Effectuality study of a 3D motion correction algorithm in C-arm CTs of severely impaired image quality during transarterial chemoembolization
BACKGROUND: To evaluate effectivity of a 3D-motion correction algorithm in C-Arm CTs (CACT) with limited image quality (IQ) during transarterial chemoembolization (TACE). METHODS: From 1/2015–5/2021, 644 CACTs were performed in patients during TACE. Of these, 27 CACTs in 26 patients (18 m, 8f; 69.7 ...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9338620/ https://www.ncbi.nlm.nih.gov/pubmed/35908026 http://dx.doi.org/10.1186/s40644-022-00473-3 |
Sumario: | BACKGROUND: To evaluate effectivity of a 3D-motion correction algorithm in C-Arm CTs (CACT) with limited image quality (IQ) during transarterial chemoembolization (TACE). METHODS: From 1/2015–5/2021, 644 CACTs were performed in patients during TACE. Of these, 27 CACTs in 26 patients (18 m, 8f; 69.7 years ± 10.7 SD) of limited IQ were included. Post-processing of the original raw-data sets (CACT(Org)) included application of a 3D-motion correction algorithm and bone segmentation (CACT(MC_no_bone)). Four radiologists (R1-4) compared the images by choosing their preferred dataset and recommending repeat acquisition in case of severe IQ-impairment. R1,2 performed additional grading of intrahepatic vessel visualization, presence/extent of movement artifacts, and overall IQ. RESULTS: R1,2 demonstrated excellent interobserver agreement for overall IQ (ICC 0.79,p < 0.01) and the five-point vessel visualization scale before and after post-processing of the datasets (ICC 0.78,p < 0.01). Post-processing caused significant improvement, with overall IQ improving from 2.63 (CACT(Org)) to 1.39 (CACT(MC_no_bone);p < 0.01) and a decrease in the mean distance of identifiable, subcapsular vessels to the liver capsule by 4 mm (p < 0.01). This proved especially true for datasets with low parenchymal and high hepatic artery contrast. A good interobserver agreement (ICC = 0.73) was recorded concerning the presence of motion artifacts, with significantly less discernible motion after post-processing (CACT(Org):1.31 ± 1.67, CACT(MC_no_bone):1.00 ± 1.34, p < 0.01). Of the 27 datasets, ≥ 23 CACT(MC_no_bone) were preferred, with identical datasets chosen by the readers to show benefit from the algorithm. CONCLUSION: Application of a 3D-motion correction algorithm significantly improved IQ in diagnostically limited CACTs during TACE, with the potential to decrease repeat acquisitions. |
---|