Cargando…

A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device

Device closure of an eccentric atrial septal defect can be challenging and needs technical modifications to avoid unnecessary complications. Here, we present a case of a 45-year-old woman who underwent device closure of an eccentric defect with a large device. The patient developed pericardial effus...

Descripción completa

Detalles Bibliográficos
Autores principales: Barik, Ramachandra, Mahapatra, Rudrapratap
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9337235/
https://www.ncbi.nlm.nih.gov/pubmed/35910697
http://dx.doi.org/10.7759/cureus.27447
_version_ 1784759708731572224
author Barik, Ramachandra
Mahapatra, Rudrapratap
author_facet Barik, Ramachandra
Mahapatra, Rudrapratap
author_sort Barik, Ramachandra
collection PubMed
description Device closure of an eccentric atrial septal defect can be challenging and needs technical modifications to avoid unnecessary complications. Here, we present a case of a 45-year-old woman who underwent device closure of an eccentric defect with a large device. The patient developed pericardial effusion and left-sided pleural effusion due to injury to the junction of right atrium and superior vena cava because of the malalignment of the delivery sheath and left atrial disc before the device was pulled across the eccentric defect despite releasing the left atrial disc in the left atrium in place of the left pulmonary vein. These two serious complications were managed conservatively with close monitoring of the case during and after the procedure.
format Online
Article
Text
id pubmed-9337235
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-93372352022-07-30 A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device Barik, Ramachandra Mahapatra, Rudrapratap Cureus Cardiac/Thoracic/Vascular Surgery Device closure of an eccentric atrial septal defect can be challenging and needs technical modifications to avoid unnecessary complications. Here, we present a case of a 45-year-old woman who underwent device closure of an eccentric defect with a large device. The patient developed pericardial effusion and left-sided pleural effusion due to injury to the junction of right atrium and superior vena cava because of the malalignment of the delivery sheath and left atrial disc before the device was pulled across the eccentric defect despite releasing the left atrial disc in the left atrium in place of the left pulmonary vein. These two serious complications were managed conservatively with close monitoring of the case during and after the procedure. Cureus 2022-07-29 /pmc/articles/PMC9337235/ /pubmed/35910697 http://dx.doi.org/10.7759/cureus.27447 Text en Copyright © 2022, Barik et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Cardiac/Thoracic/Vascular Surgery
Barik, Ramachandra
Mahapatra, Rudrapratap
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device
title A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device
title_full A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device
title_fullStr A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device
title_full_unstemmed A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device
title_short A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large Device
title_sort case of device closure of an eccentric atrial septal defect using a large device
topic Cardiac/Thoracic/Vascular Surgery
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9337235/
https://www.ncbi.nlm.nih.gov/pubmed/35910697
http://dx.doi.org/10.7759/cureus.27447
work_keys_str_mv AT barikramachandra acaseofdeviceclosureofaneccentricatrialseptaldefectusingalargedevice
AT mahapatrarudrapratap acaseofdeviceclosureofaneccentricatrialseptaldefectusingalargedevice
AT barikramachandra caseofdeviceclosureofaneccentricatrialseptaldefectusingalargedevice
AT mahapatrarudrapratap caseofdeviceclosureofaneccentricatrialseptaldefectusingalargedevice