doi.org/10.1111/jce.15461 Text en © 2022 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Original Articles
Aryana, Arash
Irastorza, Ramiro M.
Berjano, Enrique
Cohen, Richard J.
Kraus, Jeffrey
Haghighi‐Mood, Ali
Reddy, Vivek Y.
d'Avila, André
Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
title Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
title_full Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
title_fullStr Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
title_full_unstemmed Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
title_short Radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
title_sort radiofrequency ablation using a novel insulated‐tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9325397/
https://www.ncbi.nlm.nih.gov/pubmed/35322477
http://dx.doi.org/10.1111/jce.15461
work_keys_str_mv AT aryanaarash radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT irastorzaramirom radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT berjanoenrique radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT cohenrichardj radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT krausjeffrey radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT haghighimoodali radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT reddyviveky radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation
AT davilaandre radiofrequencyablationusinganovelinsulatedtipablationcathetercancreateuniformlesionscomparableinsizetoconventionalirrigatedablationcatheterswhileusingafractionoftheenergyandirrigation