A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device

Device closure of an eccentric atrial septal defect can be challenging and needs technical modifications to avoid unnecessary complications. Here, we present a case of a 45-year-old woman who underwent device closure of an eccentric defect with a large device. The patient developed pericardial effus...

Descripción completa

Detalles Bibliográficos
Autores principales: Barik, Ramachandra, Mahapatra, Rudrapratap
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9337235/
https://www.ncbi.nlm.nih.gov/pubmed/35910697
http://dx.doi.org/10.7759/cureus.27447
Descripción
Sumario:Device closure of an eccentric atrial septal defect can be challenging and needs technical modifications to avoid unnecessary complications. Here, we present a case of a 45-year-old woman who underwent device closure of an eccentric defect with a large device. The patient developed pericardial effusion and left-sided pleural effusion due to injury to the junction of right atrium and superior vena cava because of the malalignment of the delivery sheath and left atrial disc before the device was pulled across the eccentric defect despite releasing the left atrial disc in the left atrium in place of the left pulmonary vein. These two serious complications were managed conservatively with close monitoring of the case during and after the procedure.