Tissue Dielectric Constant of the Lower Leg as an Index of Skin Water: Temporal Variations

Tissue dielectric constant (TDC) measurements are a convenient, reliable, and accurate way to noninvasively access local tissue water content and its change with time or treatment. The method has been widely used for upper extremity lymphedema assessments but less so for lower extremities. For lower...

Descripción completa

Detalles Bibliográficos
Autor principal: Mayrovitz, Harvey N
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9339369/
https://www.ncbi.nlm.nih.gov/pubmed/35923478
http://dx.doi.org/10.7759/cureus.26506
Descripción
Sumario:Tissue dielectric constant (TDC) measurements are a convenient, reliable, and accurate way to noninvasively access local tissue water content and its change with time or treatment. The method has been widely used for upper extremity lymphedema assessments but less so for lower extremities. For lower extremity assessments, it would be useful to have an estimate of the expected inter-leg TDC differentials of normal non-affected legs to help interpret measured inter-leg differentials when such measurements are done in patients with lower extremity edema or lymphedema. The goals of this study were to determine normal inter-leg TDC differences, quantify TDC temporal changes during 60 minutes of supine lying in a group of 10 middle-aged women (42-57 years), and evaluate the change in TDC values as measured throughout the day from 0800 to 2000 hours in a group of 10 younger women (23-28 years). The overall findings indicate that TDC measurements, as an assessment parameter of lower extremity localized skin-to-fat tissue water, are only minimally dependent on potential confounding factors such as 60-minute supine lying or the time of day at which the measurements are made. If the small sample size of the present study is reflective of the larger population, an inter-leg TDC ratio exceeding a value of 1.2 is suggested as a threshold for judging unilateral lower extremity edema or lymphedema. This possibility needs assessment in women with these conditions.