Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients
BACKGROUND: Quantitative pupillometry is part of multimodal neuroprognostication of comatose patients after out-of-hospital cardiac arrest (OHCA). However, the reproducibility, repeatability, and reliability of quantitative pupillometry in this setting have not been investigated. METHODS: In a prosp...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9333219/ https://www.ncbi.nlm.nih.gov/pubmed/35901103 http://dx.doi.org/10.1371/journal.pone.0272303 |
_version_ | 1784758824010252288 |
---|---|
author | Nyholm, Benjamin Obling, Laust Hassager, Christian Grand, Johannes Møller, Jacob Othman, Marwan Kondziella, Daniel Kjaergaard, Jesper |
author_facet | Nyholm, Benjamin Obling, Laust Hassager, Christian Grand, Johannes Møller, Jacob Othman, Marwan Kondziella, Daniel Kjaergaard, Jesper |
author_sort | Nyholm, Benjamin |
collection | PubMed |
description | BACKGROUND: Quantitative pupillometry is part of multimodal neuroprognostication of comatose patients after out-of-hospital cardiac arrest (OHCA). However, the reproducibility, repeatability, and reliability of quantitative pupillometry in this setting have not been investigated. METHODS: In a prospective blinded validation study, we compared manual and quantitative measurements of pupil size. Observer and device variability for all available parameters are expressed as mean difference (bias), limits of agreement (LoA), and reliability expressed as intraclass correlation coefficients (ICC) with a 95% confidence interval. RESULTS: Fifty-six unique quadrupled sets of measurement derived from 14 sedated and comatose patients (mean age 70±12 years) were included. For manually measured pupil size, inter-observer bias was -0.14±0.44 mm, LoA of -1.00 to 0.71 mm, and ICC at 0.92 (0.86–0.95). For quantitative pupillometry, we found bias at 0.03±0.17 mm, LoA of -0.31 to 0.36 mm and ICCs at 0.99. Quantitative pupillometry also yielded lower bias and LoA and higher ICC for intra-observer and inter-device measurements. Correlation between manual and automated pupillometry was better in larger pupils, and quantitative pupillometry had less variability and higher ICC, when assessing small pupils. Further, observers failed to detect 26% of the quantitatively estimated abnormal reactivity with manual assessment. We found ICC >0.91 for all quantitative pupillary response parameters (except for latency with ICC 0.81–0.91). CONCLUSION: Automated quantitative pupillometry has excellent reliability and twice the reproducibility and repeatability than manual pupillometry. This study further presents novel estimates of variability for all quantitative pupillary response parameters with excellent reliability. |
format | Online Article Text |
id | pubmed-9333219 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-93332192022-07-29 Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients Nyholm, Benjamin Obling, Laust Hassager, Christian Grand, Johannes Møller, Jacob Othman, Marwan Kondziella, Daniel Kjaergaard, Jesper PLoS One Research Article BACKGROUND: Quantitative pupillometry is part of multimodal neuroprognostication of comatose patients after out-of-hospital cardiac arrest (OHCA). However, the reproducibility, repeatability, and reliability of quantitative pupillometry in this setting have not been investigated. METHODS: In a prospective blinded validation study, we compared manual and quantitative measurements of pupil size. Observer and device variability for all available parameters are expressed as mean difference (bias), limits of agreement (LoA), and reliability expressed as intraclass correlation coefficients (ICC) with a 95% confidence interval. RESULTS: Fifty-six unique quadrupled sets of measurement derived from 14 sedated and comatose patients (mean age 70±12 years) were included. For manually measured pupil size, inter-observer bias was -0.14±0.44 mm, LoA of -1.00 to 0.71 mm, and ICC at 0.92 (0.86–0.95). For quantitative pupillometry, we found bias at 0.03±0.17 mm, LoA of -0.31 to 0.36 mm and ICCs at 0.99. Quantitative pupillometry also yielded lower bias and LoA and higher ICC for intra-observer and inter-device measurements. Correlation between manual and automated pupillometry was better in larger pupils, and quantitative pupillometry had less variability and higher ICC, when assessing small pupils. Further, observers failed to detect 26% of the quantitatively estimated abnormal reactivity with manual assessment. We found ICC >0.91 for all quantitative pupillary response parameters (except for latency with ICC 0.81–0.91). CONCLUSION: Automated quantitative pupillometry has excellent reliability and twice the reproducibility and repeatability than manual pupillometry. This study further presents novel estimates of variability for all quantitative pupillary response parameters with excellent reliability. Public Library of Science 2022-07-28 /pmc/articles/PMC9333219/ /pubmed/35901103 http://dx.doi.org/10.1371/journal.pone.0272303 Text en © 2022 Nyholm et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Nyholm, Benjamin Obling, Laust Hassager, Christian Grand, Johannes Møller, Jacob Othman, Marwan Kondziella, Daniel Kjaergaard, Jesper Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
title | Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
title_full | Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
title_fullStr | Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
title_full_unstemmed | Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
title_short | Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
title_sort | superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9333219/ https://www.ncbi.nlm.nih.gov/pubmed/35901103 http://dx.doi.org/10.1371/journal.pone.0272303 |
work_keys_str_mv | AT nyholmbenjamin superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT oblinglaust superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT hassagerchristian superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT grandjohannes superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT møllerjacob superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT othmanmarwan superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT kondzielladaniel superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients AT kjaergaardjesper superiorreproducibilityandrepeatabilityinautomatedquantitativepupillometrycomparedtostandardmanualassessmentandquantitativepupillaryresponseparameterspresenthighreliabilityincriticallyillcardiacpatients |