Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700
PURPOSE: To assess the comparability of ocular biometry measurements and refractive outcomes between ANTERION and IOL Master 700. METHODS: This comparative prospective study was conducted from December 2020 to February 2021. A total of 225 patients who had cataracts were enrolled for the study and d...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9332939/ https://www.ncbi.nlm.nih.gov/pubmed/35502033 http://dx.doi.org/10.4103/ijo.IJO_2433_21 |
_version_ | 1784758769774755840 |
---|---|
author | Panda, Amita Nanda, Ashok Sahoo, Kalyani |
author_facet | Panda, Amita Nanda, Ashok Sahoo, Kalyani |
author_sort | Panda, Amita |
collection | PubMed |
description | PURPOSE: To assess the comparability of ocular biometry measurements and refractive outcomes between ANTERION and IOL Master 700. METHODS: This comparative prospective study was conducted from December 2020 to February 2021. A total of 225 patients who had cataracts were enrolled for the study and different parameters such as anterior keratometry (Steep K, Flat K) with axis, Sim K, central corneal thickness (CCT), aqueous depth (AQD), lens thickness (LT), and axial length (AL), white-to-white (WTW) were evaluated in IOL master 700 first and then with ANTERION. Finally, 203 patients completed the 6-week follow-up and the postoperative refraction was done at the 6(th) week. To assess the agreement between the devices, intraclass coefficient (ICC) and Bland–Altman analysis with 95% limits of agreement (LoA) were used. To analyze the agreement for postoperative residual refractive error between the two devices, Kappa statistics were used. RESULTS: The mean difference for steep K, flat K, and Sim K between ANTERION and IOL Master 700 were − 0.18 +/−0.26 D, −0.13+/− 0.28D, −0.15 +/−0.23, respectively. The CCT, ACD, and LT also showed excellent agreement (ICC > 0.9) but the similarity for the keratometry axis was not up to the mark (ICC = 0.794). For postoperative refractive outcomes, the Kappa value was 0.437, indicating moderate agreement. CONCLUSION: ANTERION showed a good agreement for the majority of parameters with IOLMaster 700 in measuring ocular biometry, except for the keratometry. The accuracy of the intraocular lens power calculations was clinically acceptable with both biometers though the IOL power given by ANTERION remained slightly on the hypermetropic side. |
format | Online Article Text |
id | pubmed-9332939 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-93329392022-07-29 Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 Panda, Amita Nanda, Ashok Sahoo, Kalyani Indian J Ophthalmol Special Focus, Cataract Surgery, Original Article PURPOSE: To assess the comparability of ocular biometry measurements and refractive outcomes between ANTERION and IOL Master 700. METHODS: This comparative prospective study was conducted from December 2020 to February 2021. A total of 225 patients who had cataracts were enrolled for the study and different parameters such as anterior keratometry (Steep K, Flat K) with axis, Sim K, central corneal thickness (CCT), aqueous depth (AQD), lens thickness (LT), and axial length (AL), white-to-white (WTW) were evaluated in IOL master 700 first and then with ANTERION. Finally, 203 patients completed the 6-week follow-up and the postoperative refraction was done at the 6(th) week. To assess the agreement between the devices, intraclass coefficient (ICC) and Bland–Altman analysis with 95% limits of agreement (LoA) were used. To analyze the agreement for postoperative residual refractive error between the two devices, Kappa statistics were used. RESULTS: The mean difference for steep K, flat K, and Sim K between ANTERION and IOL Master 700 were − 0.18 +/−0.26 D, −0.13+/− 0.28D, −0.15 +/−0.23, respectively. The CCT, ACD, and LT also showed excellent agreement (ICC > 0.9) but the similarity for the keratometry axis was not up to the mark (ICC = 0.794). For postoperative refractive outcomes, the Kappa value was 0.437, indicating moderate agreement. CONCLUSION: ANTERION showed a good agreement for the majority of parameters with IOLMaster 700 in measuring ocular biometry, except for the keratometry. The accuracy of the intraocular lens power calculations was clinically acceptable with both biometers though the IOL power given by ANTERION remained slightly on the hypermetropic side. Wolters Kluwer - Medknow 2022-05 2022-04-28 /pmc/articles/PMC9332939/ /pubmed/35502033 http://dx.doi.org/10.4103/ijo.IJO_2433_21 Text en Copyright: © 2022 Indian Journal of Ophthalmology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Special Focus, Cataract Surgery, Original Article Panda, Amita Nanda, Ashok Sahoo, Kalyani Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 |
title | Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 |
title_full | Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 |
title_fullStr | Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 |
title_full_unstemmed | Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 |
title_short | Comparison of ocular biometry and refractive outcome between ANTERION and IOL Master 700 |
title_sort | comparison of ocular biometry and refractive outcome between anterion and iol master 700 |
topic | Special Focus, Cataract Surgery, Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9332939/ https://www.ncbi.nlm.nih.gov/pubmed/35502033 http://dx.doi.org/10.4103/ijo.IJO_2433_21 |
work_keys_str_mv | AT pandaamita comparisonofocularbiometryandrefractiveoutcomebetweenanterionandiolmaster700 AT nandaashok comparisonofocularbiometryandrefractiveoutcomebetweenanterionandiolmaster700 AT sahookalyani comparisonofocularbiometryandrefractiveoutcomebetweenanterionandiolmaster700 |