Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children
PURPOSE: To compare the difference and agreement of axial length (AL) and anterior segment parameters obtained from the MYAH device with Pentacam AXL and IOLMaster 700 in myopic children. METHODS: The study included 60 eyes of 60 myopic children. AL, keratometry (K), and horizontal corneal diameter...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Netherlands
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9340674/ https://www.ncbi.nlm.nih.gov/pubmed/35908260 http://dx.doi.org/10.1007/s10792-022-02444-w |
_version_ | 1784760443574681600 |
---|---|
author | Sabur, Huri Takes, Omer |
author_facet | Sabur, Huri Takes, Omer |
author_sort | Sabur, Huri |
collection | PubMed |
description | PURPOSE: To compare the difference and agreement of axial length (AL) and anterior segment parameters obtained from the MYAH device with Pentacam AXL and IOLMaster 700 in myopic children. METHODS: The study included 60 eyes of 60 myopic children. AL, keratometry (K), and horizontal corneal diameter (CD) were measured with Pentacam AXL, IOLMaster 700, and MYAH, respectively. The Friedman test was used to assess the differences. The Intraclass Correlation Coefficient (ICC) and Bland–Altman plots were used to assess the consistency of measurements. RESULTS: The mean age was 10.2 ± 1.8 years (7–16 years). No statistically significant difference was determined between the Pentacam AXL, IOLMaster 700, and MYAH devices in terms of mean AL values (23.61 ± 1.42, 23.62 ± 1.45, 23.61 ± 1.42, respectively) (p = 0.06). The difference between devices in the mean steep K, flat K, mean K, and CD was statistically significant but clinically insignificant (steep K; 44.45 ± 1.25, 44.59 ± 1.23, 44.51 ± 1.24, flat K; 43.29 ± 1.28, 43.43 ± 1.29, 43.35 ± 1.30, mean K; 43.85 ± 1.21, 44.00 ± 1.19, 43.94 ± 1.20, and CD; 11.90 ± 0.34, 12.11 ± 0.38, 11.96 ± 0.31, respectively; p < 0.05). ICC and Bland–Altman plot analysis revealed a high correlation between the three devices in AL, steep K, flat K, mean K, and CD measurements. CONCLUSION: There was a quite good agreement between the MYAH, Pentacam AXL, and IOLMaster 700 devices regarding AL and anterior segment parameters. MYAH provides reliable measurements and will be a good option in the diagnosis of and follow-up with myopic children. |
format | Online Article Text |
id | pubmed-9340674 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Springer Netherlands |
record_format | MEDLINE/PubMed |
spelling | pubmed-93406742022-08-01 Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children Sabur, Huri Takes, Omer Int Ophthalmol Original Paper PURPOSE: To compare the difference and agreement of axial length (AL) and anterior segment parameters obtained from the MYAH device with Pentacam AXL and IOLMaster 700 in myopic children. METHODS: The study included 60 eyes of 60 myopic children. AL, keratometry (K), and horizontal corneal diameter (CD) were measured with Pentacam AXL, IOLMaster 700, and MYAH, respectively. The Friedman test was used to assess the differences. The Intraclass Correlation Coefficient (ICC) and Bland–Altman plots were used to assess the consistency of measurements. RESULTS: The mean age was 10.2 ± 1.8 years (7–16 years). No statistically significant difference was determined between the Pentacam AXL, IOLMaster 700, and MYAH devices in terms of mean AL values (23.61 ± 1.42, 23.62 ± 1.45, 23.61 ± 1.42, respectively) (p = 0.06). The difference between devices in the mean steep K, flat K, mean K, and CD was statistically significant but clinically insignificant (steep K; 44.45 ± 1.25, 44.59 ± 1.23, 44.51 ± 1.24, flat K; 43.29 ± 1.28, 43.43 ± 1.29, 43.35 ± 1.30, mean K; 43.85 ± 1.21, 44.00 ± 1.19, 43.94 ± 1.20, and CD; 11.90 ± 0.34, 12.11 ± 0.38, 11.96 ± 0.31, respectively; p < 0.05). ICC and Bland–Altman plot analysis revealed a high correlation between the three devices in AL, steep K, flat K, mean K, and CD measurements. CONCLUSION: There was a quite good agreement between the MYAH, Pentacam AXL, and IOLMaster 700 devices regarding AL and anterior segment parameters. MYAH provides reliable measurements and will be a good option in the diagnosis of and follow-up with myopic children. Springer Netherlands 2022-07-31 2023 /pmc/articles/PMC9340674/ /pubmed/35908260 http://dx.doi.org/10.1007/s10792-022-02444-w Text en © The Author(s), under exclusive licence to Springer Nature B.V. 2022, Springer Nature or its licensor holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law. This article is made available via the PMC Open Access Subset for unrestricted research re-use and secondary analysis in any form or by any means with acknowledgement of the original source. These permissions are granted for the duration of the World Health Organization (WHO) declaration of COVID-19 as a global pandemic. |
spellingShingle | Original Paper Sabur, Huri Takes, Omer Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children |
title | Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children |
title_full | Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children |
title_fullStr | Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children |
title_full_unstemmed | Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children |
title_short | Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children |
title_sort | agreement of axial length and anterior segment parameters measured with the myah device compared to pentacam axl and iolmaster 700 in myopic children |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9340674/ https://www.ncbi.nlm.nih.gov/pubmed/35908260 http://dx.doi.org/10.1007/s10792-022-02444-w |
work_keys_str_mv | AT saburhuri agreementofaxiallengthandanteriorsegmentparametersmeasuredwiththemyahdevicecomparedtopentacamaxlandiolmaster700inmyopicchildren AT takesomer agreementofaxiallengthandanteriorsegmentparametersmeasuredwiththemyahdevicecomparedtopentacamaxlandiolmaster700inmyopicchildren |